Virtual and real bench testing of a new percutaneous valve device: a case study

被引:17
作者
Biglino, Giovanni [1 ,2 ]
Capelli, Claudio [1 ,2 ]
Binazzi, Alia [1 ,2 ,3 ]
Reggiani, Rita [1 ,2 ,3 ]
Cosentino, Daria [4 ]
Migliavacca, Francesco [3 ]
Bonhoeffer, Philipp [5 ]
Taylor, Andrew M. [1 ,2 ]
Schievano, Silvia [1 ,2 ]
机构
[1] UCL Inst Cardiovasc Sci, Ctr Cardiovasc Imaging, London WC1N 3JH, England
[2] Great Ormond St Hosp Sick Children, London WC1N 3JH, England
[3] Politecn Milan, Dept Struct Engn, Lab Biol Struct Mech, I-20133 Milan, Italy
[4] UCL, Dept Mech Engn, London, England
[5] CNR Reg Toscana, Fdn G Monasterio, Pisa, Italy
基金
英国工程与自然科学研究理事会;
关键词
percutaneous valve; congenital heart disease; rapid prototyping; finite element analysis; stent-graft; validation; PULMONARY VALVE; RIGHT-VENTRICLE; OUTFLOW TRACT; IMPLANTATION;
D O I
10.4244/EIJV8I1A19
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Aims: To validate patient-specific computational testing of a second-generation device for percutaneous pulmonary valve implantation (PPVI), against realistic in vitro data. Methods and results: Tests were initially carried out in a simple loading mode, performing a compliance test on a rapid prototyped cylinder. This model was reproduced computationally and validated against the experimental data. A second-generation PPVI stent-graft, with no valve mounted, was then deployed in a simplified cylindrical geometry, measuring its displacement when subjected to a pressure pulse. Experimental and computational measurements were in good agreement. Finally, having selected a patient regarded as unsuitable for first-generation PPVI, but potentially suitable for a second-generation device, the stent-graft was studied in the rapidly prototyped patient-specific right ventricular outflow tract (RVOT). Stent positioning and radial displacements with pulsatile flow were observed in a mock circuit using fluoroscopy imaging. Stent deformation and anchoring were measured both in vitro and computationally. Both tests indicated that the stent was well anchored in the RVOT, especially in the distal position, and its central region was rounded, ensuring, were a valve present, optimal valve function. Conclusion: We suggest that an experimentally validated computational model can be used for preclinical device characterisation and patient selection.
引用
收藏
页码:120 / 128
页数:9
相关论文
共 31 条
[1]   A simple method for noninvasive estimation of pulmonary vascular resistance [J].
Abbas, AE ;
Fortuin, FD ;
Schiller, NB ;
Appleton, CP ;
Moreno, CA ;
Lester, SJ .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2003, 41 (06) :1021-1027
[2]  
Abunassar C, 2011, WORKSH COMP METH CAR
[3]  
[Anonymous], 2010, AB AN US MAN V 6 10, VV, P314
[4]  
[Anonymous], FULLCURE MAT DAT
[5]   Verification and validation in computational engineering and science: basic concepts [J].
Babuska, I ;
Oden, JT .
COMPUTER METHODS IN APPLIED MECHANICS AND ENGINEERING, 2004, 193 (36-38) :4057-4066
[6]  
Biglino G., 2011, ASAIO J, V57, P74
[7]   Percutaneous replacement of pulmonary valve in a right-ventricle to pulmonary-artery prosthetic conduit with valve dysfunction [J].
Bonhoeffer, P ;
Boudjemline, Y ;
Saliba, Z ;
Merckx, J ;
Aggoun, Y ;
Bonnet, D ;
Acar, P ;
Le Bidois, J ;
Sidi, D ;
Kachaner, J .
LANCET, 2000, 356 (9239) :1403-1405
[8]   Are animal experiments the crux for decision making in whether new heart valves can be brought to clinical practice? [J].
Bonhoeffer, Philipp .
EUROINTERVENTION, 2010, 5 (06) :643-+
[9]  
Bowles C, 2010, CARDIOVASC ENG TECHN, V1, P10
[10]   Patient-specific reconstructed anatomies and computer simulations are fundamental for selecting medical device treatment: application to a new percutaneous pulmonary valve [J].
Capelli, Claudio ;
Taylor, Andrew M. ;
Migliavacca, Francesco ;
Bonhoeffer, Philipp ;
Schievano, Silvia .
PHILOSOPHICAL TRANSACTIONS OF THE ROYAL SOCIETY A-MATHEMATICAL PHYSICAL AND ENGINEERING SCIENCES, 2010, 368 (1921) :3027-3038