Evaluation of Intraocular Reactivity to Metallic and Ethylene Oxide Contaminants of Medical Devices in a Rabbit Model

被引:12
作者
Calogero, Don [1 ]
Buchen, Shelley Y. [2 ]
Tarver, Michelle E. [1 ]
Hilmantel, Gene [1 ]
Lucas, Anne D. [3 ]
Eydelman, Malvina B. [1 ]
机构
[1] US FDA, Ctr Devices & Radiol Hlth, Off Device Evaluat, Div Ophthalm Neurol & Ear Nose & Throat Devices, Silver Spring, MD USA
[2] US FDA, ORISE, Silver Spring, MD USA
[3] US FDA, Ctr Devices & Radiol Hlth, Off Sci & Engn Labs, Div Biol, Silver Spring, MD USA
关键词
ANTERIOR SEGMENT SYNDROME; PLASMA GAS; SURGERY;
D O I
10.1016/j.ophtha.2012.04.009
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Objective: To evaluate the intraocular reactivity to metallic and ethylene oxide (EO) contaminants of ophthalmic devices in rabbits. Design: Two experimental animal studies. Participants: Thirty-five New Zealand white rabbits. Methods: A metallic exposure study and an EO exposure study were performed. In the first study, both eyes of 25 rabbits were equally allocated to intracameral injections of alumina 0.2 mu g, alumina 20 mu g, copper sulfate 0.4 mu g, copper sulfate 20 mu g, or an aqueous control. In the second study, 10 rabbits were allocated (5 per group) to receive intracamerally an ophthalmic viscosurgical device (OVD) exposed to EO or not exposed to EO (control). All eyes were examined by slit lamp at baseline and 3, 6, 9, 24, 48, and 72 hours after exposure, with dilated indirect ophthalmoscopy being performed at 24 and 72 hours. Tonometry was performed only in the first study. Main Outcome Measures: Grade of corneal clouding, anterior chamber (AC) flare, AC cells, AC fibrin, iridal hyperemia, cell and fibrin on the lens surface, vitreous haze and cells, lens opacities, intraocular pressure, and onset time. Results: For metallic compounds at the study's low doses, mean inflammatory grades were 0.2 or less above the control for all responses at all time points. For the high-dose alumina, mean inflammatory grades peaked at 6 to 9 hours at 0.5 to 0.7 above the control responses for conjunctival congestion, iris hyperemia, AC cells, flare, and fibrin and declined over the remaining time points. For the high-dose copper sulfate, mean inflammatory grades peaked between 3 and 24 hours at 1.2 to 1.8 above the control responses for conjunctival congestion, iris hyperemia, AC cells, flare, fibrin, and corneal clouding, then subsequently declined. The intraocular pressure changes appeared significant for only high-dose copper sulfate, with mean declines of 4.3 to 7.5 mmHg at 6 to 72 hours. No clinically meaningful differences in ocular inflammation were observed between the OVD exposed to EO and the OVD not exposed to EO. Conclusions: Alumina and copper sulfate did not cause clinically meaningful ocular inflammation at the low study levels (levels expected with ophthalmic devices). Ethylene oxide exposure of an OVD was not associated with inflammation. Financial Disclosure(s): The authors have no proprietary or commercial interest in any materials discussed in this article. Ophthalmology 2012;xx:xxx (C) 2012 by the American Academy of Ophthalmology.
引用
收藏
页码:e36 / e42
页数:7
相关论文
共 18 条
  • [1] AITKENHEAD P, 1988, J ASSOC OFF ANA CHEM, V71, P729
  • [2] [Anonymous], 1995, 109937 ANSIAAMIISO
  • [3] [Anonymous], 2011, EUR PHARM
  • [4] Choi Jin Seok, 2008, Korean J Ophthalmol, V22, P220, DOI 10.3341/kjo.2008.22.4.220
  • [5] Duffy RE, 2000, ARCH OPHTHALMOL-CHIC, V118, P1167
  • [6] INTRAOCULAR SAFETY EVALUATION OF ETHYLENE-OXIDE AND STERILANT RESIDUES
    EDELHAUSER, HF
    ANTOINE, ME
    PEDERSON, HJ
    HIDDEMAN, JW
    HARRIS, RG
    [J]. JOURNAL OF TOXICOLOGY-CUTANEOUS AND OCULAR TOXICOLOGY, 1983, 2 (01): : 7 - 39
  • [7] Outbreak of toxic anterior segment syndrome following cataract surgery associated with impurities in autoclave steam moisture
    Hellinger, Walter C.
    Hasan, Saiyid A.
    Bacalis, Laura P.
    Thornblom, Deborah M.
    Beckmann, Susan C.
    Blackmore, Carina
    Forster, Terri S.
    Tirey, Jason F.
    Ross, Mary J.
    Nilson, Christian D.
    Mamalis, Nick
    Crook, Julia E.
    Bendel, Rick E.
    Shetty, Rajesh
    Stewart, Michael W.
    Bolling, James P.
    Edelhauser, Henry F.
    [J]. INFECTION CONTROL AND HOSPITAL EPIDEMIOLOGY, 2006, 27 (03) : 294 - 298
  • [8] Standardization of uveitis nomenclature for reporting clinical data. Results of the First International Workshop
    Jabs, DA
    Nussenblatt, RB
    Rosenbaum, JT
    [J]. AMERICAN JOURNAL OF OPHTHALMOLOGY, 2005, 140 (03) : 509 - 516
  • [9] Postoperative sterile endophthalmitis (TASS) associated with the MemoryLens
    Jehan, FS
    Mamalis, N
    Spencer, TS
    Fry, LL
    Kerstine, RS
    Olson, RJ
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2000, 26 (12) : 1773 - 1777
  • [10] Residual ethylene oxide in medical devices and device material
    Lucas, AD
    Merritt, K
    Hitchins, VM
    Woods, TO
    McNamee, SG
    Lyle, DB
    Brown, SA
    [J]. JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS, 2003, 66B (02): : 548 - 552