Low-dose cone-beam CT (LD-CBCT) reconstruction for image-guided radiation therapy (IGRT) by three-dimensional dual-dictionary learning

被引:10
作者
Song, Ying [1 ]
Zhang, Weikang [2 ]
Zhang, Hong [1 ]
Wang, Qiang [1 ]
Xiao, Qing [1 ]
Li, Zhibing [1 ]
Wei, Xing [1 ]
Lai, Jialu [1 ]
Wang, Xuetao [1 ]
Li, Wan [1 ]
Zhong, Quan [1 ]
Gong, Pan [1 ]
Zhong, Renming [1 ]
Zhao, Jun [2 ]
机构
[1] Sichuan Univ, West China Hosp, Dept Radiotherapy, Ctr Canc, 37 Guo Xue Alley, Chengdu 610065, Peoples R China
[2] Shanghai Jiao Tong Univ, Sch Biomed Engn, 800 Dongchuan Rd, Shanghai 610065, Peoples R China
基金
中国国家自然科学基金;
关键词
Low dose; CBCT reconstruction; Radiation therapy; Dictionary learning; COMPUTED-TOMOGRAPHY; PROJECTION;
D O I
10.1186/s13014-020-01630-3
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background To develop a low-dose cone beam CT (LD-CBCT) reconstruction method named simultaneous algebraic reconstruction technique and dual-dictionary learning (SART-DDL) joint algorithm for image guided radiation therapy (IGRT) and evaluate its imaging quality and clinical application ability. Methods In this retrospective study, 62 CBCT image sets from February 2018 to July 2018 at west china hospital were randomly collected from 42 head and neck patients (mean [standard deviation] age, 49.7 [11.4] years, 12 females and 30 males). All image sets were retrospectively reconstructed by SART-DDL (resultant D-CBCT image sets) with 18% less clinical raw projections. Reconstruction quality was evaluated by quantitative parameters compared with SART and Total Variation minimization (SART-TV) joint reconstruction algorithm with pairedttest. Five-grade subjective grading evaluations were done by two oncologists in a blind manner compared with clinically used Feldkamp-Davis-Kress algorithm CBCT images (resultant F-CBCT image sets) and the grading results were compared by paired Wilcoxon rank test. Registration results between D-CBCT and F-CBCT were compared. D-CBCT image geometry fidelity was tested. Results The mean peak signal to noise ratio of D-CBCT was 1.7 dB higher than SART-TV reconstructions (P < .001, SART-DDL vs SART-TV, 36.36 +/- 0.55 dB vs 34.68 +/- 0.28 dB). All D-CBCT images were recognized as clinically acceptable without significant difference with F-CBCT in subjective grading (P > .05). In clinical registration, the maximum translational and rotational difference was 1.8 mm and 1.7 degree respectively. The horizontal, vertical and sagittal geometry fidelity of D-CBCT were acceptable. Conclusions The image quality, geometry fidelity and clinical application ability of D-CBCT are comparable to that of the F-CBCT for head-and-neck patients with 18% less projections by SART-DDL.
引用
收藏
页数:9
相关论文
共 24 条
  • [1] K-SVD: An algorithm for designing overcomplete dictionaries for sparse representation
    Aharon, Michal
    Elad, Michael
    Bruckstein, Alfred
    [J]. IEEE TRANSACTIONS ON SIGNAL PROCESSING, 2006, 54 (11) : 4311 - 4322
  • [2] [Anonymous], 2010, SPRINGER
  • [3] MR-guidance - a clinical study to evaluate a shuttle- based MR-linac connection to provide MR-guided radiotherapy
    Bostel, Tilman
    Nicolay, Nils H.
    Grossmann, Joerg G.
    Mohr, Angela
    Delorme, Stefan
    Echner, Gernot
    Haering, Peter
    Debus, Juergen
    Sterzing, Florian
    [J]. RADIATION ONCOLOGY, 2014, 9
  • [4] Low dose CBCT reconstruction via prior contour based total variation (PCTV) regularization: a feasibility study
    Chen, Yingxuan
    Yin, Fang-Fang
    Zhang, Yawei
    Zhang, You
    Ren, Lei
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2018, 63 (08)
  • [5] Accurate patient dosimetry of kilovoltage cone-beam CT in radiation therapy
    Ding, George X.
    Duggan, Dennis M.
    Coffey, Charles W.
    [J]. MEDICAL PHYSICS, 2008, 35 (03) : 1135 - 1144
  • [6] Hao J, 2012, IEEE NUCL SCI CONF R, P3422
  • [7] GPU-based fast cone beam CT reconstruction from undersampled and noisy projection data via total variation
    Jia, Xun
    Lou, Yifei
    Li, Ruijiang
    Song, William Y.
    Jiang, Steve B.
    [J]. MEDICAL PHYSICS, 2010, 37 (04) : 1757 - 1760
  • [8] Radiation dose from cone beam computed tomography for image-guided radiation therapy
    Kan, Monica W. K.
    Leung, Lucullus H. T.
    Wong, Wicger
    Lam, Nelson
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2008, 70 (01): : 272 - 279
  • [9] Kulkarni A, 2015, IEEE INT SYMP CIRC S, P970, DOI 10.1109/ISCAS.2015.7168797
  • [10] Liu J, 2016, IEEE INT S BIOM IM