Electrothermal vaporization of mineral acid solutions in inductively coupled plasma mass spectrometry: comparison with sample nebulization

被引:13
作者
Mora, J [1 ]
Gras, L [1 ]
van Veen, EH [1 ]
de Loos-Vollebregt, MTC [1 ]
机构
[1] Delft Univ Technol, Mat Sci Lab, NL-2628 AL Delft, Netherlands
关键词
acids; electrothermal vaporization; nebulization; microwave desolvation; spectral interferences; modifier; inductively coupled plasma mass spectrometry;
D O I
10.1016/S0584-8547(99)00029-4
中图分类号
O433 [光谱学];
学科分类号
0703 ; 070302 ;
摘要
The analytical behaviour of an electrothermal vaporization (ETV) device for the introduction of mineral acid solutions in inductively coupled plasma mass spectrometry (ICP-MS) was evaluated. Water, nitric acid, hydrochloric acid, perchloric acid and sulphuric acid in concentrations within the 0.05-1.0 mol l(-1) range were studied. For all the acids tested, increasing the acid concentration increases the ion signal and deteriorates the precision. The magnitude of the signal enhancement depends on the analyte and on the acid considered. Acid solutions give rise to ion signals that are between 2 and 10 times higher than those with water. Among the acids tested, sulphuric acid provides the highest signals. The addition of palladium reduces matrix effects due to the acids and increases the signal in ETV ICP-MS. In comparison with conventional sample nebulization (CS), the ETV sample introduction system provides higher sensitivities (between 2 and 20 times higher) at the same acid concentration. The magnitude of this improvement is similar to that obtained with a microwave desolvation system (MWDS). The ETV sample introduction system gives rise to the lowest background signals from matrix-induced species. Due to this fact, the limits of detection (LODs) obtained for the isotopes affected by any interference are lower for ETV sample introduction than those obtained with the CS and the MWDS. For the isotopes that do not suffer from matrix-induced spectral interferences, the ETV gives rise to LODs higher than those obtained with the CS. For these isotopes the lowest LODs are obtained with MWDS. (C) 1999 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:959 / 974
页数:16
相关论文
共 46 条
[1]   REDUCTION OF POLYATOMIC ION INTERFERENCES IN INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY BY CRYOGENIC DESOLVATION [J].
ALVES, LC ;
WIEDERIN, DR ;
HOUK, RS .
ANALYTICAL CHEMISTRY, 1992, 64 (10) :1164-1169
[2]  
[Anonymous], AEROSOL TECHNOLOGY
[3]   Determination of rare earth elements in wine by inductively coupled plasma mass spectrometry using a microconcentric nebulizer [J].
Augagneur, S ;
Medina, B ;
Szpunar, J ;
Lobinski, R .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1996, 11 (09) :713-721
[4]   Characterization of an ultrasonic nebulizer membrane separation interface with inductively coupled plasma mass spectrometry for the determination of trace elements by solvent extraction - Invited lecture [J].
Brenner, IB ;
Zander, A ;
Plantz, M ;
Zhu, J .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1997, 12 (03) :273-279
[5]   Vaporization and atomization of the platinum group elements in the graphite furnace investigated by electrothermal vaporization inductively coupled plasma mass spectrometry [J].
Byrne, JP ;
Gregoire, DC ;
Benyounes, ME ;
Chakrabarti, CL .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1997, 52 (11) :1575-1586
[6]   EVALUATION OF A MODIFIED COMMERCIAL GRAPHITE-FURNACE FOR REDUCTION OF ISOBARIC INTERFERENCES IN ARGON INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY [J].
CAREY, JM ;
EVANS, EH ;
CARUSO, JA ;
SHEN, WL .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1991, 46 (13) :1711-1721
[7]  
DEBRAH E, 1995, ATOM SPECTROSC, V16, P197
[8]   THE ROLE OF CHEMICAL MODIFIERS IN ANALYTE TRANSPORT LOSS INTERFERENCES WITH ELECTROTHERMAL VAPORIZATION ICP-MASS SPECTROMETRY [J].
EDIGER, RD ;
BERES, SA .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1992, 47 (07) :907-922
[9]   INTERFERENCES IN INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY - A REVIEW [J].
EVANS, EH ;
GIGLIO, JJ .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1993, 8 (01) :1-18
[10]   Simultaneous determination of arsenic, antimony and selenium in aqueous matrices by electrothermal vaporization inductively coupled plasma mass spectrometry [J].
Fairman, B ;
Catterick, T .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1997, 12 (08) :863-866