Performance of a multi leaf collimator system for MR-guided radiation therapy

被引:21
作者
Cai, Bin [1 ]
Li, Harold [1 ]
Yang, Deshan [1 ]
Rodriguez, Vivian [1 ]
Curcuru, Austen [1 ]
Wang, Yuhe [1 ]
Wen, Jie [2 ]
Kashani, Rojano [1 ]
Mutic, Sasa [1 ]
Green, Olga [1 ]
机构
[1] Washington Univ, Dept Radiat Oncol, St Louis, MO 63110 USA
[2] Washington Univ, Med Ctr, Mallinckrodt Inst Radiol, St Louis, MO 63110 USA
关键词
MLC; MR-IGRT; tri-Co60; QUALITY-ASSURANCE; CLINICAL IMPLEMENTATION; DELIVERY; IMRT; DOSIMETRY; MLC;
D O I
10.1002/mp.12571
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The purpose of this study was to investigate and characterize the performance of a Multi Leaf Collimator (MLC) designed for Cobalt-60 based MR-guided radiation therapy system in a 0.35 T magnetic field. Methods: The MLC design and unique assembly features in the ViewRay MRIdian system were first reviewed. The RF cage shielding of MLC motor and cables were evaluated using ACR phantoms with real-time imaging and quantified by signal-to-noise ratio. The dosimetric characterizations, including the leaf transmission, leaf penumbra, tongue-and-groove effect, were investigated using radiosensitive films. The output factor of MLC-defined fields was measured with ionization chambers for both symmetric fields from 2.1 x 2.1 cm(2) to 27.3 x 27.3 cm(2) and asymmetric fields from 10.5 x 10.5 cm(2) to 10.5 x 2.0 cm(2). Multi leaf collimator (MLC) positional accuracy was assessed by delivering either a picket fence (PF) style pattern on radiochromic films with wire-jig phantom or double and triple-rectangular patterns on ArcCheck-MR (Sun Nuclear, Melbourne, FL, USA) with gamma analysis as the pass/fail indicator. Leaf speed tests were performed to assess the capability of full range leaf travel within manufacture's specifications. Multi leaf collimator plan delivery reproducibility was tested by repeatedly delivering both open fields and fields with irregular shaped segments over 1-month period. Results: Comparable SNRs within 4% were observed for MLC moving and stationary plans on vendor-reconstructed images, and the direct k-space reconstructed images showed that the three SNRs are within 1%. The maximum leaf transmission for all three MLCs was less than 0.35% and the average leakage was 0.153 +/- 0.006%, 0.151 +/- 0.008%, and 0.159 +/- 0.015% for head 1, 2, and 3, respectively. Both the leaf edge and leaf end penumbra showed comparable values within 0.05 cm, and the measured values are within 0.1 cm with TPS values. The leaf edge TG effect indicated 10% underdose and the leaf end TG showed a shifted dose distribution with 0.3 cm offset. The leaf positioning test showed a 0.2 cm accuracy in the PF style test, and a gamma passing rate above 96% was observed with a 3%/2 mm criteria when comparing the measured double/triple-rectangular pattern fluence with TPS calculated fluence. The average leaf speed when executing the test plan fell in a range from 1.86 to 1.95 cm/s. The measured and TPS calculated output factors were within 2% for squared fields and within 3% for rectangular fields. The reproducibility test showed the deviation of output factors were well within 2% for square fields and the gamma passing rate within 1.5% for fields with irregular segments. The Monte Carlo predicted output factors were within 2% compared to TPS values. 15 out of the 16 IMRT plans have gamma passing rate more than 98% compared to the TPS fluence with an average passing rate of 99.1 +/- 0.6%. Conclusion: The MRIdian MLC has a good RF noise shielding design, low radiation leakage, good positioning accuracy, comparable TG effect, and can be modeled by an independent Monte Carlo calculation platform. (C) 2017 American Association of Physicists in Medicine
引用
收藏
页码:6504 / 6514
页数:11
相关论文
共 35 条
[1]   Online Magnetic Resonance Image Guided Adaptive Radiation Therapy: First Clinical Applications [J].
Acharya, Sahaja ;
Fischer-Valuck, Benjamin W. ;
Kashani, Rojano ;
Parikh, Parag ;
Yang, Deshan ;
Zhao, Tianyu ;
Green, Olga ;
Wooten, Omar ;
Li, H. Harold ;
Hu, Yanle ;
Rodriguez, Vivian ;
Olsen, Lindsey ;
Robinson, Clifford ;
Michalski, Jeff ;
Mutic, Sasa ;
Olsen, Jeffrey .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2016, 94 (02) :394-403
[2]  
Boyer A, 2001, 50 RAD THER COMM
[3]  
BRAHME A, 1988, STRAHLENTHER ONKOL, V164, P343
[4]   Guidance document on delivery, treatment planning, and clinical implementation of IMRT: Report of the IMRT subcommittee of the AAPM radiation therapy committee [J].
Ezzell, GA ;
Galvin, JM ;
Low, D ;
Palta, JR ;
Rosen, I ;
Sharpe, MB ;
Xia, P ;
Xiao, Y ;
Xing, L ;
Yu, CX .
MEDICAL PHYSICS, 2003, 30 (08) :2089-2115
[5]   IMRT commissioning: Multiple institution planning and dosimetry comparisons, a report from AAPM Task Group 119 [J].
Ezzell, Gary A. ;
Burmeister, Jay W. ;
Dogan, Nesrin ;
LoSasso, Thomas J. ;
Mechalakos, James G. ;
Mihailidis, Dimitris ;
Molineu, Andrea ;
Palta, Jatinder R. ;
Ramsey, Chester R. ;
Salter, Bill J. ;
Shi, Jie ;
Xia, Ping ;
Yue, Ning J. ;
Xiao, Ying .
MEDICAL PHYSICS, 2009, 36 (11) :5359-5373
[6]   First MR images obtained during megavoltage photon irradiation from a prototype integrated linac-MR system [J].
Fallone, B. G. ;
Murray, B. ;
Rathee, S. ;
Stanescu, T. ;
Steciw, S. ;
Vidakovic, S. ;
Blosser, E. ;
Tymofichuk, D. .
MEDICAL PHYSICS, 2009, 36 (06) :2084-2088
[7]   Dosimetric evaluation of the conformation of the multileaf collimator to irregularly shaped fields [J].
Frazier, A ;
Du, M ;
Wong, J ;
Vicini, F ;
Taylor, R ;
Yu, C ;
Matter, R ;
Martinez, A ;
Yan, D .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1995, 33 (05) :1229-1238
[8]  
Glasgow G., 1999, MODERN TECHNOLOGY RA
[9]   Commissioning and Quality Assurance of the First Commercial Hybrid MRI-IMRT System [J].
Green, O. Pechenaya ;
Goddu, S. ;
Mutic, S. .
MEDICAL PHYSICS, 2012, 39 (06) :3785-3785
[10]   Generalized Autocalibrating Partially Parallel Acquisitions (GRAPPA) [J].
Griswold, MA ;
Jakob, PM ;
Heidemann, RM ;
Nittka, M ;
Jellus, V ;
Wang, JM ;
Kiefer, B ;
Haase, A .
MAGNETIC RESONANCE IN MEDICINE, 2002, 47 (06) :1202-1210