Several modified piezoelectric quartz crystals for determination of omethoate

被引:2
作者
Fang, C [1 ]
Wang, XB [1 ]
Zhang, WM [1 ]
Zhou, XY [1 ]
机构
[1] Wuhan Univ, Dept Chem, Wuhan 430072, Peoples R China
关键词
PQC; PQC sensor; organophosphorus compounds; omethoate;
D O I
10.1016/S0039-9140(98)00375-0
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Determination of omethoate by modified PQC (piezoelectric quartz crystal) sensor is described. Several modified films were studied and compared. Omethoate of 0.1-10 ppm can be determined directly and sensitively with PVC-dioctyl sebacate film modified PQC. The modified crystal has good reversibility and reproductivity. The lifetime is about 1 month. (C) 1999 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:253 / 259
页数:7
相关论文
共 19 条
[1]  
AGRUILAR M, 1997, J CHROMATOGR, V778, P201
[2]  
DAI Q, 1996, CHEM SENSOR, V16, P107
[3]   PIEZOELECTRIC CRYSTAL COATING FOR DETECTION OF ORGANO-PHOSPHORUS COMPOUNDS [J].
GUILBAULT, GG ;
AFFOLTER, J ;
TOMITA, Y ;
KOLESAR, ES .
ANALYTICAL CHEMISTRY, 1981, 53 (13) :2057-2060
[4]   DETECTION OF ORGANO-PHOSPHORUS COMPOUNDS WITH A COATED PIEZOELECTRIC CRYSTAL [J].
GUILBAULT, GG ;
KRISTOFF, J ;
OWEN, D .
ANALYTICAL CHEMISTRY, 1985, 57 (08) :1754-1756
[5]  
JAMES HT, 1995, ANAL CHEM, V67, P3372
[6]   CHEMICAL SENSORS [J].
JANATA, J ;
JOSOWICZ, M ;
DEVANEY, DM .
ANALYTICAL CHEMISTRY, 1994, 66 (12) :R207-R228
[7]  
JANATA J, 1988, ANAL CHEM, V60, pR64
[8]  
JOHN JM, 1989, ANALYST, V114, P1173
[9]  
KING WH, 1969, RES DEV, V20, P28
[10]   EVALUATION OF COATING MATERIALS USED ON PIEZOELECTRIC SENSORS FOR THE DETECTION OF ORGANOPHOSPHORUS COMPOUNDS IN THE VAPOR-PHASE [J].
MILANKO, OS ;
MILINKOVIC, SA ;
RAJAKOVIC, LV .
ANALYTICA CHIMICA ACTA, 1992, 269 (02) :289-300