Entrance surface dose and image quality: comparison of adult chest and abdominal X-ray examinations in general practitioner clinics, public and private hospitals in Malaysia

被引:1
作者
Hambali, Ahmad Shariff [2 ]
Ng, Kwan-Hoong [1 ]
Abdullah, Basri Johan Jeet [1 ]
Wang, Hwee-Beng [2 ]
Jamal, Noriah [3 ]
Spelic, David C. [4 ]
Suleiman, Orhan H. [4 ]
机构
[1] Univ Malaya, Dept Biomed Imaging, Kuala Lumpur 50603, Malaysia
[2] Fed Govt Adm Ctr, Engn Serv Div, Minist Hlth Malaysia, Putrajaya 62590, Malaysia
[3] Agensi Nuklear Malaysia, Med Technol Div, Kajang 43000, Malaysia
[4] US FDA, Ctr Devices & Radiol Hlth, Silver Spring, MD 20903 USA
关键词
DIAGNOSTIC-RADIOLOGY; RADIOGRAPHY; REDUCTION; ABDOMEN;
D O I
10.1093/rpd/ncp007
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
This study was undertaken to compare the entrance surface dose (ESD) and image quality of adult chest and abdominal X-ray examinations conducted at general practitioner (GP) clinics, and public and private hospitals in Malaysia. The surveyed facilities were randomly selected within a given category (28 GP clinics, 20 public hospitals and 15 private hospitals). Only departmental X-ray units were involved in the survey. Chest examinations were done at all facilities, while only hospitals performed abdominal examinations. This study used the x-ray attenuation phantoms and protocols developed for the Nationwide Evaluation of X-ray Trends (NEXT) survey program in the United States. The ESD was calculated from measurements of exposure and clinical geometry. An image quality test tool was used to evaluate the low-contrast detectability and high-contrast detail performance under typical clinical conditions. The median ESD value for the adult chest X-ray examination was the highest (0.25 mGy) at GP clinics, followed by private hospitals (0.22 mGy) and public hospitals (0.17 mGy). The median ESD for the adult abdominal X-ray examination at public hospitals (3.35 mGy) was higher than that for private hospitals (2.81 mGy). Results of image quality assessment for the chest X-ray examination show that all facility types have a similar median spatial resolution and low-contrast detectability. For the abdominal X-ray examination, public hospitals have a similar median spatial resolution but larger low-contrast detectability compared with private hospitals. The results of this survey clearly show that there is room for further improvement in performing chest and abdominal X-ray examinations in Malaysia.
引用
收藏
页码:25 / 34
页数:10
相关论文
共 23 条
[1]  
*AM COLL RAD, 2006, ACR TECHN STAND DIAG
[2]  
[Anonymous], 1996, IAEA SAF SER
[3]  
*C RAD CONTR PROGR, C RAD CONTR PROGR E, V52
[4]  
*C RAD CONTR PROGR, NEXT 2002 PROT SURV
[5]  
*C RAD CONTR PROGR, NEXT 2001 PROT SURV
[6]  
*C RAD CONTR PROGR, C RAD CONTR PROGR DI
[7]  
Dosimetry Working Party of the Institute of physical sciences in medicine, 1992, NAT PROT PAT DOS MEA
[8]   Reference values for diagnostic radiology: Application and impact [J].
Gray, JE ;
Archer, BR ;
Butler, PF ;
Hobbs, BB ;
Mettler, FA ;
Pizzutiello, RJ ;
Schueler, BA ;
Strauss, KJ ;
Suleiman, OH ;
Yaffe, MJ .
RADIOLOGY, 2005, 235 (02) :354-358
[9]  
HART D, 1996, R289 NRPB
[10]  
*IAEA COMM EUR COM, 1995, IAEATECDOC796 CEC