An improved sliding-load calibration procedure using a semiparametric circle-fitting procedure

被引:7
作者
Vandersteen, G
Rolain, Y
Schoukens, J
Verschueren, A
机构
[1] Vrije Universiteit Brüssel (VUB/TW-ELEC)
关键词
calibration; estimation;
D O I
10.1109/22.598437
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Circle-fitting problems often occur in microwave engineering when dealing with variable delays, e.g., during calibration using a sliding load, This paper proposes an efficient semiparametric circle-fitting procedure, which takes into account the phase relationships over the frequencies. It produces more accurate results than the standard sliding-load calibration, requires only three positions on the sliding load for the whole frequency band, and is more robust to the settings of the positions of the sliding load, The proposed method also has the ability to detect whether or not the sliding load is defective or out of its specifications, This can be done by using only three positions on the sliding load, Optimal-position settings are then proposed, The performance of the proposed method is illustrated on sliding-load measurements up to 50 GHz, demonstrating the ability of detecting modeling errors and showing that the accuracy of the proposed method using three positions is comparable to the standard method with six positions.
引用
收藏
页码:1027 / 1033
页数:7
相关论文
共 7 条
[1]  
Cadzow J. A., 1990, IEEE ASSP Magazine, V7, P12, DOI 10.1109/53.62941
[2]  
Fletcher R., 1981, PRACTICAL METHODS OP
[3]  
Golub G, 2013, Matrix Computations, V4th
[4]  
Golub G. H., 1979, Proc. Of The 1979 Army Numerical Analysis And Comp. Conf., P1
[5]   CIRCLE FITTING PROCEDURE AND ITS ERROR ANALYSIS [J].
KASA, I .
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 1976, 25 (01) :8-14
[6]  
Rizzi P.A., 1988, MICROWAVE ENG PASSIV
[7]  
SCHOUKENS J, 1997, AUTOMATICA, V33