Passive vortex currents in magneto- and electrocardiography: comparison of magnetic and electric signal strengths

被引:28
作者
Dutz, S
Bellemann, ME
Leder, U
Haueisen, J
机构
[1] Univ Appl Sci, Dept Biomed Engn, D-07745 Jena, Germany
[2] Univ Jena, Dept Neurol, Biomagnet Ctr, D-07747 Jena, Germany
[3] Tech Univ Ilmenau, Inst Biomed Engn, D-98684 Ilmenau, Germany
[4] Univ Jena, Clin Internal Med 3, D-07747 Jena, Germany
关键词
D O I
10.1088/0031-9155/51/1/011
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Vortex currents may be of importance in the early diagnosis of myocardial infarction caused by an occlusion of a coronary artery. We investigated the influence of a passive vortex current distribution, modelled by different conductivities in a hollow cylinder, on the localization error and on the signal strength in both the magnetocardiogram and the electrocardiogram. The hollow cylinder was mounted in a realistically shaped physical torso phantom. A platinum dipole was inserted into the cylinder. The compartment boundaries were modelled with two special ionic exchange membranes. The conductivity ratio of the cylinder compartment to the torso compartment was varied from 0.25 to 100. We compared the simultaneously measured magnetic and electric signal strengths as a function of this conductivity ratio. We found that an increasing conductivity ratio causes only a slight increase (about 19%) of the magnetic signal strength but a strong decrease (about 81%) of the electric signal strength. Using a homogeneous torso model, the dipole localization errors were, depending on the conductivity ratio, up to 16 mm. In conclusion, passive vortex currents might partially explain the differences between magnetocardiographic and electrocardiographic recordings observed both experimentally and clinically.
引用
收藏
页码:145 / 151
页数:7
相关论文
共 15 条
[1]   A SIMULATION OF CARDIAC ACTION CURRENTS HAVING CURL [J].
BARACH, JP .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1993, 40 (01) :49-58
[2]   Magnetocardiography and 32-lead potential mapping: Repolarization in normal subjects during pharmacologically induced stress [J].
Brockmeier, K ;
Schmitz, L ;
Chavez, JB ;
Burghoff, M ;
Koch, H ;
Zimmermann, R ;
Trahms, L .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 1997, 8 (06) :615-626
[3]   Dependence of the inverse solution accuracy in magnetocardiography on the boundary-element discretization [J].
Haueisen, J ;
Schreiber, J ;
Brauer, H ;
Knösche, TR .
IEEE TRANSACTIONS ON MAGNETICS, 2002, 38 (02) :1045-1048
[4]   MYOCARDIAL INFARCT SIZE AND LOCATION IN RELATION TO THE CORONARY VASCULAR BED AT RISK IN MAN [J].
LEE, JT ;
IDEKER, RE ;
REIMER, KA .
CIRCULATION, 1981, 64 (03) :526-534
[5]   Vortex shaped current sources in a physical torso phantom [J].
Liehr, M ;
Haueisen, J ;
Goernig, M ;
Seidel, P ;
Nenonen, J ;
Katila, T .
ANNALS OF BIOMEDICAL ENGINEERING, 2005, 33 (02) :240-247
[6]   The study of steady magnetic fields associated with primary and secondary ST shift in ischaemic rabbit hearts [J].
Monteiro, EC ;
DellaPenna, S ;
DiDonato, L ;
DiLuzio, S ;
Pasquarelli, A ;
Erne, SN ;
Romani, GL .
PHYSIOLOGICAL MEASUREMENT, 1997, 18 (03) :191-200
[7]   A SIMPLEX-METHOD FOR FUNCTION MINIMIZATION [J].
NELDER, JA ;
MEAD, R .
COMPUTER JOURNAL, 1965, 7 (04) :308-313
[8]   Current multipole expansion to estimate lateral extent of neuronal activity: A theoretical analysis [J].
Nolte, G ;
Curie, G .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2000, 47 (10) :1347-1355
[9]   MYOCARDIAL-ISCHEMIA AND INFARCTION - ANATOMIC AND BIOCHEMICAL SUBSTRATES FOR ISCHEMIC CELL-DEATH AND VENTRICULAR ARRHYTHMIAS [J].
REIMER, KA ;
IDEKER, RE .
HUMAN PATHOLOGY, 1987, 18 (05) :462-475
[10]   ELECTRICALLY SILENT MAGNETIC-FIELDS [J].
ROTH, BJ ;
WIKSWO, JP .
BIOPHYSICAL JOURNAL, 1986, 50 (04) :739-745