Percutaneous valve replacement:: Significance of different delivery systems in vitro and in vivo

被引:9
作者
Attmann, T
Lutter, G
Quaden, R
Jahnke, T
Rumberg, K
Cremer, J
Muller-Hulsbeck, S
机构
[1] Univ Kiel, Sch Med, Dept Cardiovasc Surg, D-24105 Kiel, Germany
[2] Univ Kiel, Sch Med, Dept Radiol, D-24105 Kiel, Germany
关键词
equipment design; heart diseases; heart valve prosthesis; stents;
D O I
10.1007/s00270-005-0244-4
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Backgraound and purpose: Percutaneous heart valve replacement is an exciting growing field in cardiovascular medicine yet still with some major problems. Only sophisticated improvement of the instruments could make it a real alternative to conventional surgery. Therefore, the aim of this study was to evaluate different delivery devices for percutaneous heart valve replacement in vitro and in vivo. Methods: A catheter prototype designed by our group, and two commercially available devices for the delivery of esophageal stents and aortic endoprostheses, were tested. After in vitro experiments, an ovine animal model of transfemoral pulmonary valve implantation was established using biological valved self-expanding stents. Only the delivery device for aortic endografts (Medtronic, Talent, Santa Rosa, CA, USA) allowed fast in vitro procedures without material fatigue. This device was chosen for the in vivo tests. Results: Technical success was achieved in 9 of 10 animals (90%). One animal died after perforation of the ventricular wall. Orthotopic pulmonary placement was performed in 6 animals and intentional supravalvular valved stent placement in 3 animals. Conclusions: An adequate in vitro model for this evolving field of interventional heart valve replacement is presented. Furthermore, the present study pinpoints the key characteristics that are mandatory for a delivery system in percutaneous pulmonary valve implantation. With regard to the delivery device's ductility observed during this "venous" study, an approach to transfemoral aortic valve implantation seems feasible.
引用
收藏
页码:406 / 412
页数:7
相关论文
共 23 条
  • [1] TRANSLUMINAL IMPLANTATION OF ARTIFICIAL-HEART VALVES - DESCRIPTION OF A NEW EXPANDABLE AORTIC-VALVE AND INITIAL RESULTS WITH IMPLANTATION BY CATHETER TECHNIQUE IN CLOSED CHEST PIGS
    ANDERSEN, HR
    KNUDSEN, LL
    HASENKAM, JM
    [J]. EUROPEAN HEART JOURNAL, 1992, 13 (05) : 704 - 708
  • [2] Advances in experimental percutaneous pulmonary valve replacemen
    Attmann, T
    Jahnke, T
    Quaden, R
    Boening, A
    Muller-Hulsbeck, S
    Cremer, J
    Lutter, G
    [J]. ANNALS OF THORACIC SURGERY, 2005, 80 (03) : 969 - 975
  • [3] Transcatheter implantation of a bovine valve in pulmonary position - A lamb study
    Bonhoeffer, P
    Boudjemline, Y
    Saliba, Z
    Hausse, AO
    Aggoun, Y
    Bonnet, D
    Sidi, D
    Kachaner, J
    [J]. CIRCULATION, 2000, 102 (07) : 813 - 816
  • [4] Percutaneous insertion of the pulmonary valve
    Bonhoeffer, P
    Boudjemline, Y
    Qureshi, SA
    Le Bidois, J
    Iserin, L
    Acar, P
    Merckx, J
    Kachaner, J
    Sidi, D
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2002, 39 (10) : 1664 - 1669
  • [5] Steps toward percutaneous aortic valve replacement
    Boudjemline, Y
    Bonhoeffer, P
    [J]. CIRCULATION, 2002, 105 (06) : 775 - 778
  • [6] Stent-graft repair versus open surgery for the descending aorta:: A case-control study
    Brandt, M
    Hussel, K
    Walluscheck, KP
    Müller-Hülsbeck, S
    Jahnke, T
    Rahimi, A
    Cremer, J
    [J]. JOURNAL OF ENDOVASCULAR THERAPY, 2004, 11 (05) : 535 - 538
  • [7] Outcome after aortic valve replacement in octogenarians
    Chiappini, B
    Camurri, N
    Loforte, A
    Di Marco, L
    Di Bartolomeo, R
    Marinelli, G
    [J]. ANNALS OF THORACIC SURGERY, 2004, 78 (01) : 85 - 89
  • [8] The potential impact of percutaneous pulmonary valve stent implantation on right ventricular outflow tract re-intervention
    Coats, L
    Tsang, V
    Khambadkone, S
    van Doorn, C
    Cullen, S
    Deanfield, J
    de Leval, MR
    Bonhoeffer, P
    [J]. EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2005, 27 (04) : 536 - 542
  • [9] Percutaneous transcatheter implantation of an aortic valve prosthesis for calcific aortic stenosis - First human case description
    Cribier, A
    Eltchaninoff, H
    Bash, A
    Borenstein, N
    Tron, C
    Bauer, F
    Derumeaux, G
    Anselme, F
    Laborde, F
    Leon, MB
    [J]. CIRCULATION, 2002, 106 (24) : 3006 - 3008
  • [10] Outcomes in nonagenarians after heart valve replacement operation
    Edwards, MB
    Taylor, KM
    [J]. ANNALS OF THORACIC SURGERY, 2003, 75 (03) : 830 - 834