Comparison of fluid volume estimates in chronic hemodialysis patients by bioimpedance, direct isotopic, and dilution methods

被引:91
作者
Raimann, Jochen G. [1 ]
Zhu, Fansan [1 ]
Wang, Jack [2 ]
Thijssen, Stephan [1 ]
Kuhlmann, Martin K. [3 ]
Kotanko, Peter [1 ]
Levin, Nathan W. [1 ]
Kaysen, George A. [4 ]
机构
[1] Renal Res Inst, New York, NY 10065 USA
[2] St Lukes Roosevelt Hosp, New York, NY USA
[3] Vivantes Klinikum Friedrichshain, Berlin, Germany
[4] Univ Calif Davis, Davis, CA 95616 USA
关键词
chronic hemodialysis; hemodialysis; nutrition; water-electrolyte balance; BIOELECTRICAL-IMPEDANCE ANALYSIS; BODY CELL MASS; X-RAY ABSORPTIOMETRY; WATER; SPECTROSCOPY; POTASSIUM; VALIDATION; BROMIDE; SINGLE;
D O I
10.1038/ki.2013.358
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
Bioimpedance analysis (BIA) is accepted for the assessment of total-body water (TBW), intracellular fluid (ICF) and extracellular fluid (ECF). We aimed to compare precision and accuracy of single and multi-frequency-BIA to direct estimation methods (DEMs) of TBW, ECF, and ICF in hemodialysis patients. Linear regression analysis of volume estimates in 49 patients by single-and multi-frequency-BIA correlated significantly with DEMs. Bland-Altman analysis (BAA) found systemic bias for ECF single-frequency-BIA vs. ECF-DEMs. No other systematic biases were found. Proportional errors were found by BAA of ICF and ECF assessments with single-and multi-frequency bioimpedance spectroscopy compared to the DEMs. Comparisons of indirect methods (IEMs) to DEMs showed no significant differences and proportional errors. Root mean-squared-error analysis suggested slightly better accuracy and precision of ICF single-frequency-BIA vs. DEMs over ICF multi-frequency-BIA and IEMs to DEMs, and slightly better performance for ECF multi-frequency-BIA over both respective other methods. Compared to DEMs, there is slightly better accuracy for ECF multi-over single-frequency-BIA and ICF single-over multi-frequency-BIA. However the margin of differences between direct and indirect methods suggests that none of the analyzed methods served as a true "gold standard'', because indirect methods are almost equally precise compared to DEMs.
引用
收藏
页码:898 / 908
页数:11
相关论文
共 28 条
[1]  
[Anonymous], 2010, R LANG ENV STAT COMP
[2]  
[Anonymous], 2005, HUMAN BODY COMPOSITI
[3]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[4]   Effect of body mass index (BMI) on estimation of extracellular volume (ECV) in hemodialysis (FID) patients using segmental and whole body bioimpedance analysis [J].
Carter, M ;
Morris, AT ;
Zhu, FS ;
Zaluska, W ;
Levin, NW .
PHYSIOLOGICAL MEASUREMENT, 2005, 26 (02) :S93-S99
[5]   MULTIFREQUENCY BIOELECTRICAL-IMPEDANCE ESTIMATES THE DISTRIBUTION OF BODY-WATER [J].
CHA, KC ;
CHERTOW, GM ;
GONZALEZ, J ;
LAZARUS, JM ;
WILMORE, DW .
JOURNAL OF APPLIED PHYSIOLOGY, 1995, 79 (04) :1316-1319
[6]  
Cole K.S., 1968, Membranes, Ions and Impulses: A Chapter of Classical Biophysics
[7]   Dispersion and absorption in dielectrics II Direct current characteristics [J].
Cole, KS ;
Cole, RH .
JOURNAL OF CHEMICAL PHYSICS, 1942, 10 (02) :98-105
[8]   Predicting body cell mass with bioimpedance by using theoretical methods: A technological review [J].
DeLorenzo, A ;
Andreoli, A ;
Matthie, J ;
Withers, P .
JOURNAL OF APPLIED PHYSIOLOGY, 1997, 82 (05) :1542-1558
[9]   Single- and multi-frequency bioelectrical impedance analyses to analyse body composition in maintenance haemodialysis patients: comparison with dual-energy x-ray absorptiometry [J].
Donadio, C. ;
Ben Halim, A. ;
Caprio, F. ;
Grassi, G. ;
Khedr, B. ;
Mazzantini, M. .
PHYSIOLOGICAL MEASUREMENT, 2008, 29 (06) :S517-S524
[10]   Human body composition: In vivo methods [J].
Ellis, KJ .
PHYSIOLOGICAL REVIEWS, 2000, 80 (02) :649-680