Corneal Subbasal Nerve Density: A Comparison of Two Confocal Microscopes

被引:43
作者
Erie, Elizabeth A. [1 ]
McLaren, Jay W. [1 ]
Kittleson, Katrina M. [1 ]
Patel, Sanjay V. [1 ]
Erie, Jay C. [1 ]
Bourne, William M. [1 ]
机构
[1] Mayo Clin, Dept Ophthalmol, Rochester, MN 55905 USA
来源
EYE & CONTACT LENS-SCIENCE AND CLINICAL PRACTICE | 2008年 / 34卷 / 06期
关键词
Confocal microscopy; Subbasal nerve density; ConfoScan; 4; Tandem scanning;
D O I
10.1097/ICL.0b013e31818b74f4
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose. To compare subbasal nerve densities estimated from images recorded by the tandem scanning and the ConfoScan 4 confocal microscopes. Methods. Confocal microscopy was used to estimate subbasal nerve density in 62 corneas of 40 subjects (18 corneas of 18 normal subjects and 44 corneas of 22 patients between 1 and 12 months after LASIK). At each examination, corneas were scanned first by using tandem scanning and then by using a ConfoScan 4 confocal microscope. Subbasal nerves from 2 to 4 scans per cornea were traced by using a semi-automated nerve analysis program. Nerve density was expressed as total nerve length divided by the sample area (mu m/mm(2)). Differences in nerve density between instruments were examined by using paired tests. Results. In normal corneas, subbasal nerve density was 10,658 +/- 5,581 mu m/mm(2) (mean +/- SD) with the ConfoScan 4 and 5,534 +/- 1,850 mu m/mm(2) with the tandem scanning microscope (P<0.0001). One to 12 months after LASIK, mean subbasal nerve density was 2,477 +/- 3,514 mu m/mm(2) estimated with the ConfoScan 4 and 844 +/- 983 mu m/mm(2) estimated with the tandem scanning (P = 0.0003). Estimates of nerve density were correlated between instruments (r = 0.71, P<0.0001), although the mean difference between instruments was 2,308 +/- 3,885 mu m/mm(2) (P<0.0001). Conclusions. Mean subbasal nerve density estimated with the ConfoScan 4 was 2 to 3 times higher than density estimated with the tandem scanning confocal microscope. These differences must be considered when comparing subbasal nerve densities between studies that use different confocal microscopes.
引用
收藏
页码:322 / 325
页数:4
相关论文
共 13 条
[1]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[2]   Corneal reinnervation after LASIK: Prospective 3-year longitudinal study [J].
Calvillo, MP ;
McLaren, JW ;
Hodge, DO ;
Bourne, WM .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2004, 45 (11) :3991-3996
[3]   Subbasal nerve fiber regeneration after LASIK and LASEK assessed by noncontact esthesiometry and in vivo confoccal microscopy: Prospective study [J].
Darwish, Taym ;
Brahma, Arun ;
O'Donnell, Clare ;
Efron, Nathan .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2007, 33 (09) :1515-1521
[4]  
Erie Jay C, 2003, Trans Am Ophthalmol Soc, V101, P293
[5]   The effect of age on the corneal subbasal nerve plexus [J].
Erie, JC ;
McLaren, JW ;
Hodge, DO ;
Bourne, WM .
CORNEA, 2005, 24 (06) :705-709
[6]  
McLaren Jay W, 2005, Eye Contact Lens, V31, P28, DOI 10.1097/01.ICL.0000151948.92593.C3
[7]  
McLaren Jay W, 2007, Eye Contact Lens, V33, P185, DOI 10.1097/ICL.0b013e31802b3114
[8]   Corneal thickness measurement by confocal microscopy, ultrasound, and scanning slit methods [J].
McLaren, JW ;
Nau, CB ;
Erie, JC ;
Bourne, WM .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2004, 137 (06) :1011-1020
[9]   Design and validation of a tool for neurite tracing and analysis in fluorescence microscopy images [J].
Meijering, E ;
Jacob, M ;
Sarria, JCF ;
Steiner, P ;
Hirling, H ;
Unser, M .
CYTOMETRY PART A, 2004, 58A (02) :167-176
[10]   Corneal nerves:: structure, contents and function [J].
Müller, LJ ;
Marfurt, CF ;
Kruse, F ;
Tervo, TMT .
EXPERIMENTAL EYE RESEARCH, 2003, 76 (05) :521-542