Selective enzyme immobilization on arrayed microelectrodes for the application of sensing neurotransmitters

被引:20
作者
Tseng, Tina T-C [1 ]
Yao, Judy [2 ]
Chan, Wen-Chin [1 ]
机构
[1] Natl Taiwan Univ Sci & Technol, Dept Chem Engn, Taipei 10607, Taiwan
[2] Univ Calif Los Angeles, Dept Chem & Biomol Engn, Los Angeles, CA 90095 USA
关键词
Biosensors; Immobilized enzymes; Glutamate; Microelectrode array; Chitosan; Biomedical; Bioprocess monitoring; L-GLUTAMATE OXIDASE; CENTRAL-NERVOUS-SYSTEM; EXTRACELLULAR GLUTAMATE; MULTISITE MICROELECTRODE; IN-VIVO; BIOSENSOR; BRAIN; MICROSENSORS; DOPAMINE; ELECTROCHEMISTRY;
D O I
10.1016/j.bej.2013.04.019
中图分类号
Q81 [生物工程学(生物技术)]; Q93 [微生物学];
学科分类号
071005 ; 0836 ; 090102 ; 100705 ;
摘要
The method for selective enzyme immobilization on arrayed microelectrodes has been investigated in order to fabricate a near-real time biosensor for potential ex vivo or in vivo applications (e.g. in fermentation processes, in neuroscience studies, etc.). In this study, chitosan is used as the enzyme immobilization material, permselective polymers (polypyrrole and Nafion (R)) are used as the exclusion layers for charged interferents, and glutamate oxidase is used as the model enzyme for immobilization. To fabricate a glutamate sensor in the array, glutamate oxidase is selectively immobilized on a microelectrode site which is closely arrayed to other microelectrodes (platinum microelectrodes with similar to 100 mu m separation) with the electrode dimension of similar to 30 mu m x 140 mu m by adsorbing the enzyme on the electrodeposited chitosan matrix on the electrode surface. When testing the glutamate sensor with glutamate at physiological concentrations, it has a linear detection range up to 217 mu m, a response time similar to 1 s, a limit of detection 2.5 +/- 1.2 mu m, and a sensitivity 38.1 +/- 5.4 nA/mu m cm(2) (n = 8); on the other hand, the closely arrayed control sensor without the chitosan film showed no signal upon the addition of glutamate. Little interferent current was observed. The sensor can retain up to 70% of its initial sensitivity for at least 9 days and no observable decrease in sensor sensitivity was detected after 20 times of continuous operations. Successful selective glutamate oxidase immobilization on closely packed microelectrodes for monitoring glutamate is demonstrated. (C) 2013 Elsevier B.V. All rights reserved.
引用
收藏
页码:146 / 153
页数:8
相关论文
共 32 条
  • [1] A MEDIATED AMPEROMETRIC ENZYME ELECTRODE USING TETRATHIAFULVALENE AND L-GLUTAMATE OXIDASE FOR THE DETERMINATION OF L-GLUTAMIC ACID
    ALMEIDA, NF
    MULCHANDANI, AK
    [J]. ANALYTICA CHIMICA ACTA, 1993, 282 (02) : 353 - 361
  • [2] ORGANIC ELECTROCHEMISTRY IN THE SOLID STATE-OVEROXIDATION OF POLYPYRROLE
    BECK, F
    BRAUN, P
    OBERST, M
    [J]. BERICHTE DER BUNSEN-GESELLSCHAFT-PHYSICAL CHEMISTRY CHEMICAL PHYSICS, 1987, 91 (09): : 967 - 974
  • [3] Improved ceramic-based multisite microelectrode for rapid measurements of L-glutamate in the CNS
    Burmeister, JJ
    Pomerleau, F
    Palmer, M
    Day, BK
    Huettl, P
    Gerhardt, GA
    [J]. JOURNAL OF NEUROSCIENCE METHODS, 2002, 119 (02) : 163 - 171
  • [4] CARLSSON A, 1959, PHARMACOL REV, V11, P490
  • [5] AAPS-FDA workshop white paper:: Microdialysis principles, application and regulatory perspectives
    Chaurasia, Chandra S.
    Mueller, Markus
    Bashaw, Edward D.
    Benfeldt, Eva
    Bolinder, Jan
    Bullock, Ross
    Bungay, Peter M.
    DeLange, Elizabeth C. M.
    Derendorf, Hartmut
    Elmquist, William F.
    Hammarlund-Udenaes, Margareta
    Joukhadar, Christian
    Kellogg, Dean L., Jr.
    Lunte, Craig E.
    Nordstrom, Carl Henrik
    Rollema, Hans
    Sawchuk, Ronald J.
    Cheung, Belinda W. Y.
    Shah, Vinod P.
    Stahle, Lars
    Ungerstedt, Urban
    Welty, Devin F.
    Yeo, Helen
    [J]. PHARMACEUTICAL RESEARCH, 2007, 24 (05) : 1014 - 1025
  • [6] Chefer Vladimir I, 2009, Curr Protoc Neurosci, VChapter 7, DOI 10.1002/0471142301.ns0701s47
  • [7] GLUTAMATE OXIDASE ENZYME ELECTRODES - MICROSENSORS FOR NEUROTRANSMITTER DETERMINATION USING ELECTROCHEMICALLY POLYMERIZED PERMSELECTIVE FILMS
    COOPER, JM
    FOREMAN, PL
    GLIDLE, A
    LING, TW
    PRITCHARD, DJ
    [J]. JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1995, 388 (1-2): : 143 - 149
  • [8] COOPER JM, 1994, J MATER SCI-MATER EL, V5, P111
  • [9] An electrochemical method for making enzyme microsensors. Application to the detection of dopamine and glutamate
    Cosnier, S
    Innocent, C
    Allien, L
    Poitry, S
    Tsacopoulos, M
    [J]. ANALYTICAL CHEMISTRY, 1997, 69 (05) : 968 - 971
  • [10] A very thin overoxidized polypyrrole membrane as coating for fast time response and selective H2O2 amperometric sensor
    Debiemme-Chouvy, Catherine
    [J]. BIOSENSORS & BIOELECTRONICS, 2010, 25 (11) : 2454 - 2457