Monte Carlo-based determination of radiation leakage dose around a dedicated IOERT accelerator

被引:6
作者
Baghani, Hamid Reza [1 ]
Aghdam, Seyed Rashid Hosseini [2 ]
Robatjazi, Mostafa [3 ]
Mahdavi, Seyed Rabi [4 ]
机构
[1] Hakim Sabzevari Univ, Phys Dept, Tohid Shahr St,POB 397, Sabzevar 617976487, Iran
[2] Shahid Beheshti Univ, Radiat Med Dept, Tehran, Iran
[3] Sabzevar Univ Med Sci, Dept Med Phys & Radiol Sci, Sabzevar, Iran
[4] Iran Univ Med Sci, Med Phys Dept, Tehran, Iran
关键词
Intraoperative electron radiotherapy; Radiation leakage; Monte Carlo simulation; Dosimetry; GAFCHROMIC EBT2 FILM; ELECTRON-BEAMS; DOSIMETRIC CHARACTERISTICS; IONIZATION CHAMBERS; LINEAR-ACCELERATOR; THERAPY; IORT; COMMITTEE;
D O I
10.1007/s00411-019-00786-1
中图分类号
Q [生物科学];
学科分类号
07 ; 0710 ; 09 ;
摘要
Evaluating the stray radiation around medical electron accelerators is a mandatory issue. Surveying the radiation leakage dose is important for patients, technicians, and health physicists, due to radiation protection aspects. Consequently, radiation leakage dose around the head of a mobile-dedicated intraoperative radiotherapy accelerator (LIAC), at different electron energies and field sizes have been evaluated in this study. More specifically, the MCNPX Monte Carlo code was used to model the LIAC head, connected applicator, and employed water phantom. Radiation leakage dose around the LIAC head was calculated for different energy and field sizes through tuning the Monte Carlo results to the practically measured doses. These measurements were performed using an Advance Markus ionization chamber inside an automated MP3-XS water phantom. The good agreement between the calculated dose distributions within the water tank and corresponding dose measurements show that the simulation model of the LIAC head is appropriate for radiation leakage assessment. The obtained radiation leakage dose distribution highly depends on the electron energy and applicator diameter. With increasing the electron energy, the leakage dose decreased, while increasing the field size increased the leakage dose. It is concluded that the rate of stray radiation and leakage dose around the LIAC head in both vertical and horizontal planes were acceptable according to the recommended radiation protection criteria. To meet the recommended dose limit (100 mu Sv/week for controlled areas), the maximum number of patients should be kept to four patients per week inside a standard and unshielded operating room.
引用
收藏
页码:263 / 276
页数:14
相关论文
共 24 条
  • [1] Aghdam MRH, 2017, J RADIOTHER PRACT, V16, P29, DOI 10.1017/S1460396916000455
  • [2] Andreo P., 2000, IAEA TECHNICAL REPOR
  • [3] [Anonymous], ANN ICRP
  • [4] [Anonymous], 2014, MOBILE ELECT ACCELER
  • [5] [Anonymous], 2005, 151 NCRP
  • [6] Comparing the dosimetric characteristics of the electron beam from dedicated intraoperative and conventional radiotherapy accelerators
    Baghani, Hamid Reza
    Aghamiri, Seyed Mahmoud Reza
    Mahdavi, Seyed Rabi
    Akbari, Mohammad Esmail
    Mirzaei, Hamid Reza
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2015, 16 (02): : 62 - 72
  • [7] Dosimetric evaluation of Gafchromic EBT2 film for breast intraoperative electron radiotherapy verification
    Baghani, Hamid Reza
    Aghamiri, S. Mahmoud Reza
    Mahdavi, S. Rabi
    Robatjazi, Mostafa
    Zadeh, Arezo Rahim
    Akbari, Mohammad Esmail
    Mirzaei, Hamid Reza
    Nafissi, Nahid
    Samsami, Majid
    [J]. PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2015, 31 (01): : 37 - 42
  • [8] Intraoperative radiation therapy using mobile electron linear accelerators: Report of AAPM radiation therapy committee task group no. 72
    Beddar, A. Sam
    Biggs, Peter J.
    Chang, Sha
    Ezzell, Gary A.
    Faddegon, Bruce A.
    Hensley, Frank W.
    Mills, Michael D.
    [J]. MEDICAL PHYSICS, 2006, 33 (05) : 1476 - 1489
  • [9] Biggs P, 2011, CURR CLIN ONCOL, P51, DOI 10.1007/978-1-61779-015-7_3
  • [10] The Italian affair: The employment of parallel-plate ionization chambers for dose measurements in high dose-per-pulse IORT electron beams
    Cella, L.
    Liuzzi, R.
    Salvatore, M.
    [J]. MEDICAL PHYSICS, 2010, 37 (06) : 2918 - 2924