The effect of rib and lung heterogeneities on the computed dose to lung in Ir-192 High-Dose-Rate breast brachytherapy: Monte Carlo versus a treatment planning system

被引:2
|
作者
Yazdi, Hossein Salehi [3 ]
Shamsaei, Mojtaba [3 ]
Jaberi, Ramin [2 ]
Shabani, Hamid Reza [1 ]
Allahverdi, Mahmoud [1 ]
Vaezzadeh, Seyed Ali [1 ]
机构
[1] Univ Tehran Med Sci, Dept Med Phys & Biomed Engn, Tehran, Iran
[2] Imam Khomeini Hosp, Inst Canc, Dept Radiotherapy Oncol, Tehran, Iran
[3] Amirkabir Univ Technol, Dept Nucl Engn & Phys, Tehran, Iran
关键词
Breast cancer; brachytherapy; heterogeneity; MCNPX; treatment planning system; DOSIMETRIC PARAMETERS; PROSTATE-CANCER; INHOMOGENEITIES; POSITION; MODEL;
D O I
10.4103/0973-1482.103519
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Aims: This study investigates to what extent the dose received by lungs from a commercially available treatment planning system, Ir-192 high-dose-rate (HDR), in breast brachytherapy, is accurate, with the emphasis on tissue heterogeneities, and taking into account the presence of ribs, in dose delivery to the lung. Materials and Methods: A computed tomography (CT) scan of a breast was acquired and transferred to the 3-D treatment planning system and was also used to construct a patient-equivalent phantom. An implant involving 13 plastic catheters and 383 programmed source dwell positions were simulated, using the Monte Carlo N-Particle eXtended (MCNPX) code. The Monte Carlo calculations were compared with the corresponding commercial treatment planning system (TPS) in the form of percentage isodose and cumulative dose-volume histogram (DVH) in the breast, lungs, and ribs. Results: The comparison of the Monte Carlo results and the TPS calculations showed that a percentage of isodose greater than 75 in the breast, which was located rather close to the implant or away from the breast curvature surface and lung boundary, were in good agreement. TPS calculations overestimated the dose to the lung for lower isodose contours that were lying near the breast surface and the boundary of breast and lung and were relatively away from the implant. Conclusions: Taking into account the ribs and entering the actual data for breasts, ribs, and lungs, revealed an average overestimation of the dose by a factor of 8 in the lung for TPS calculations. Therefore, the accuracy of the TPS results may be limited to regions near the implants where the treatment is planned, and is a more conservative approach for regions at boundaries with curvatures or tissues with a different material than that in the breast.
引用
收藏
页码:394 / 398
页数:5
相关论文
共 50 条
  • [21] Dosimetric characteristics of the 192Ir high-dose-rate afterloading brachytherapy source
    Sadeghi, Mahdi
    Taghdiri, Fatemah
    Saidi, Pooneh
    JAPANESE JOURNAL OF RADIOLOGY, 2011, 29 (05) : 324 - 329
  • [22] Dosimetric characteristics of the 192Ir high-dose-rate afterloading brachytherapy source
    Mahdi Sadeghi
    Fatemah Taghdiri
    Pooneh Saidi
    Japanese Journal of Radiology, 2011, 29 : 324 - 329
  • [23] Monte Carlo dosimetry of a new 192Ir pulsed dose rate brachytherapy source
    Karaiskos, P
    Angelopoulos, A
    Pantelis, E
    Papagiannis, P
    Sakelliou, L
    Kouwenhoven, E
    Baltas, D
    MEDICAL PHYSICS, 2003, 30 (01) : 9 - 16
  • [24] A Monte Carlo investigation of the dosimetric characteristics of the VariSource 192Ir high dose rate brachytherapy source
    Karaiskos, P
    Angelopoulos, A
    Baras, P
    Sakelliou, L
    Sandilos, P
    Dardoufas, K
    Vlachos, L
    MEDICAL PHYSICS, 1999, 26 (08) : 1498 - 1502
  • [25] HIGH-DOSE RATE (HDR) IR-192 BRACHYTHERAPY APPLICATION OF BASAL CELL CARCINOMA
    Basaric, Borko
    Petrovic, Borislava
    Teodorovic, Milan
    Baucal, Milutin
    Rutonjski, Laza
    Cudic, Ozren
    Djuran, Branislav
    Mitric-Askovic, Milana
    RAD 2015: THE THIRD INTERNATIONAL CONFERENCE ON RADIATION AND APPLICATIONS IN VARIOUS FIELDS OF RESEARCH, 2015, : 65 - 68
  • [26] Interstitial brachytherapy for recurrent breast cancer using a high dose rate Ir-192 remote afterloading system: A report of two cases
    Yoden E.
    Hiratsuka J.
    Imajo Y.
    Yamamoto S.
    Kurebayashi J.
    Shimozurna K.
    Sonoo H.
    Breast Cancer, 2000, 7 (3) : 252 - 255
  • [27] Deep reinforcement learning for treatment planning in high-dose-rate cervical brachytherapy
    Pu, Gang
    Jiang, Shan
    Yang, Zhiyong
    Hu, Yuanjing
    Liu, Ziqi
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2022, 94 : 1 - 7
  • [28] Characterization of microMOSFET detectors for in vivo dosimetry in high-dose-rate brachytherapy with 192Ir
    Ruiz-Arrebola, Samuel
    Fabregat-Borras, Rosa
    Rodriguez, Eduardo
    Fernandez-Montes, Manuel
    Perez-Macho, Mercedes
    Ferri, Maria
    Garcia, Ana
    Cardenal, Juan
    Pacheco, Maria T.
    Anchuelo, Javier
    Tornero-Lopez, Ana M.
    Prada, Pedro J.
    Guirado, Damian
    MEDICAL PHYSICS, 2020, 47 (05) : 2242 - 2253
  • [29] Tumor Control and Normal Tissue Complications in High-dose-rate Brachytherapy for Cervical Cancer Patients Using Ir-192 Radioactive Source
    Kaur, Gurpreet
    Garg, Pardeep
    Dangwal, Vinod Kumar
    Singh, Baltej
    Gaur, Garima
    Grover, Romikant
    Singh, Simrandeep
    Sharma, Rachana
    JOURNAL OF MEDICAL PHYSICS, 2024, 49 (03) : 363 - 369
  • [30] THUBrachy: fast Monte Carlo dose calculation tool accelerated by heterogeneous hardware for high-dose-rate brachytherapy
    Hu, An-Kang
    Qiu, Rui
    Liu, Huan
    Wu, Zhen
    Li, Chun-Yan
    Zhang, Hui
    Li, Jun-Li
    Yang, Rui-Jie
    NUCLEAR SCIENCE AND TECHNIQUES, 2021, 32 (03)