Pattern electroretinogram to detect glaucoma: comparing the PERGLA and the PERG Ratio protocols

被引:22
作者
Bach, Michael [1 ]
Ramharter-Sereinig, Anke [2 ]
机构
[1] Univ Eye Hosp, D-79106 Freiburg, Germany
[2] Univ Augenklin Innsbruck, Innsbruck, Austria
关键词
Pattern ERG; PERG; Glaucoma; Methodology; Skin electrode; DTL electrode; GANGLION-CELL FUNCTION; OCULAR HYPERTENSION; CONTRAST ADAPTATION; EVOKED-POTENTIALS; LONG-TERM; FIBER; SKIN; ERG; VARIABILITY; PROGRESSION;
D O I
10.1007/s10633-013-9412-z
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
To detect early glaucoma, Porciatti and Ventura suggested in 2004 the pattern electroretinogram (PERG) protocol "PERGLA" with the following features: (1) skin electrodes, (2) steady-state reversal (15 rps) of horizontal 1.6 cpd gratings and (3) Fourier-based analysis. We compared this to our "PERG Ratio" protocol which uses (1) corneal (DTL) electrodes, (2) nearly the same reversal rate, but 2 check sizes of 0.8A degrees and 16A degrees and (3) Fourier-based analysis using the ratio of amplitudes to the 2 check sizes. We examined 16 eyes with glaucoma (age 64 +/- A 7 years) and a control group (n = 13, age 59 +/- A 8 years). Responses to all stimuli were simultaneously recorded with both electrode types using PERGLA-type gratings and checkerboards as necessary for the PERG Ratio. The median intra-individual test-retest coefficient of variation in normals pooled across stimuli for skin was 10.0 (95 % CI 1-85 %) and for DTL 9.95 (0.5-49 %), and they are thus nearly identical. The amplitudes obtained from skin were on average 30 % of those with DTL; the skin's signal-to-noise ratio (SNR) was 66 % of DTL electrodes. Glaucoma detection was assessed via receiver operating characteristics (ROCs). Using skin electrodes, ROC area-under-curve (AUC) was 72/76/72 % using gratings (PERGLA), checks or the PERG Ratio, respectively. Using DTL electrodes, the ROC areas were 60/67/77 %; the last value corresponds to the PERG Ratio protocol. Our results suggest that skin electrodes are a valid alternative to corneal electrodes: their advantages being no direct eye contact and smaller normal amplitude range compared to DTL; disadvantages: amplitude reduced to 30 % of DTL and SNR ratio reduced to 66 % of DTL. Efficacy in detecting glaucoma was a little higher with the PERG Ratio protocol (ROC AUC: 77 %, PERGLA: 72 %), but not significantly so.
引用
收藏
页码:227 / 238
页数:12
相关论文
共 71 条
[1]  
[Anonymous], 2000, JAMA, V284, P3043
[2]  
[Anonymous], 2006, R LANG ENV STAT COMP
[3]   A 3-YEAR FOLLOW-UP-STUDY OF OCULAR HYPERTENSION BY PATTERN ELECTRORETINOGRAM [J].
ARAI, M ;
YOSHIMURA, N ;
SAKAUE, H ;
CHIHARA, E ;
HONDA, Y .
OPHTHALMOLOGICA, 1993, 207 (04) :187-195
[4]  
Aulhorn E, 1977, DOC OPHTHALMOL P SER, P75
[5]   Raster-scan cathode-ray tubes for vision research - Limits of resolution in space, time and intensity, and some solutions [J].
Bach, M ;
Meigen, T ;
Strasburger, H .
SPATIAL VISION, 1997, 10 (04) :403-414
[6]   Different effect of dioptric defocus vs. light scatter on the Pattern Electroretinogram (PERG) [J].
Bach, M ;
Mathieu, M .
DOCUMENTA OPHTHALMOLOGICA, 2004, 108 (01) :99-106
[7]   Electrophysiological approaches for early detection of glaucoma [J].
Bach, M .
EUROPEAN JOURNAL OF OPHTHALMOLOGY, 2001, 11 :S41-S49
[8]   Check size tuning of the pattern electroretingoram: A reappraisal [J].
Bach, M ;
Holder, GE .
DOCUMENTA OPHTHALMOLOGICA, 1996, 92 (03) :193-202
[9]   PATTERN ELECTRORETINOGRAM IN GLAUCOMA AND OCULAR HYPERTENSION [J].
BACH, M ;
SPEIDELFIAUX, A .
DOCUMENTA OPHTHALMOLOGICA, 1989, 73 (02) :173-181
[10]   Do's and don'ts in Fourier analysis of steady-state potentials [J].
Bach, M ;
Meigen, T .
DOCUMENTA OPHTHALMOLOGICA, 1999, 99 (01) :69-82