Evaluating Component Migration: Comparing Two Generations of the INBONE® Total Ankle Replacement

被引:3
|
作者
Brigido, Stephen A. [1 ]
Wobst, Garrett M. [2 ]
Galli, Melissa M. [3 ]
Protzman, Nicole M. [4 ]
机构
[1] Coordinated Hlth, Foot & Ankle Reconstruct, Foot & Ankle Dept, Bethlehem, PA USA
[2] Avera Orthoped Surg Specialists, Aberdeen, SD USA
[3] CORE Inst, Phoenix, AZ USA
[4] Coordinated Hlth, Clin Educ & Res Dept, Allentown, PA USA
关键词
ankle joint surgery; joint prosthesis; osteoarthritis; subsidence; total ankle replacement; FOLLOW-UP; PERIOPERATIVE COMPLICATIONS; HINTEGRA ANKLE; SHORT-TERM; ARTHROPLASTY;
D O I
10.1053/j.jfas.2015.03.004
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Although total ankle replacement (TAR) designs have radically evolved, the compressive forces at the ankle can cause aseptic loosening, talar subsidence, and implant failure. The purpose of the present report was to compare the implant migration associated with the INBONE (R) I, a TAR system with a stemmed talar component, and the newer generation INBONE (R) II, a TAR system without a stemmed talar component (Wright Medical Technology, Inc., Arlington, TN). Because core decompression could weaken the integrity of the talus, we hypothesized that the stemmed component would result in greater implant migration. A total of 35 consecutive patients (age 58.2 +/- 12.1 years; 23 men) were included. Of these 35 patients, 20 (57.1%) had been treated with the INBONE (R) I and 15 (42.9%) with the INBONE (R) II. To assess implant migration, using anteroposterior radiographs, the distance from the apex of the tibial component to the most distal aspect of the talar stem or to the mid-saddle of the nonstemmed component was measured. The measurements were recorded from the immediate postoperative radiographs and the 12-month postoperative radiographs. Implant migration was quantified as the difference between the 12-month and the immediate postoperative measurements. Despite our hypothesis, no significant difference was found in implant migration between the INBONE (R) I (0.7 +/- 1.2 mm) and INBONE (R) II (0.6 +/- 1.3 mm, p = .981). However, previously published data have suggested that implant migration can continue for >= 2 years after surgery. Therefore, additional investigations with larger sample sizes and longer follow-up periods are needed to draw definitive conclusions. (C) 2015 by the American College of Foot and Ankle Surgeons. All rights reserved.
引用
收藏
页码:892 / 895
页数:4
相关论文
共 50 条
  • [41] Statistical shape modeling of shape variability of the human distal tibia: implication for implant design of the tibial component for total ankle replacement
    Yu, Jian
    Li, Chengke
    Lyu, Jinyang
    Cao, Shengxuan
    Zhang, Chao
    Ma, Xin
    Zhao, Dahang
    FRONTIERS IN BIOENGINEERING AND BIOTECHNOLOGY, 2025, 13
  • [42] Preoperative and Postoperative Range of Motion: A Retrospective Comparison of Two Total Ankle Replacement Systems
    Brigido, Stephen A.
    Mulhern, Jennifer L.
    Wobst, Garrett M.
    Protzman, Nicole M.
    JOURNAL OF FOOT & ANKLE SURGERY, 2015, 54 (05) : 809 - 814
  • [43] Total Ankle Replacement Using HINTEGRA, an Unconstrained, Three-Component System Surgical Technique and Pitfalls
    Barg, Alexej
    Knupp, Markus
    Henninger, Heath B.
    Zwicky, Lukas
    Hintermann, Beat
    FOOT AND ANKLE CLINICS, 2012, 17 (04) : 607 - +
  • [44] Early Failure in Total Ankle Replacements Due to Component Malposition: A Report of Two Cases
    Goldberg, Andrew J.
    Sharp, Bob
    Cooke, Paul
    FOOT & ANKLE INTERNATIONAL, 2009, 30 (08) : 783 - 787
  • [45] The Effect of Three-Component Total Ankle Replacement Malalignment on Clinical Outcome: Pain Relief and Functional Outcome in 317 Consecutive Patients
    Barg, Alexej
    Elsner, Andreas
    Anderson, Andrew E.
    Hintermann, Beat
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2011, 93A (21) : 1969 - 1978
  • [46] Influence of Tibial Component Design Features and Interference Fit on Implant-Bone Micromotion in Total Ankle Replacement: A Finite Element Study
    Clarke, Gabriel
    Johnson, Joshua E.
    Netto, Cesar de Cesar
    Anderson, Donald D.
    FOOT & ANKLE INTERNATIONAL, 2024, 45 (12) : 1414 - 1421
  • [47] The Effects of Implant Orientations and Implant-Bone Interfacial Conditions on Potential Causes of Failure of Tibial Component Due to Total Ankle Replacement
    Mondal, Subrata
    Ghosh, Rajesh
    JOURNAL OF MEDICAL AND BIOLOGICAL ENGINEERING, 2019, 39 (04) : 541 - 551
  • [48] Polka dot cementless talar component in enhancing total ankle replacement fixation: A parametric study using the finite element analysis approach
    Moideen, Irwan Shah Mohd
    Lim, Chin Tat
    Yeow, Raye C. H.
    Chong, Desmond Y. R.
    COMPUTERS IN BIOLOGY AND MEDICINE, 2022, 141
  • [49] Ankle Replacement or Ankle Fusion: Who Reigns Supreme? Commentary on an article by Marisa R. Benich, BS, et al.: "Comparison of Treatment Outcomes of Arthrodesis and Two Generations of Ankle Replacement Implants"
    DiGiovanni, Christopher W.
    Guss, Daniel
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2017, 99 (21)
  • [50] In vivo kinematics of three-component mobile-bearing total ankle replacement in rheumatoid ankle with talocalcaneal arthrodesis and spontaneous talocalcaneal fusion
    Iwamoto, Keiji
    Shi, Kenrin
    Tomita, Tetsuya
    Hashimoto, Jun
    Yamazaki, Takaharu
    Yoshikawa, Hideki
    Sugamoto, Kazuomi
    MODERN RHEUMATOLOGY, 2014, 24 (06) : 897 - 903