Magnetic field interactions in adjustable hydrocephalus shunts

被引:44
作者
Lavinio, Andrea [2 ,4 ]
Harding, Sally [3 ]
Van Der Boogaard, Floor [2 ]
Czosnyka, Marek [1 ,2 ]
Smielewski, Peter [2 ]
Richards, Hugh K. [2 ]
Pickard, John D. [2 ,3 ]
Czosnyka, Zofia H. [1 ,2 ]
机构
[1] Univ Cambridge, United Kingdom Shunt Evaluat Lab, Cambridge CB2 1TN, England
[2] Univ Cambridge, Neurosurg Unit, Dept Clin Neurosci, Cambridge CB2 1TN, England
[3] Univ Cambridge, Wolfson Brain Imaging Ctr, Cambridge CB2 1TN, England
[4] Univ Brescia, Inst Anaesthesiol & Intens Care Med, I-25121 Brescia, Italy
基金
英国医学研究理事会;
关键词
hydrocephalus; magnetic field; magnetic resonance imaging; safety; shunt;
D O I
10.3171/PED/2008/2/9/222
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Object. Exposing patients with ventricular shunts to magnetic fields and MR imaging procedures poses a significant risk of unintentional changes in shunt settings. Shunt valves can also generate considerable imaging artifacts. The purpose of this study was to determine the magnetic field safety and MR imaging compatibility of 5 adjustable models of hydrocephalus shunts. Methods. The Codman Hakim (regular and with SiphonGuard), Miethke ProGAV, Medtronic Strata, Sophysa Sophy and Polaris programmable valves were tested in a low-intensity magnetic field, and then translational attraction (TA), magnetic torque (MT), and volume of artifacts on T1-weighted spin echo (SE) and gradient echo (GE) pulse sequences in a 3-T MR imaging unit were measured. Results. The ProGAV and Polaris valves were immune to unintentional reprogramming by magnetic fields up to 3 T. Other valves randomly changed settings, starting from the intensity of field: Sophy valve 24 mT, Strata valve 30 mT, and both Codman Hakim programmable valves from 42 mT. Shunt performances in the 3-T MR imaging unit are reported in the order of compatibility: 1) Codman Hakim regular, TA = 0.005 N, MT = 0.000 Nm, GE = 30 cm(3), SE = 2 cm(3); 2) Miethke ProGAV, TA = 0.001 N, MT = 1.4 X 10(-3) Nm, GE = 231 cm(3), SE = 13 cm(3); 3) Codman Hakim with SiphonGuard, TA = 0.005 N, MT = 2.3 X 10(-3) Nm, GE = 233 cm(3), SE = 19 cm(3); 4) Medtronic Strata, TA = 0.27 N, MT = 18.0 X 10-3 Nm, GE = 484 cm(3), SE = 86 cm(3); 5) Sophysa Sophy, TA = 0.82 N, MT = 38.9 X 10(-3) Nm, GE = 758 cm(3), SE = 72 cm(3); and 6) Sophysa Polaris, TA = 0.80 N, MT = 39.6 X 10(-3) Nm, GE = 954 cm(3), SE = 100 cm(3). Conclusions. All valves, with the exception of the Polaris and ProGAV models, are prone to unintentional reprogramming when exposed to heterogeneous magnetic fields stronger than 40 mT. All tested valves can be considered safe for 3-T MR imaging. All valves generated a distortion of the MR image, especially the GE sequences.
引用
收藏
页码:222 / 228
页数:7
相关论文
共 12 条
[1]   In vitro hydrodynamic properties of the Miethke proGAV hydrocephalus shunt [J].
Allin D.M. ;
Czosnyka Z.H. ;
Czosnyka M. ;
Richards H.K. ;
Pickard J.D. .
Cerebrospinal Fluid Research, 3 (1)
[2]   Adjustment and malfunction of a programmable valve after exposure to toy magnets - Case report [J].
Anderson, RCE ;
Walker, ML ;
Viner, JM ;
Kestle, JRW .
JOURNAL OF NEUROSURGERY, 2004, 101 (02) :222-225
[3]   Laboratory testing of hydrocephalus shunts - Conclusion of the UK Shunt Evaluation Programme [J].
Czosnyka, Z ;
Czosnyka, M ;
Richards, HK ;
Pickard, JD .
ACTA NEUROCHIRURGICA, 2002, 144 (06) :525-538
[4]  
Czosnyka ZH, 2005, ACT NEUR S, V95, P223
[5]   Effect of 3-tesla magnetic resonance imaging on various pressure programmable shunt valves [J].
Inoue, T ;
Kuzu, Y ;
Ogasawara, K ;
Ogawa, A .
JOURNAL OF NEUROSURGERY, 2005, 103 (02) :163-165
[6]   Effect of vagus nerve stimulator magnet on programmable shunt settings [J].
Jandial, R ;
Aryan, HE ;
Hughes, SA ;
Levy, ML .
NEUROSURGERY, 2004, 55 (03) :627-629
[7]   Experience with the Strata valve in the management of shunt overdrainage [J].
Kondageski, Charles ;
Thompson, Dominic ;
Reynolds, Mandy ;
Hayward, Richard D. .
JOURNAL OF NEUROSURGERY, 2007, 106 (02) :95-102
[8]   Effect of 3 T MR1 on the function of shunt valves - Evaluation of Paedi GAV, Dual Switch and proGAV [J].
Lindner, D ;
Preul, C ;
Trantakis, C ;
Moeller, H ;
Meixensberger, J .
EUROPEAN JOURNAL OF RADIOLOGY, 2005, 56 (01) :56-59
[9]   Early programmable valve malfunctions in pediatric hydrocephalus [J].
Mangano, FT ;
Menendez, JA ;
Habrock, T ;
Narayan, P ;
Leonard, JR ;
Park, TS ;
Smyth, MD .
JOURNAL OF NEUROSURGERY, 2005, 103 (06) :501-507
[10]   Adjustable vs set-pressure valves decrease the risk of proximal shunt obstruction in the treatment of pediatric hydrocephalus [J].
McGirt, Matthew J. ;
Buck, Donald W., II ;
Sciubba, Daniel ;
Woodworth, Graeme F. ;
Carson, Benjamin ;
Weingart, Jon ;
Jallo, George .
CHILDS NERVOUS SYSTEM, 2007, 23 (03) :289-295