Calculating dose and output factors for wedged photon radiotherapy fields using a convolution/superposition method

被引:41
作者
Liu, HH [1 ]
Mackie, TR [1 ]
McCullough, EC [1 ]
机构
[1] UNIV WISCONSIN,DEPT MED PHYS,MADISON,WI 53706
关键词
convolution/superposition; beam modifiers; wedges; compensators; wedge generated radiation; extra-focal photons; beam hardening; wedge factors; output factors; asymmetric fields;
D O I
10.1118/1.597959
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
We have developed a convolution/superposition method to calculate dose distributions in photon treatment fields with beam modifiers such as physical wedges. The dose component due to wedge generated radiation was accounted for by using an extended phantom model, which integrated a wedge, an air gap, and a patient phantom as the calculation phantom. The inhomogeneities in the extended phantom and the effect of beam hardening by the wedge were both corrected for in the convolution dose calculation. The calculated dose was verified by Monte Carlo simulation of the same extended phantom. A new dual photon source model was also used in the convolution method to account for both primary photons from the target and extra-focal photons from the primary collimator and flattening filter. Thus, realistic photon energy fluence distributions in the extended phantom were used for the dose calculation. The calculated dose distributions and the wedge factors agreed with the measured data within 2% for a variety of treatment fields including asymmetric fields. Our results showed that the wedge-generated radiation could contribute a significant fraction of the total dose in patients. This dose component depends on a specific field configuration, thus wedge factor changes with photon energy, wedge angle, field size, depth, and patient phantom SSD. The variation of the wedge factor can be predicted accurately by our convolution approach with the extended phantom model, which allows for more accurate dose or monitor unit computation for photon fields with beam modifiers. (C) 1997 American Association of Physicists in Medicine.
引用
收藏
页码:1714 / 1728
页数:15
相关论文
共 34 条
[1]   MODELING TRANSMISSION AND SCATTER FOR PHOTON-BEAM ATTENUATORS [J].
AHNESJO, A ;
WEBER, L ;
NILSSON, P .
MEDICAL PHYSICS, 1995, 22 (11) :1711-1720
[2]   A PENCIL BEAM MODEL FOR PHOTON DOSE CALCULATION [J].
AHNESJO, A ;
SAXNER, M ;
TREPP, A .
MEDICAL PHYSICS, 1992, 19 (02) :263-273
[3]   COLLIMATOR SCATTER IN PHOTON THERAPY BEAMS [J].
AHNESJO, A .
MEDICAL PHYSICS, 1995, 22 (03) :267-278
[4]   COLLAPSED CONE CONVOLUTION OF RADIANT ENERGY FOR PHOTON DOSE CALCULATION IN HETEROGENEOUS MEDIA [J].
AHNESJO, A .
MEDICAL PHYSICS, 1989, 16 (04) :577-592
[5]   MONITOR UNIT CALCULATION FOR LARGE WEDGED HIGH-ENERGY PHOTON BEAMS [J].
CHEN, Z ;
REINSTEIN, LE ;
RAMSAY, EB .
PHYSICS IN MEDICINE AND BIOLOGY, 1995, 40 (03) :465-470
[6]   BEAM PROFILES ALONG THE NONWEDGED DIRECTION FOR LARGE WEDGED FIELDS [J].
CHUI, CS ;
LOSASSO, T .
MEDICAL PHYSICS, 1994, 21 (11) :1685-1690
[7]  
HEUKELOM S, 1994, RADIOTHER ONCOL, V32, P66
[8]   The variation in output of symmetric, asymmetric and irregularly shaped wedged radiotherapy fields [J].
Hounsell, AR ;
Wilkinson, JM .
PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (10) :2155-2172
[9]   Effects of scatter generated by beam-modifying absorbers in megavoltage photon beams [J].
Islam, MK ;
VanDyk, J .
MEDICAL PHYSICS, 1995, 22 (12) :2075-2081
[10]   SEPARATION OF DOSE-GRADIENT EFFECT FROM BEAM-HARDENING EFFECT ON WEDGE FACTORS IN PHOTON FIELDS [J].
KALEND, AM ;
WU, A ;
YODER, V ;
MAITZ, A .
MEDICAL PHYSICS, 1990, 17 (04) :701-704