A Novel Smartphone App for the Measurement of Ultra-Short-Term and Short-Term Heart Rate Variability: Validity and Reliability Study

被引:21
作者
Chen, Yung-Sheng [1 ]
Lu, Wan-An [2 ]
Pagaduan, Jeffrey C. [3 ]
Kuo, Cheng-Deng [4 ,5 ]
机构
[1] Univ Taipei, Dept Exercise & Hlth Sci, Taipei, Taiwan
[2] Fo Guang Univ, Inst Cultural Asset & Reinvent, Yilan, Taiwan
[3] Univ Tasmania, Coll Hlth & Med, Sch Hlth Sci, Launceston, Tas, Australia
[4] Taipei Vet Gen Hosp, Dept Med Res, 201,Sect 2,Shipai Rd, Taipei 112, Taiwan
[5] Tanyu Res Lab, Taipei, Taiwan
关键词
heart rate variability; smartphone; reproducibility; limits of agreement; autonomic nervous function; HRV; AGREEMENT; EXERCISE; INDEXES; REST;
D O I
10.2196/18761
中图分类号
R19 [保健组织与事业(卫生事业管理)];
学科分类号
摘要
Background: Smartphone apps for heart rate variability (HRV) measurement have been extensively developed in the last decade. However, ultra-short-term HRV recordings taken by wearable devices have not been examined. Objective: The aims of this study were the following: (1) to compare the validity and reliability of ultra-short-term and short-term HRV time-domain and frequency-domain variables in a novel smartphone app, Pulse Express Pro (PEP), and (2) to determine the agreement of HRV assessments between an electrocardiogram (ECG) and PEP. Methods: In total, 60 healthy adults were recruited to participate in this study (mean age 22.3 years [SD 3.0 years], mean height 168.4 cm [SD 8.0 cm], mean body weight 64.2 kg [SD 11.5 kg]). A 5-minute resting HRV measurement was recorded via ECG and PEP in a sitting position Standard deviation of normal R-R interval (SDNN), root mean square of successive R-R interval (RMSSD), proportion of NN50 divided by the total number of RR intervals (pNN50), normalized very-low-frequency power (nVLF), normalized low-frequency power (nLF), and normalized high-frequency power (nHF) were analyzed within 9 time segments of HRV recordings: 0-1 minute, 1-2 minutes, 2-3 minutes, 3-4 minutes, 4-5 minutes, 0-2 minutes, 0-3 minutes, 0-4 minutes, and 0-5 minutes (standard) Standardized differences (ES), intraclass correlation coefficients (ICC), and the Spearman product-moment correlation were used to compare the validity and reliability of each time segment to the standard measurement (0-5 minutes). Limits of agreement were assessed by using Bland-Altman plot analysis. Results: Compared to standard measures in both ECG and PEP, pNN50, SDNN, and RMSSD variables showed trivial ES (<0.2) and very large to nearly perfect ICC and Spearman correlation coefficient values in all time segments (>0.8). The nVLF, nLF, and nHF demonstrated a variation of ES (from trivial to small effects, 0.01-0.40), ICC (from moderate to nearly perfect, 0.39-0.96), and Spearman correlation coefficient values (from moderate to nearly perfect, 0.40-0.96). Furthermore, the Bland-Altman plots showed relatively narrow values of mean difference between the ECG and PEP after consecutive 1-minute recordings for SDNN, RMSSD, and pNN50. Acceptable limits of agreement were found after consecutive 3-minute recordings for nLF and nHF. Conclusions: Using the PEP app to facilitate a 1-minute ultra-short-term recording is suggested for time-domain HRV indices (SDNN, RMSSD, and pNN50) to interpret autonomic functions during stabilization. When using frequency-domain HRV indices (nLF and nHF) via the PEP app, a recording of at least 3 minutes is needed for accurate measurement.
引用
收藏
页数:16
相关论文
共 44 条
[1]  
Ali K, 2001, 2001 IEEE ANT PROP S, P376, DOI [10.1109/aps.2001.960112, DOI 10.1109/APS.2001.960112]
[2]   Reliability of Ultra-Short-Term Analysis as a Surrogate of Standard 5-Min Analysis of Heart Rate Variability [J].
Baek, Hyun Jae ;
Cho, Chul-Ho ;
Cho, Jaegeol ;
Woo, Jong-Min .
TELEMEDICINE AND E-HEALTH, 2015, 21 (05) :404-414
[3]   Effects of Gender and Aging on Differential Autonomic Responses to Orthostatic Maneuvers [J].
Barantke, Melanie ;
Krauss, Timothy ;
Ortak, Jasmin ;
Lieb, Wolfgang ;
Reppel, Michael ;
Burgdorf, Christof ;
Pramstaller, Peter P. ;
Schunkert, Heribert ;
Bonnemeier, Hendrik .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2008, 19 (12) :1296-1303
[4]   The LF/HF ratio does not accurately measure cardiac sympatho-vagal balance [J].
Billman, George E. .
FRONTIERS IN PHYSIOLOGY, 2013, 4
[5]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[6]   Monitoring training status ith HR measures: do all roads lead to Rome? [J].
Buchheit, Martin .
FRONTIERS IN PHYSIOLOGY, 2014, 5
[7]  
Camm AJ, 1996, CIRCULATION, V93, P1043
[8]   Ultra-short term HRV features as surrogates of short term HRV: a case study on mental stress detection in real life [J].
Castaldo, R. ;
Montesinos, L. ;
Melillo, P. ;
James, C. ;
Pecchia, L. .
BMC MEDICAL INFORMATICS AND DECISION MAKING, 2019, 19 (1)
[9]   Consecutive ultra-short-term heart rate variability to track dynamic changes in autonomic nervous system during and after exercise [J].
Chen, Xiang ;
Yao, Ruijie ;
Yin, Gezhen ;
Li, Jin .
PHYSIOLOGICAL MEASUREMENT, 2017, 38 (07) :1384-1395
[10]  
Chen Yabo, 2015, Proceedings of the CSU-EPSA, V27, P1, DOI 10.3969/j.issn.1003-8930.2015.05.01