Improvement of registration accuracy for navigatedcontrol drill in mastoidectomy (Autopilot)

被引:7
作者
Strauss, G. [1 ,3 ]
Dittrich, E. [3 ]
Baumberger, C.
Hofer, M. [3 ]
Strauss, M. [2 ]
Stopp, S. [2 ]
Koulechov, K. [2 ]
Dietz, A. [3 ]
Lueth, T. [2 ]
机构
[1] Univ Leipzig, ICCAS, Klin & Poliklin HNO Heilkunde Plast Operat, D-04103 Leipzig, Germany
[2] Tech Univ Munich, Lehrstuhl Mikro & Med Geratetechn, D-8000 Munich, Germany
[3] Univ Leipzig, BMBF Innovat Ctr Comp Assisted Surg ICCAS, Leipzig, Germany
关键词
navigated control; mastoidectomy; drill; facial nerve; computer-assisted surgery; autopilot;
D O I
10.1055/s-2007-995629
中图分类号
R76 [耳鼻咽喉科学];
学科分类号
100213 ;
摘要
Background: The goal of this study is the improvement of the surgical accuracy of a navigate-controlled drill for mastoidectomy in a lab test. Methods: For lab tests an artificial model of the temporal bone with color-coded injury identification of the facial nerve (solution of 0.5 mm) was used. Two different registration methods were examined: (group 1) navigation bow with 4 integrated markers at the upper jaw; (group 2) landmark registration with 4 titanium micro screws. An optical navigation system was used. The targets were illustrated by 3 titanium screws within the range of the planum mastoideum. The accuracy of the navigate-controlled drill in drilling the planned cavity were evaluated at 20 temporal bone models. The measurement of the registration accuracy was evaluated by deviation between the target screw and the calculated position in the navigation system. The evaluation of the resulted cavities was done by 5 senior surgeons with the help of the microscope. Results: The registration accuracy shows a maximum deviation between the real position and the calculated position of 1, 73 mm in group of 1 and 0.93 mm in group 2. In group 1 the nerve was hurt in 5/20 cases and a maximum deviation of - 1.5 mm (Std 0.25 mm) (drilled beyond the nerve) was measured. In group 2 the nerve was not hurt, a maximum deviation of 0.5 mm (too early stopped before the nerve) was measured. Conclusions: Significantly better results of the registration and drilling accuracy show up in group group 2. Thus the preconditions for clinical use are fulfilled.
引用
收藏
页码:560 / 564
页数:5
相关论文
共 14 条
  • [1] Caversaccio Marco, 2002, Acta Otorhinolaryngol Belg, V56, P51
  • [2] GRUNERT R, 2006, ELEPHANT ANATOMICAL, P312
  • [3] Computer-assisted paranasal sinus surgery using an optoelectronic digitizer system
    Hauser, R
    Westermann, B
    Reinhardt, H
    Probst, R
    [J]. LARYNGO-RHINO-OTOLOGIE, 1996, 75 (04) : 199 - 207
  • [4] Hofer M, 2007, STUD HEALTH TECHNOL, V125, P179
  • [5] Hofer M, 2006, STUD HEALTH TECHNOL, V119, P201
  • [6] Computer-Assisted-Surgery in otorhinolaryngology.
    Klimek, L
    Mosges, R
    [J]. LARYNGO-RHINO-OTOLOGIE, 1998, 77 (05) : 275 - 282
  • [7] Maassen MM, 2006, HNO, V54, P105, DOI 10.1007/s00106-005-1283-5
  • [8] Marmulla R, 2005, HNO, V53, P148, DOI 10.1007/s00106-004-1101-5
  • [9] Schmerber S, 2001, Comput Aided Surg, V6, P1
  • [10] Strauss G, 2006, LARYNGO RHINO OTOL, V85, P559, DOI 10.1055/s-2005-921222