Developments of scintillator-based soft x-ray diagnostic in LHD with CsI:Tl and P47 scintillators

被引:6
|
作者
Bando, T. [1 ]
Ohdachi, S. [1 ,2 ]
Suzuki, Y. [1 ,2 ]
机构
[1] SOKENDAI Grad Univ Adv Studies, 322-6 Oroshi Cho, Toki, Gifu 5095292, Japan
[2] Natl Inst Fus Sci, 322-6 Oroshi Cho, Toki, Gifu 5095292, Japan
来源
REVIEW OF SCIENTIFIC INSTRUMENTS | 2016年 / 87卷 / 11期
关键词
D O I
10.1063/1.4960418
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Multi-channel soft x-ray (SX) diagnostic has been used in the large helical device (LHD) to research magnetohydrodynamic equilibria and activities. However, in the coming deuterium plasma experiments of LHD, it will be difficult to use semiconductor systems near LHD. Therefore, a new type of SX diagnostic, a scintillator-based type diagnostic, has been investigated in order to avoid damage from the radiation. A fiber optic plate coated by P47 scintillator will be used to detect SX emission. Scintillation light will be transferred by pure silica core optical fibers and detected by photomultiplier tubes. A vertically elongated section of LHD will be covered by a 13 ch. array. Effects from the Deuterium Deuterium neutrons can be negligible when the scintillator is covered by a Pb plate 4 cm in thickness to avoid gamma-rays. Published by AIP Publishing.
引用
收藏
页数:4
相关论文
共 50 条
  • [1] Development of a scintillator-based optical soft x-ray (OSXR) diagnostic system for KSTAR tokamak
    Lee, M. W.
    Kim, Junghee
    Kim, M. H.
    Lim, Soobin
    Kim, Jayhyun
    Sung, C.
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2023, 94 (05):
  • [2] Structured CsI(Tl) scintillators for x-ray imaging applications
    Radiation Monitoring Devices, Inc, Watertown, United States
    IEEE Trans Nucl Sci, 3 pt 1 (492-496):
  • [3] Structured CsI(Tl) scintillators for X-ray imaging applications
    Nagarkar, VV
    Gupta, TK
    Miller, SR
    Klugerman, Y
    Squillante, MR
    Entine, G
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1998, 45 (03) : 492 - 496
  • [4] Structured CsI(Tl) scintillators for X-ray imaging applications
    Nagarkar, VV
    Gupta, TK
    Miller, S
    Klugarman, Y
    Squillante, MR
    Entine, G
    1997 IEEE NUCLEAR SCIENCE SYMPOSIUM - CONFERENCE RECORD, VOLS 1 & 2, 1998, : 226 - 230
  • [5] A Pixelated CsI(Tl) Scintillator for CMOS-based X-ray Image Sensor
    Cha, Bo Kyung
    Kim, Byoung-Jik
    Cho, Gyuseong
    Jeon, Sung Chae
    Bae, Jun Hyung
    Chi, Yong Ki
    Lim, Gyu-Ho
    Kim, Young-Hee
    2006 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOL 1-6, 2006, : 1139 - 1143
  • [6] Portal imaging with a CsI(Tl) transparent scintillator x-ray detector
    Zeman, HD
    Samant, SS
    Lovhoiden, G
    Weinberg, B
    Sawant, A
    PHYSICS OF MEDICAL IMAGING, 1998, 3336 : 175 - 186
  • [7] A scintillator-based hard X-ray imaging telescope - CASTER
    Cherry, M. L.
    Blose, P. F.
    Case, G. L.
    Cravens, J. P.
    Guzik, T. G.
    Hurley, K. C.
    Isbert, J. B.
    Kippen, R. M.
    Macri, J. M.
    McConnell, M. L.
    Miller, R. S.
    Paciesas, W. S.
    Ryan, J. M.
    Schaefer, B. E.
    Stacy, J. G.
    Vestrand, W. T.
    Wefel, J. P.
    Welch, C. E.
    Proceedings of the 29th International Cosmic Ray Conference, Vol 5: OG 2.5, 2.6 & 2.7, 2005, : 407 - 410
  • [8] 3-D modeling of scintillator-based X-ray detectors
    Vieira da Rocha, Jose Gerardo
    Lanceros-Mendez, Senentxu
    IEEE SENSORS JOURNAL, 2006, 6 (05) : 1236 - 1242
  • [9] Scintillator-based flat-panel x-ray imaging detectors
    Granfors, Paul R.
    Albagli, Douglas
    JOURNAL OF THE SOCIETY FOR INFORMATION DISPLAY, 2009, 17 (06) : 535 - 542
  • [10] Emission properties of scintillator crystal CsI:Tl excited by X-ray and UV light
    Wu, Zheng-Long
    Wang, Jin-Mei
    Wang, Ya-Fang
    Chen, Luan
    Yang, Bai-Rui
    Faguang Xuebao/Chinese Journal of Luminescence, 2010, 31 (06): : 831 - 835