Verification of step-and-shoot IMRT delivery using a fast video-based electronic portal imaging device

被引:33
作者
Zeidan, OA [1 ]
Li, JG [1 ]
Ranade, M [1 ]
Stell, AM [1 ]
Dempsey, JF [1 ]
机构
[1] Univ Florida, Dept Radiat Oncol, Gainesville, FL 32610 USA
关键词
EPID; IMRT; MLC; fluence; CCD camera; IMRT treatment plan verification;
D O I
10.1118/1.1644518
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
We present an investigation into the use of a fast video-based electronic portal-imaging device (EPID) to study intensity modulated radiation therapy (IMRT) delivery. The aim of this study is to test the feasibility of using an EPID system to independently measure the orchestration of collimator leaf motion and beam fluence; simultaneously measuring both the delivered field fluence and shape as it exits the accelerator head during IMRT delivery. A fast EPID that consists of a terbium-doped gadolinium oxysulphide (GdO2S:Tb) scintillator coupled with an inexpensive commercial 30 frames-per-second (FPS) CCD-video recorder (16.7 ms shutter time) was employed for imaging IMRT delivery. The measurements were performed on a Varian 2100 C/D linear accelerator equipped with a 120-leaf multileaf-collimator (MLQ. A characterization of the EPID was performed that included measurements of spatial resolution, linac pulse-rate dependence, linear output response, signal uniformity, and imaging artifacts. The average pixel intensity for fields imaged with the EPID was found to be linear in the delivered monitor units of static non-lMRT fields between 3 X 3 and 15 X 15 cm(2). A systematic increase of the average pixel intensity was observed with increasing field size, leading to a maximum variation of 8%. Deliveries of a clinical step-andshoot mode leaf sequence were imaged at 600 MU/min. Measurements from this IMRT delivery were compared with experimentally validated MLC controller log files and were found to agree to within 5%. An analysis of the EPID image data allowed identification of three types of errors: (1) 5 out of 35 segments were undelivered; (2) redistributing all of the delivered segment MUs; and (3) leaf movement during segment delivery. Measurements with the EPID at lower dose rates showed poor agreement with log files due to an aliasing artifact. The study was extended to use a highspeed camera (1-1000 FPS and 10 mus shutter time) with our EPID to image the same delivery to demonstrate the feasibility of imaging without aliasing artifacts. High-speed imaging was shown to be a promising direction toward validating IMRT deliveries with reasonable image resolution and noise. (C) 2004 American Association of Physicists in Medicine.
引用
收藏
页码:463 / 476
页数:14
相关论文
共 29 条
  • [1] Physical characteristics of a commercial electronic portal imaging device
    Althof, VGM
    deBoer, JCJ
    Huizenga, H
    Stroom, JC
    Visser, AG
    Swanenburg, BN
    [J]. MEDICAL PHYSICS, 1996, 23 (11) : 1845 - 1855
  • [2] Fast 2D phantom dosimetry for scanning proton beams
    Boon, SN
    van Luijk, P
    Schippers, JM
    Meertens, H
    Denis, JM
    Vynckier, S
    Medin, J
    Grusell, E
    [J]. MEDICAL PHYSICS, 1998, 25 (04) : 464 - 475
  • [3] Verification of segmented beam delivery using a commercial electronic portal imaging device
    Curtin-Savard, AJ
    Podgorsak, EB
    [J]. MEDICAL PHYSICS, 1999, 26 (05) : 737 - 742
  • [4] Leaf trajectory calculation for dynamic multileaf collimation to realize optimized fluence profiles
    Dirkx, MLP
    Heijmen, BJM
    van Santvoort, JPC
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1998, 43 (05) : 1171 - 1184
  • [5] Ezzell G A, 2001, J Appl Clin Med Phys, V2, P138, DOI 10.1120/1.1386508
  • [6] Measurement of the response of Gd2O2S:Tb phosphor to 6 MV x-rays
    Glendinning, AG
    Hunt, SG
    Bonnett, DE
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2001, 46 (02) : 517 - 530
  • [7] Verification of dynamic multileaf collimation using an electronic portal imaging device
    James, HV
    Atherton, S
    Budgell, GJ
    Kirby, MC
    Williams, PC
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (02) : 495 - 509
  • [8] Monte Carlo simulations of the imaging performance of metal plate/phosphor screens used in radiotherapy
    Kausch, C
    Schreiber, B
    Kreuder, F
    Schmidt, R
    Dössel, O
    [J]. MEDICAL PHYSICS, 1999, 26 (10) : 2113 - 2124
  • [9] Intensity modulated radiotherapy dose delivery error from radiation field offset inaccuracy
    Kung, JH
    Chen, GTY
    [J]. MEDICAL PHYSICS, 2000, 27 (07) : 1617 - 1622
  • [10] Validation of dynamic MLC-controller log files using a two-dimensional diode array
    Li, JG
    Dempsey, JF
    Ding, L
    Liu, CR
    Palta, JR
    [J]. MEDICAL PHYSICS, 2003, 30 (05) : 799 - 805