Sperm quality assessment via separation and sedimentation in a microfluidic device

被引:34
作者
Chen, Chang-Yu [1 ]
Chiang, Tsun-Chao [1 ]
Lin, Cheng-Ming [1 ]
Lin, Shu-Sheng [1 ]
Jong, De-Shien [3 ]
Tsai, Vincent F. -S. [2 ]
Hsieh, Ju-Ton [2 ]
Wo, Andrew M. [1 ]
机构
[1] Natl Taiwan Univ, Inst Appl Mech, Taipei 106, Taiwan
[2] Natl Taiwan Univ Hosp, Dept Urol, Taipei, Taiwan
[3] Natl Taiwan Univ, Dept Anim Sci & Technol, Taipei 106, Taiwan
关键词
ANALYZER IIB; FERTILIZATION; POPULATION; PARAMETERS; SYSTEM; MALES; MEN;
D O I
10.1039/c3an00900a
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A major reason for infertility is due to male factors, including the quality of spermatozoa, which is a primary factor and often difficult to assess, particularly the total sperm concentration and its motile percentage. This work presents a simple microfluidic device to assess sperm quality by quantifying both total and motile sperm counts. The key design feature of the microfluidic device is two channels separated by a permeative phase-guide structure, where one channel is filled with raw semen and the other with pure buffer. The semen sample was allowed to reach equilibrium in both chambers, whereas non-motile sperms remained in the original channel, and roughly half of the motile sperms would swim across the phase-guide barrier into the buffer channel. Sperms in each channel agglomerated into pellets after centrifugation, with the corresponding area representing total and motile sperm concentrations. Total sperm concentration up to 10(8) sperms per ml and motile percentage in the range of 10-70% were tested, encompassing the cutoff value of 40% stated by World Health Organization standards. Results from patient samples show compact and robust pellets after centrifugation. Comparison of total sperm concentration between the microfluidic device and the Makler chamber reveal they agree within 5% and show strong correlation, with a coefficient of determination of R-2 = 0.97. Motile sperm count between the microfluidic device and the Makler chamber agrees within 5%, with a coefficient of determination of R-2 = 0.84. Comparison of results from the Makler Chamber, sperm quality analyzer, and the microfluidic device revealed that results from the microfluidic device agree well with the Makler chamber. The sperm microfluidic chip analyzes both total and motile sperm concentrations in one spin, is accurate and easy to use, and should enable sperm quality analysis with ease.
引用
收藏
页码:4967 / 4974
页数:8
相关论文
共 40 条
  • [1] Evolution of semen quality in North-eastern Spain: a study in 22,759 infertile men over a 36 year period
    Andolz, P
    Bielsa, MA
    Vila, J
    [J]. HUMAN REPRODUCTION, 1999, 14 (03) : 731 - 735
  • [2] Development of a novel home sperm test
    Björndahl, L
    Kirkman-Brown, J
    Hart, G
    Rattle, S
    Barratt, CLR
    [J]. HUMAN REPRODUCTION, 2006, 21 (01) : 145 - 149
  • [3] Application of computer-assisted semen analysis to explain variations in pig fertility
    Broekhuijse, M. L. W. J.
    Sostaric, E.
    Feitsma, H.
    Gadella, B. M.
    [J]. JOURNAL OF ANIMAL SCIENCE, 2012, 90 (03) : 779 - 789
  • [4] Analysis of sperm concentration and motility in a microfluidic device
    Chen, Yu-An
    Huang, Zi-Wei
    Tsai, Fang-Sheng
    Chen, Chang-Yu
    Lin, Cheng-Ming
    Wo, Andrew M.
    [J]. MICROFLUIDICS AND NANOFLUIDICS, 2011, 10 (01) : 59 - 67
  • [5] Passively driven integrated microfluidic system for separation of motile sperm
    Cho, BS
    Schuster, TG
    Zhu, XY
    Chang, D
    Smith, GD
    Takayama, S
    [J]. ANALYTICAL CHEMISTRY, 2003, 75 (07) : 1671 - 1675
  • [6] SpermCheck® Fertility, an immunodiagnostic home test that detects normozoospermia and severe oligozoospermia
    Coppola, M. A.
    Klotz, K. L.
    Kim, K. -a.
    Cho, H. Y.
    Kang, J.
    Shetty, J.
    Howards, S. S.
    Flickinger, C. J.
    Herr, J. C.
    [J]. HUMAN REPRODUCTION, 2010, 25 (04) : 853 - 861
  • [7] Improving the density of jammed disordered packings using ellipsoids
    Donev, A
    Cisse, I
    Sachs, D
    Variano, E
    Stillinger, FH
    Connelly, R
    Torquato, S
    Chaikin, PM
    [J]. SCIENCE, 2004, 303 (5660) : 990 - 993
  • [8] Computer-assisted sperm analysis parameters in young fertile sperm donors and relationship with age
    Freour, Thomas
    Jean, Miguel
    Mirallie, Sophie
    Barriere, Paul
    [J]. SYSTEMS BIOLOGY IN REPRODUCTIVE MEDICINE, 2012, 58 (02) : 102 - 106
  • [9] Assessment of sperm quality analyzer IIB: Comparison with manual semen analysis and CASA
    Fuse, H
    Akashi, T
    Nozaki, T
    Nishio, R
    Mizuno, I
    [J]. ARCHIVES OF ANDROLOGY, 2005, 51 (01): : 65 - 67
  • [10] Microfluidic sedimentation cytometer for milk quality and bovine mastitis monitoring
    Garcia-Cordero, Jose L.
    Barrett, Louise M.
    O'Kennedy, Richard
    Ricco, Antonio J.
    [J]. BIOMEDICAL MICRODEVICES, 2010, 12 (06) : 1051 - 1059