Influence of femur size and morphology on load transfer in the resurfaced femoral head: A large scale, multi-subject finite element study

被引:30
作者
Bryan, Rebecca [2 ]
Nair, Prasanth B. [3 ,4 ]
Taylor, Mark [1 ,2 ]
机构
[1] Flinders Univ S Australia, Med Device Res Inst, Sch Comp Sci Engn & Math, Adelaide, SA, Australia
[2] Univ Southampton, Bioengn Sci Res Grp, Fac Engn & Environm, Southampton, Hants, England
[3] Univ Southampton, Sch Engn Sci, Computat Engn & Design Grp, Southampton, Hants, England
[4] Univ Toronto, Inst Aerosp Studies, Toronto, ON, Canada
关键词
Femoral resurfacing; Arthroplasty; Statistical model; Automated implantation; Finite element analysis; HIP-ARTHROPLASTY; NECK FRACTURES; RISK-FACTORS; INTERFACE;
D O I
10.1016/j.jbiomech.2012.05.015
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Femoral resurfacing has become an increasingly popular procedure, especially for young, active patients. The procedure is known to alter load transfer through the proximal femur and this has been linked with the most commonly observed complication, neck fracture. An intriguing observation noted by registry data and clinical studies is an inverse relationship between implant size and revision rate. While computational analysis has become an established part of biomedical engineering, the majority of work uses a single or small set of bone models, with a single implant size, due to the constraints of time and data availability. Therefore, it has been infeasible to run a study incorporating natural inter-patient variability or the performance of smaller implants could not be meaningfully studied. In previous work a statistical model of the whole femur was used to generate large numbers of unique, realistic, FE-ready femur models describing both geometry and material properties. The current study demonstrates a methodology for virtually implanting and performing stress analysis of cemented femoral resurfacing components, with model specific sizing and orientation. Automated analysis of 400 generated femurs, in both implanted and intact configurations showed the strain changes induced by resurfacing. This produced a statistically meaningful number of results and allowed the examination of outliers. Results showed increased femoral neck strain change's potentially increasing the risk of neck fracture, associated with smaller, less dense femurs and smaller implant sizes; agreeing with clinical observations. The study demonstrates a methodology for more comprehensive analyses, based on populations rather than individuals. (C) 2012 Elsevier Ltd. All rights reserved.
引用
收藏
页码:1952 / 1958
页数:7
相关论文
共 35 条
[1]  
Amstutz H C, 1998, J Orthop Sci, V3, P169, DOI 10.1007/s007760050038
[2]   Fracture of the neck of the femur after surface arthroplasty of the hip [J].
Amstutz, HC ;
Campbell, PA ;
Le Duff, MJ .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2004, 86A (09) :1874-1877
[3]  
[Anonymous], 2008, AUSTR ORTH ASS NAT J
[4]   Statistical modelling of the whole human femur incorporating geometric and material properties [J].
Bryan, Rebecca ;
Mohan, P. Surya ;
Hopkins, Andrew ;
Galloway, Francis ;
Taylor, Mark ;
Nair, Prasanth B. .
MEDICAL ENGINEERING & PHYSICS, 2010, 32 (01) :57-65
[5]   Use of a statistical model of the whole femur in a large scale, multi-model study of femoral neck fracture risk [J].
Bryan, Rebecca ;
Nair, Prasanth B. ;
Taylor, Mark .
JOURNAL OF BIOMECHANICS, 2009, 42 (13) :2171-2176
[6]   Stress shielding and stress concentration of contemporary epiphyseal hip prostheses [J].
Cristofolini, L. ;
Juszczyk, M. ;
Taddei, F. ;
Field, R. E. ;
Rushton, N. ;
Viceconti, M. .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2009, 223 (H1) :27-44
[7]  
Feldesman MR, 1996, AM J PHYS ANTHROPOL, V100, P207
[8]   Musculo-skeletal loading conditions at the hip during walking and stair climbing [J].
Heller, MO ;
Bergmann, G ;
Deuretzbacher, G ;
Dürselen, L ;
Pohl, M ;
Claes, L ;
Haas, NP ;
Duda, GN .
JOURNAL OF BIOMECHANICS, 2001, 34 (07) :883-893
[9]   Assessment of mobility after multi-level surgery for cerebral palsy [J].
Graham, H. K. ;
Harvey, A. .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2007, 89B (08) :993-994
[10]   INTERFACE STRESSES IN THE RESURFACED HIP - FINITE-ELEMENT ANALYSIS OF LOAD TRANSMISSION IN THE FEMORAL-HEAD [J].
HUISKES, R ;
STRENS, PHGE ;
VANHECK, J ;
SLOOFF, TJJH .
ACTA ORTHOPAEDICA SCANDINAVICA, 1985, 56 (06) :474-478