Retention and wear of resin matrix attachments for implant overdentures

被引:24
作者
Wichmann, Niklas [1 ]
Kern, Matthias [1 ]
Taylor, Thomas [2 ]
Wille, Sebastian [1 ]
Passia, Nicole [1 ]
机构
[1] Univ Kiel, Sch Dent, Dept Prosthodont Propaedeut & Dent Mat, Arnold Heller Str 16, D-24105 Kiel, Germany
[2] UConn Sch Dent Med, Dept Reconstruct Sci, Farmington, CT USA
关键词
Implant-overdenture; Attachment system; Wear; PEEK; PEKK; Implant; MANDIBULAR 2-IMPLANT OVERDENTURES; LOCATOR ATTACHMENTS; STUD ATTACHMENTS; PREFABRICATED ATTACHMENTS; CONVENTIONAL DENTURES; PATIENTS SATISFACTION; PATIENT SATISFACTION; DENTAL PROSTHESES; LONG-TERM; SYSTEMS;
D O I
10.1016/j.jmbbm.2020.103901
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Objectives: The aim of the present laboratory study was to investigate the retentive properties and the wear of three different resin matrix attachments for implant overdentures as well as to assess the effects of implant angulation. Methods and materials: Three attachment systems with either polyetheretherketone inserts (PEEK; Novaloc, Straumann, Basel, Switzerland), polyetherketoneketone inserts (PEKK; CM Loc, Cendres + Metaux, Biel, Switzerland) or nylon inserts (Locator R-Tx, Zest Dental Solutions, Escondido, California, USA) were evaluated. The patrices were connected to their implant analogues and fixed in a resin cast at implant angulations of 0 degrees and 15 degrees. The corresponding matrices with inserts were fixed in a stylized unilaterally removable dental prothesis. To simulate masticatory forces 30,000 insertion and removal cycles with an eccentric load of 100 N at a distance of 12 mm were performed in a chewing simulator. The retention forces were measured. The wear patterns were examined using a scanning electron microscope (SEM). The statistical analysis was performed using a one-way ANOVA followed by a Games-Howell test and a Friedman test. Results: All attachment systems revealed a significant decrease in retention after testing, except PEEK 0 degrees/0 degrees and PEEK 15 degrees/15 degrees. PEEK and PEKK attachment systems showed significantly higher retentive forces than the nylon attachment system between 5000 and 30,000 cycles. The implant angulation did not significantly influence the retention behavior for any of the three attachment systems. Patrices showed no signs of wear, whereas the matrix inserts displayed signs of deformation. Conclusion: Within the limitations of the present study, attachments with PEEK and PEKK inserts combined with titanium patrices are favorable for long-term use, both for orthogonal and tilted implants. All three attachment systems showed a high variability of the retentive forces at baseline and for subsequent cycles. This should be taken into consideration for clinical use.
引用
收藏
页数:10
相关论文
共 48 条
  • [1] THE IN VITRO EFFECT OF DIFFERENT IMPLANT ANGULATIONS AND CYCLIC DISLODGEMENT ON THE RETENTIVE PROPERTIES OF AN OVERDENTURE ATTACHMENT SYSTEM
    Al-Ghafli, Suhail Ali
    Michalakis, Konstantinos X.
    Hirayama, Hiroshi
    Kang, Kiho
    [J]. JOURNAL OF PROSTHETIC DENTISTRY, 2009, 102 (03) : 140 - 147
  • [2] Alsabeeha NHM, 2009, INT J PROSTHODONT, V22, P429
  • [3] Comparative study of four retentive anchor systems for implant supported overdentures - retention force changes
    Bayer, Stefan
    Steinheuser, Daniela
    Gruener, Manfred
    Keilig, Ludger
    Enkling, Norbert
    Stark, Helmut
    Mues, Sebastian
    [J]. GERODONTOLOGY, 2009, 26 (04) : 268 - 272
  • [4] CLINICAL LONG-TERM STUDY OF COMPLETE DENTURE WEARERS
    BERGMAN, B
    CARLSSON, GE
    [J]. JOURNAL OF PROSTHETIC DENTISTRY, 1985, 53 (01) : 56 - 61
  • [5] In vitro retention force changes of prefabricated attachments for overdentures
    Besimo, CE
    Guarneri, A
    [J]. JOURNAL OF ORAL REHABILITATION, 2003, 30 (07) : 671 - 678
  • [6] Improving masticatory performance, bite force, nutritional state and patient's satisfaction with implant overdentures: a systematic review of the literature
    Boven, G. C.
    Raghoebar, G. M.
    Vissink, A.
    Meijer, H. J. A.
    [J]. JOURNAL OF ORAL REHABILITATION, 2015, 42 (03) : 220 - 233
  • [7] Biomechanics of polyaryletherketone rod composites and titanium rods for posterior lumbosacral instrumentation
    Bruner, Harlan J.
    Guan, Yabo
    Yoganandan, Narayan
    Pintar, Frank A.
    Maiman, Dennis J.
    Slivka, Michael A.
    [J]. JOURNAL OF NEUROSURGERY-SPINE, 2010, 13 (06) : 766 - 772
  • [8] Retentive Properties of Two Stud Attachments with Polyetherketoneketone or Nylon Insert in Mandibular Implant Overdentures
    Choi, Jae-Won
    Yun, Bo-Hyeok
    Jeong, Chang-Mo
    Huh, Jung-Bo
    [J]. INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2018, 33 (05) : 1079 - 1088
  • [9] Stud attachments for the mandibular implant-retained overdentures: Prosthetic complications. A literature review
    Daou, Elie E.
    [J]. SAUDI DENTAL JOURNAL, 2013, 25 (02) : 53 - 60
  • [10] Effect of Distal Implant Inclination on Dislodging Forces of Different Locator Attachments Used for Mandibular Overdentures: An In Vitro Study
    ELsyad, Moustafa Abdou
    Emera, Radwa Mohsen
    Ashmawy, Tarek Mohy
    [J]. JOURNAL OF PROSTHODONTICS-IMPLANT ESTHETIC AND RECONSTRUCTIVE DENTISTRY, 2019, 28 (02): : E666 - E674