Evaluation of the Biotyper MALDI-TOF MS system for identification of Staphylococcus species

被引:30
作者
Zhu, Wenming [1 ]
Sieradzki, Krzysztof [1 ]
Albrecht, Valerie [1 ]
McAllister, Sigrid [1 ]
Lin, Wen [1 ]
Stuchlik, Olga [2 ]
Limbago, Brandi [1 ]
Pohl, Jan [2 ]
Rasheed, J. Kamile [1 ]
机构
[1] Ctr Dis Control Prevent, Div Healthcare Qual Promot, Atlanta, GA USA
[2] Ctr Dis Control & Prevent, Div Sci Resources, Atlanta, GA 30333 USA
关键词
MALDI-TOF MS; Staphylococcus; Biochemical identification; DESORPTION IONIZATION-TIME; FLIGHT MASS-SPECTROMETRY; ASSISTED-LASER-DESORPTION/IONIZATION; COAGULASE-NEGATIVE STAPHYLOCOCCI; SCORE CUTOFF VALUES; EXTRACTION METHOD; BACTERIA; AUREUS; DISCRIMINATION; MICROBIOLOGY;
D O I
10.1016/j.mimet.2015.07.014
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The Bruker Biotyper MALDI-TOF MS (Biotyper) system, with a modified 30 minute formic acid extraction method, was evaluated by its ability to identify 216 clinical Staphylococcus isolates from the CDC reference collection comprising 23 species previously identified by conventional biochemical tests. 16S rDNA sequence analysis was used to resolve discrepancies. Of these, 209 (96.8%) isolates were correctly identified: 177 (84.7%) isolates had scores >= 2.0, while 32 (15.3%) had scores between 1.70 and 1.99. The Biotyper identification was inconsistent with the biochemical identification for seven (3.2%) isolates, but the Biotyper identifications were confirmed by 16S rDNA analysis. The distribution of low scores was strongly species-dependent, e.g. only 5% of Staphylococcus epidermidis and 4.8% of Staphylococcus aureus isolates scored below 2.0, while 100% of Staphylococcus cohnii, 75% of Staphylococcus sciuri, and 60% of Staphylococcus caprae produced low but accurate Biotyper scores. Our results demonstrate that the Biotyper can reliably identify Staphylococcus species with greater accuracy than conventional biochemicals. Broadening of the reference database by inclusion of additional examples of under-represented species could further optimize Biotyper results. Published by Elsevier B.V.
引用
收藏
页码:14 / 17
页数:4
相关论文
共 27 条
  • [1] Glycopeptide resistance in coagulase-negative staphylococci isolated in blood cultures from patients with hematological malignancies during three decades
    Ahlstrand, E.
    Svensson, K.
    Persson, L.
    Tidefelt, U.
    Soderquist, B.
    [J]. EUROPEAN JOURNAL OF CLINICAL MICROBIOLOGY & INFECTIOUS DISEASES, 2011, 30 (11) : 1349 - 1354
  • [2] Comparison of Direct Colony Method versus Extraction Method for Identification of Gram-Positive Cocci by Use of Bruker Biotyper Matrix-Assisted Laser Desorption Ionization-Time of Flight Mass Spectrometry
    Alatoom, Adnan A.
    Cunningham, Scott A.
    Ihde, Sherry M.
    Mandrekar, Jayawant
    Patel, Robin
    [J]. JOURNAL OF CLINICAL MICROBIOLOGY, 2011, 49 (08) : 2868 - 2873
  • [3] Bannerman TL, 2007, MANUAL OF CLINICAL MICROBIOLOGY, 9TH ED, P390
  • [4] Bernardo K, 2002, PROTEOMICS, V2, P747, DOI 10.1002/1615-9861(200206)2:6<747::AID-PROT747>3.0.CO
  • [5] 2-V
  • [6] Matrix-assisted laser desorption ionization time-of-flight mass spectrometry, a revolution in clinical microbial identification
    Bizzini, A.
    Greub, G.
    [J]. CLINICAL MICROBIOLOGY AND INFECTION, 2010, 16 (11) : 1614 - 1619
  • [7] Performance of Matrix-Assisted Laser Desorption Ionization-Time of Flight Mass Spectrometry for Identification of Bacterial Strains Routinely Isolated in a Clinical Microbiology Laboratory
    Bizzini, A.
    Durussel, C.
    Bille, J.
    Greub, G.
    Prod'hom, G.
    [J]. JOURNAL OF CLINICAL MICROBIOLOGY, 2010, 48 (05) : 1549 - 1554
  • [8] Matrix-Assisted Laser Desorption Ionization-Time of Flight Mass Spectrometry: a Fundamental Shift in the Routine Practice of Clinical Microbiology
    Clark, Andrew E.
    Kaleta, Erin J.
    Arora, Amit
    Wolk, Donna M.
    [J]. CLINICAL MICROBIOLOGY REVIEWS, 2013, 26 (03) : 547 - 603
  • [9] Identification of Staphylococcus aureus and determination of its methicillin resistance by matrix-assisted laser desorption/ionization, time-of-flight mass spectrometry
    Du, ZM
    Yang, RF
    Guo, ZB
    Song, YJ
    Wang, J
    [J]. ANALYTICAL CHEMISTRY, 2002, 74 (21) : 5487 - 5491
  • [10] Performances of the Vitek MS Matrix-Assisted Laser Desorption Ionization-Time of Flight Mass Spectrometry System for Rapid Identification of Bacteria in Routine Clinical Microbiology
    Dubois, Damien
    Grare, Marion
    Prere, Marie-Francoise
    Segonds, Christine
    Marty, Nicole
    Oswald, Eric
    [J]. JOURNAL OF CLINICAL MICROBIOLOGY, 2012, 50 (08) : 2568 - 2576