Sutureless bioprosthesis for aortic valve replacement: Surgical and clinical outcomes

被引:4
作者
Ferreira, Ricardo [1 ,2 ]
Rua, Nuno [3 ]
Sena, Andre [1 ]
Velho, Tiago R. [1 ,4 ]
Goncalves, Joao [1 ]
Junqueira, Nadia [1 ]
Almeida, Ana G. [2 ,5 ]
Nobre, Angelo [1 ,2 ]
Pinto, Fausto [2 ,5 ]
机构
[1] Ctr Hosp Lisboa Norte, Hosp Santa Maria, Dept Cardiothorac Surg, Lisbon, Portugal
[2] Univ Lisbon, Ctr Cardiovasc, Fac Med, Lisbon, Portugal
[3] Univ Beira Interior, Fac Ciencias Med, Covilha, Portugal
[4] Inst Gulbenkian Ciencias, Innate Immun & Inflammat Lab, Oeiras, Portugal
[5] Ctr Hosp Lisboa Norte, Hosp Santa Maria, Dept Cardiol, Lisbon, Portugal
关键词
aortic valve disease; aortic valve replacement; aortic valve stenosis; Perceval (R) valve; sutureless bioprosthesis; TRANSCATHETER; IMPLANTATION; SURGERY; METAANALYSIS; MANAGEMENT; SOCIETY;
D O I
10.1111/jocs.17113
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: Aortic valve stenosis is the most common adult valve disease in industrialized countries. The aging population and the increase in comorbidities urge the development of safer alternatives to the current surgical treatment. Sutureless bioprosthesis has shown promising results, especially in complex procedures and in patients requiring concomitant surgeries. Objectives: Assess the clinical and hemodynamic performance, safety, and durability of the Perceval (R) prosthetic valve. Methods: This single-center retrospective longitudinal cohort study collected data from all adult patients with aortic valve disease who underwent aortic valve replacement with a Perceval (R) prosthetic valve between February 2015 and October 2020. Of the 196 patients included (mean age 77.20 +/- 5.08 years; 45.4% female; mean EuroSCORE II 2.91 +/- 2.20%), the majority had aortic stenosis. Results: Overall mean cross-clamp and cardiopulmonary bypass times were 33.31 +/- 14.09 min and 45.55 +/- 19.04 min, respectively. Mean intensive care unit and hospital stay were 3.32 +/- 3.24 days and 7.70 +/- 5.82 days, respectively. Procedural success was 98.99%, as two explants occurred. Four valves were reimplanted due to intraoperative misplacement. Mean transvalvular gradients were 7.82 +/- 3.62 mmHg. Pacemaker implantation occurred in 12.8% of patients, new-onset atrial fibrillation in 21.9% and renal replacement support was necessary for 3.1%. Early mortality was 2.0%. We report no structural valve deterioration, strokes, or endocarditis, and one successfully treated valve thrombosis. Conclusions: Our study confirms the excellent clinical and hemodynamic performance and safety of a truly sutureless aortic valve, up to a 5-year follow-up. These results were consistent in isolated and concomitant interventions, solidifying this device as a viable option for the treatment of isolated aortic valve disease.
引用
收藏
页码:4774 / 4782
页数:9
相关论文
共 38 条
[11]  
Di Bacco L, 2017, INNOVATIONS, V12, P483, DOI 10.1097/IMI.0000000000000427
[12]   Effect of sutureless implantation of the Perceval S aortic valve bioprosthesis on intraoperative and early postoperative outcomes [J].
Flameng, Willem ;
Herregods, Marie-Christine ;
Hermans, Hadewich ;
Van der Mieren, Gerry ;
Vercalsteren, Monique ;
Poortmans, Gert ;
Van Hemelrijck, Jan ;
Meuris, Bart .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2011, 142 (06) :1453-1457
[13]   Sutureless valve and rapid deployment valves: a systematic review and meta-analysis of comparative studies [J].
Flynn, Campbell D. ;
Williams, Michael L. ;
Chakos, Adam ;
Hirst, Lucy ;
Muston, Benjamin ;
Tian, David H. .
ANNALS OF CARDIOTHORACIC SURGERY, 2020, 9 (05) :364-+
[14]   Sutureless Perceval Aortic Valve Replacement: Results of Two European Centers [J].
Folliguet, Thierry A. ;
Laborde, Francois ;
Zannis, Konstantinos ;
Ghorayeb, Gabriel ;
Haverich, Axel ;
Shrestha, Malakh .
ANNALS OF THORACIC SURGERY, 2012, 93 (05) :1483-1488
[15]   The German Aortic Valve Registry (GARY): in-hospital outcome [J].
Hamm, Christian W. ;
Moellmann, Helge ;
Holzhey, David ;
Beckmann, Andreas ;
Veit, Christof ;
Figulla, Hans-Reiner ;
Cremer, J. ;
Kuck, Karl-Heinz ;
Lange, Ruediger ;
Zahn, Ralf ;
Sack, Stefan ;
Schuler, Gerhard ;
Walther, Thomas ;
Beyersdorf, Friedhelm ;
Boehm, Michael ;
Heusch, Gerd ;
Funkat, Anne-Kathrin ;
Meinertz, Thomas ;
Neumann, Till ;
Papoutsis, Konstantinos ;
Schneider, Steffen ;
Welz, Armin ;
Mohr, Friedrich W. .
EUROPEAN HEART JOURNAL, 2014, 35 (24) :1588-1598
[16]  
Heimansohn D, 2016, OPERAT TECHNIQ THORA, V21, P282, DOI [10.1053/j.optechstcvs.2017.09.004, DOI 10.1053/J.OPTECHSTCVS.2017.09.004]
[17]   Aortic Stenosis: Pathophysiology, Diagnosis, and Therapy [J].
Joseph, Jessica ;
Naqvi, Syed Yaseen ;
Giri, Jay ;
Goldberg, Sheldon .
AMERICAN JOURNAL OF MEDICINE, 2017, 130 (03) :253-263
[18]   Management of Patients With Aortic Valve Stenosis [J].
Kanwar, Amrit ;
Thaden, Jeremy J. ;
Nkomo, Vuyisile T. .
MAYO CLINIC PROCEEDINGS, 2018, 93 (04) :488-508
[19]  
Kappetein AP, 2012, J AM COLL CARDIOL, V60, P1438, DOI [10.1016/j.jacc.2012.09.001, 10.1093/ejcts/ezs533]
[20]   Two-Year Outcomes after Transcatheter or Surgical Aortic-Valve Replacement [J].
Kodali, Susheel K. ;
Williams, Mathew R. ;
Smith, Craig R. ;
Svensson, Lars G. ;
Webb, John G. ;
Makkar, Raj R. ;
Fontana, Gregory P. ;
Dewey, Todd M. ;
Thourani, Vinod H. ;
Pichard, Augusto D. ;
Fischbein, Michael ;
Szeto, Wilson Y. ;
Lim, Scott ;
Greason, Kevin L. ;
Teirstein, Paul S. ;
Malaisrie, S. Chris ;
Douglas, Pamela S. ;
Hahn, Rebecca T. ;
Whisenant, Brian ;
Zajarias, Alan ;
Wang, Duolao ;
Akin, Jodi J. ;
Anderson, William N. ;
Leon, Martin B. .
NEW ENGLAND JOURNAL OF MEDICINE, 2012, 366 (18) :1686-1695