Experimental results from Al/p-CdTe/Pt X-ray detectors

被引:27
作者
Abbene, L. [1 ,2 ]
Gerardi, G. [1 ]
Turturici, A. A. [1 ]
Del Sordo, S. [2 ]
Principato, F. [1 ]
机构
[1] Univ Palermo, Dipartimento Fis, I-90128 Palermo, Italy
[2] INAF IASF Palermo, I-90146 Palermo, Italy
关键词
CdTe detectors; Diode detectors; X-ray and gamma ray spectroscopy; Digital pulse processing; Pulse shape correction; High photon counting rate; POLARIZATION PHENOMENA; RADIATION DETECTORS; RESOLUTION; SPECTROSCOPY; PROTOTYPE; PROGRESS;
D O I
10.1016/j.nima.2013.03.016
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Recently, Al/CdTe/PL detectors have been proposed for the development of high resolution X-ray spectrometers. Due to the low leakage currents, these detectors allovv high electric fields and the pixellization of anodes vvith the possibility to realize single charge carrier sensing detectors. In this work, we report on the results of electrical and spectroscopic investigations on CdTe diode detectors with Al/ CdTe/Pt electrode configuration (4.1 x 4.1 x 0.75 and 4.1 x 4.1 x 2 mm(3)). The detectors are characterized by very low leakage currents in the reverse bias operation: 0.3 nA at 25 'C and 24 pA at 25 'C under a bias voltage of 1000 V. The spectroscopic performance of the detectors at both low and high photon counting rates were also investigated with a focus on the minimization of tune instability, generally termed as polarization, looking for the optimum bias voltage and temperature. Good time stability, during a long-term operation of 10 IA, was observed for both detectors at 25 'C and by using an electric field of 5000 V/cm. The 2 mm thick detector exhibited good energy resolution of 6.1%, 2.5% and 2.0% (FWHM) at 22.1 keV, 59.5 and 122.1 key, respectively. Performance enhancements were obtained by using digital pulse processing techniques, especially at high photon counting rates (300 kcps). The 2 mm thick detector, after a digital pulse shape correction (PSC), is characterized by similar performance to the thin detector ones, opening up to the use of thick CdTe detectors without excessive performance degradations. This work was carried out in the framework of the development of portable X-ray spectrometers for both laboratory research and medical applications. (C) 2013 Elsevier BM. All rights reserved,
引用
收藏
页码:135 / 140
页数:6
相关论文
共 34 条
[1]   Spectroscopic response of a CdZnTe multiple electrode detector [J].
Abbene, L. ;
Del Sordo, S. ;
Fauci, F. ;
Gerardi, G. ;
La Manna, A. ;
Raso, G. ;
Cola, A. ;
Perillo, E. ;
Raulo, A. ;
Gostilo, V. ;
Stumbo, S. .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2007, 583 (2-3) :324-331
[2]   Performance enhancements of compound semiconductor radiation detectors using digital pulse processing techniques [J].
Abbene, L. ;
Gerardi, G. .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2011, 654 (01) :340-348
[3]   High-rate x-ray spectroscopy in mammography with a CdTe detector: A digital pulse processing approach [J].
Abbene, L. ;
Gerardi, G. ;
Principato, F. ;
Del Sordo, S. ;
Ienzi, R. ;
Raso, G. .
MEDICAL PHYSICS, 2010, 37 (12) :6147-6156
[4]   Performance of a digital CdTe X-ray spectrometer in low and high counting rate environment [J].
Abbene, L. ;
Gerardi, G. ;
Del Sordo, S. ;
Raso, G. .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2010, 621 (1-3) :447-452
[5]   Hard x-ray response of pixellated CdZnTe detectors [J].
Abbene, L. ;
Del Sordo, S. ;
Caroli, E. ;
Gerardi, G. ;
Raso, G. ;
Caccia, S. ;
Bertuccio, G. .
JOURNAL OF APPLIED PHYSICS, 2009, 105 (12)
[6]  
ABBENE L, 2012, IEEE NUCL SCI S, P94636
[7]   X-ray spectroscopy and dosimetry with a portable CdTe device [J].
Abbene, Leonardo ;
La Manna, Angelo ;
Fauci, Francesco ;
Gerardi, Gaetano ;
Stumbo, Simone ;
Raso, Giuseppe .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2007, 571 (1-2) :373-377
[8]   Direct Measurement of Mammographic X-Ray Spectra with a Digital CdTe Detection System [J].
Abbene, Leonardo ;
Gerardi, Gaetano ;
Principato, Fabio ;
Del Sordo, Stefano ;
Raso, Giuseppe .
SENSORS, 2012, 12 (06) :8390-8404
[9]   Charge transport properties in CdZnTe detectors grown by the vertical Bridgman technique [J].
Auricchio, N. ;
Marchini, L. ;
Caroli, E. ;
Zappettini, A. ;
Abbene, L. ;
Honkimaki, V. .
JOURNAL OF APPLIED PHYSICS, 2011, 110 (12)
[10]   Comparison of two portable solid state detectors with an improved collimation and alignment device for mammographic x-ray spectroscopy [J].
Bottigli, U. ;
Golosio, B. ;
Masala, G. L. ;
Oliva, P. ;
Stumbo, S. ;
Delogu, P. ;
Fantacci, M. E. ;
Abbene, L. ;
Fauci, F. ;
Raso, G. .
MEDICAL PHYSICS, 2006, 33 (09) :3469-3477