Fiber-Coupled, Time-Gated Al2O3 : C Radioluminescence Dosimetry Technique and Algorithm for Radiation Therapy With LINACs

被引:5
作者
Magne, Sylvain [1 ]
Deloule, Sybelle [2 ]
Ostrowsky, Aime [2 ]
Ferdinand, Pierre [1 ]
机构
[1] CEA, LIST, Lab Mesures Opt, F-91191 Gif Sur Yvette, France
[2] CEA, LIST, Lab Natl Henri Becquerel, F-91191 Gif Sur Yvette, France
关键词
Alumina; in vivo dosimetry; optical fiber sensor; optically stimulated luminescence; radiation therapy; radioluminescence; time discrimination; OPTICALLY STIMULATED LUMINESCENCE; MULTICHANNEL DOSIMETER; SENSORS; SIGNAL;
D O I
10.1109/TNS.2013.2263640
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
An original algorithm for real-time In Vivo Dosimetry (IVD) based on Radioluminescence (RL) of dosimetric-grade Al2O3 : C crystals is described and demonstrated in reference conditions with 12-MV photon beams from a Saturne 43 linear accelerator (LINAC), simulating External Beam Radiation Therapy (EBRT) treatments. During the course of irradiation, a portion of electrons is trapped within the Al2O3 : C crystal while another portion recombines and generates RL, recorded on-line using an optical fiber. The RL sensitivity is dose-dependent and increases in accordance with the concentration of trapped electrons. Once irradiation is completed, the Al2O3 : C crystal is reset by laser light (reusable) and the resultant OSL (Optically Stimulated Luminescence) is also collected back by the remote RL-OSL reader and finally integrated to yield the absorbed dose. During irradiation, scintillation and Cerenkov lights generated within the optical fiber ("stem effect") are removed by a time-discrimination method involving a discriminating unit and a fiber-coupled BGO scintillator placed in the irradiation room, next to the LINAC. The RL signals were then calibrated with respect to reference dose and dose rate data using an ionization chamber (IC). The algorithm relies upon the integral of the RL and provides the accumulated dose (useful to the medical physicist) at any time during irradiation, the dose rate being derived afterwards. It is tested with both step and arbitrary dose rate profiles, manually operated from the LINAC control desk. The doses measured by RL and OSL are both compared to reference doses and deviations are about +/- 2% and +/- 1% respectively, thus demonstrating the reliability of the algorithm for arbitrary profiles and wide range of dose rates. Although the calculation was done off-line, it is amenable to real-time processing during irradiation.
引用
收藏
页码:2998 / 3007
页数:10
相关论文
共 20 条
[1]   An algorithm for real-time dosimetry in intensity-modulated radiation therapy using the radioluminescence signal from Al2O3:C [J].
Andersen, C. E. ;
Marckmann, C. J. ;
Aznar, M. C. ;
Botter-Jensen, L. ;
Kjaer-Kristoffersen, F. ;
Medin, J. .
RADIATION PROTECTION DOSIMETRY, 2006, 120 (1-4) :7-13
[2]   Real-time optical-fibre luminescence dosimetry for radiotherapy:: physical characteristics and applications in photon beams [J].
Aznar, MC ;
Andersen, CE ;
Botter-Jensen, L ;
Bäck, SÅJ ;
Mattsson, S ;
Kjær-Kristoffersen, F ;
Medin, J .
PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (09) :1655-1669
[3]   Water equivalent plastic scintillation detectors in radiation therapy [J].
Beddar, A. S. .
RADIATION PROTECTION DOSIMETRY, 2006, 120 (1-4) :1-6
[4]   A comparison of BCF-12 organic scintillators and Al2O3:C crystals for real-time medical dosimetry [J].
Beierholm, Anders R. ;
Andersen, Claus E. ;
Lindvold, Lars R. ;
Kjaer-Kristoffersen, Flemming ;
Medin, Joakim .
RADIATION MEASUREMENTS, 2008, 43 (2-6) :898-903
[5]   Improved real-time dosimetry using the radioluminescence signal from Al2O3:C [J].
Damkjaer, S. M. S. ;
Andersen, C. E. ;
Aznar, M. C. .
RADIATION MEASUREMENTS, 2008, 43 (2-6) :893-897
[6]   Memory effects and systematic errors in the RL signal from fiber coupled Al2O3:C for medical dosimetry [J].
Damkjaer, S. M. S. ;
Andersen, C. E. .
RADIATION MEASUREMENTS, 2010, 45 (3-6) :671-673
[7]   CW-OSL measurement protocols using optical fibre Al2O3:C dosemeters [J].
Edmund, J. M. ;
Andersen, C. E. ;
Marckmann, C. J. ;
Aznar, M. C. ;
Akselrod, M. S. ;
Botter-Jensen, L. .
RADIATION PROTECTION DOSIMETRY, 2006, 119 (1-4) :368-374
[8]   Spectral discrimination of Cerenkov radiation in scintillating dosimeters [J].
Frelin, AM ;
Fontbonne, JM ;
Ban, G ;
Colin, J ;
Labalme, M ;
Batalla, A ;
Isambert, A ;
Vela, A ;
Leroux, T .
MEDICAL PHYSICS, 2005, 32 (09) :3000-3006
[9]  
International Atomic Energy Agency (IAEA), 2000, Technical Reports Series, V398
[10]   Elimination of Cerenkov interference in a fibre-optic-coupled radiation dosemeter [J].
Justus, Brian L. ;
Falkenstein, Paul ;
Huston, Alan L. ;
Plazas, Maria C. ;
Ning, Holly ;
Miller, Robert W. .
RADIATION PROTECTION DOSIMETRY, 2006, 120 (1-4) :20-23