Comparison of virtual monoenergetic and polyenergetic images reconstructed from dual-layer detector CT angiography of the head and neck

被引:44
作者
Neuhaus, Victor [1 ]
Hokamp, Nils Grosse [1 ]
Abdullayev, Nuran [1 ]
Maus, Volker [1 ]
Kabbasch, Christoph [1 ]
Mpotsaris, Anastasios [1 ]
Maintz, David [1 ]
Borggrefe, Jan [1 ]
机构
[1] Univ Hosp Cologne, Dept Diagnost & Intervent Radiol, Kerpener Str 62, D-50937 Cologne, Germany
关键词
Dual-energy CT; Computed tomography angiography; Head and neck; Virtual monoenergetic images; Image quality; COMPUTED TOMOGRAPHIC ANGIOGRAPHY; KILOELECTRON VOLT SETTINGS; ENERGY CT; MATERIAL DECOMPOSITION; QUALITY; OPTIMIZATION; CONTRAST; ABDOMEN; ALGORITHM; STROKE;
D O I
10.1007/s00330-017-5081-8
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
To compare the image quality of virtual monoenergetic images and polyenergetic images reconstructed from dual-layer detector CT angiography (DLCTA). Thirty patients who underwent DLCTA of the head and neck were retrospectively identified and polyenergetic as well as virtual monoenergetic images (40 to 120 keV) were reconstructed. Signals (+/- SD) of the cervical and cerebral vessels as well as lateral pterygoid muscle and the air surrounding the head were measured to calculate the CNR and SNR. In addition, subjective image quality was assessed using a 5-point Likert scale. Student's t-test and Wilcoxon test were used to determine statistical significance. Compared to polyenergetic images, although noise increased with lower keV, CNR (p < 0.02) and SNR (p > 0.05) of the cervical, petrous and intracranial vessels were improved in virtual monoenergetic images at 40 keV and virtual monoenergetic images at 45 keV were also rated superior regarding vascular contrast, assessment of arteries close to the skull base and small arterial branches (p < 0.0001 each). Compared to polyenergetic images, virtual monoenergetic images reconstructed from DLCTA at low keV ranging from 40 to 45 keV improve the objective and subjective image quality of extra- and intracranial vessels and facilitate assessment of vessels close to the skull base and of small arterial branches.
引用
收藏
页码:1102 / 1110
页数:9
相关论文
共 29 条
[1]   Comprehensive Comparison of Virtual Monoenergetic and Linearly Blended Reconstruction Techniques in Third-Generation Dual-Source Dual-Energy Computed Tomography Angiography of the Thorax and Abdomen [J].
Albrecht, Moritz H. ;
Trommer, Jesko ;
Wichmann, Julian L. ;
Scholtz, Jan-Erik ;
Martin, Simon S. ;
Lehnert, Thomas ;
Vogl, Thomas J. ;
Bodelle, Boris .
INVESTIGATIVE RADIOLOGY, 2016, 51 (09) :582-590
[2]   Advanced image-based virtual monoenergetic dual-energy CT angiography of the abdomen: optimization of kiloelectron volt settings to improve image contrast [J].
Albrecht, Moritz H. ;
Scholtz, Jan-Erik ;
Huesers, Kristina ;
Beeres, Martin ;
Bucher, Andreas M. ;
Kaup, Moritz ;
Martin, Simon S. ;
Fischer, Sebastian ;
Bodelle, Boris ;
Bauer, Ralf W. ;
Lehnert, Thomas ;
Vogl, Thomas J. ;
Wichmann, Julian L. .
EUROPEAN RADIOLOGY, 2016, 26 (06) :1863-1870
[3]   ENERGY-SELECTIVE RECONSTRUCTIONS IN X-RAY COMPUTERIZED TOMOGRAPHY [J].
ALVAREZ, RE ;
MACOVSKI, A .
PHYSICS IN MEDICINE AND BIOLOGY, 1976, 21 (05) :733-744
[4]   ASPECTS on CTA source images versus unenhanced CT - Added value in predicting final infarct extent and clinical outcome [J].
Coutts, SB ;
Lev, MH ;
Eliasziw, M ;
Roccatagliata, L ;
Hill, MD ;
Schwamm, LH ;
Pexman, JHW ;
Koroshetz, WJ ;
Hudon, ME ;
Buchan, AM ;
Gonzalez, RG ;
Demchuk, AM .
STROKE, 2004, 35 (11) :2472-2476
[5]   New Applications of Cardiac Computed Tomography Dual-Energy, Spectral, and Molecular CT Imaging [J].
Danad, Ibrahim ;
Fayad, Zahi A. ;
Willemink, Martin J. ;
Min, James K. .
JACC-CARDIOVASCULAR IMAGING, 2015, 8 (06) :710-723
[6]   Combined iterative reconstruction and image-domain decomposition for dual energy CT using total-variation regularization [J].
Dong, Xue ;
Niu, Tianye ;
Zhua, Lei .
MEDICAL PHYSICS, 2014, 41 (05)
[7]   Carotid CTA: Radiation Exposure and Image Quality with the Use of Attenuation-Based, Automated Kilovolt Selection [J].
Eller, A. ;
Wuest, W. ;
Kramer, M. ;
May, M. ;
Schmid, A. ;
Uder, M. ;
Lell, M. M. .
AMERICAN JOURNAL OF NEURORADIOLOGY, 2014, 35 (02) :237-241
[8]   AN ALGORITHM FOR NOISE SUPPRESSION IN DUAL ENERGY CT MATERIAL DENSITY IMAGES [J].
KALENDER, WA ;
KLOTZ, E ;
KOSTARIDOU, L .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1988, 7 (03) :218-224
[9]   GENERALIZED IMAGE COMBINATIONS IN DUAL KVP DIGITAL RADIOGRAPHY [J].
LEHMANN, LA ;
ALVAREZ, RE ;
MACOVSKI, A ;
BRODY, WR ;
PELC, NJ ;
RIEDERER, SJ ;
HALL, AL .
MEDICAL PHYSICS, 1981, 8 (05) :659-667
[10]   Optimizing Contrast Media Injection Protocols in State-of-the Art Computed Tomographic Angiography [J].
Lell, Michael M. ;
Jost, Gregor ;
Korporaal, Johannes Georg ;
Mahnken, Andreas Horst ;
Flohr, Thomas G. ;
Uder, Michael ;
Pietsch, Hubertus .
INVESTIGATIVE RADIOLOGY, 2015, 50 (03) :161-167