A clinical evaluation of the Maquet Quadrox-i Neonatal oxygenator with integrated arterial filter

被引:6
作者
Ginther, R. M., Jr.
Gorney, R.
Cruz, R.
机构
[1] Univ Texas SW Med Ctr Dallas, Div Pediat Cardiothorac Surg, Dallas, TX 75390 USA
[2] Childrens Med Ctr, Dallas, TX 75235 USA
来源
PERFUSION-UK | 2013年 / 28卷 / 03期
关键词
cardiopulmonary bypass; pediatric oxygenator; Quadrox- i neonatal oxygenator; integrated arterial filter; low prime oxygenator;
D O I
10.1177/0267659113475694
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
High-performance, low-prime-volume oxygenators for the pediatric patient population have become a growing market among manufacturers. In the summer of 2011, our institution clinically evaluated the performance of the newly released Maquet Quadrox-i Neonatal oxygenator with integrated arterial filter. The static priming volume, including the integrated arterial filter, is 40 ml and the maximum rated blood flow is 1.5 liters per minute (LPM). The device was used on seven pediatric patients, ranging from 3.2 to 14 kg, undergoing various congenital heart defect repairs. Data were collected to calculate gas transfer, trans-oxygenator pressure drop, and heat exchange performance. The mean cardiopulmonary bypass time was 85 minutes and the mean cross-clamp time was 56 minutes. The average oxygen transfer was 34.3 +/- 22.8 ml/O-2/min and increased with both blood flow and FiO(2). The average carbon dioxide transfer was 22.3 +/- 17.8 ml/min and increased with both blood flow and gas sweep to blood flow ratio. The average trans-oxygenator pressure drop per blood flow was 53.3 +/- 15.5 mmHg/L/min and increased with flow. The average heat exchanger performance factor was 47.6 +/- 11.6% and decreased with flow. The heat exchange performance factor at maximum observed clinical flow, 1.42 LPM, was 36.4%. During this evaluation, the Maquet Quadrox-i Neonatal oxygenator adequately performed within its operational flow in the clinical setting.
引用
收藏
页码:194 / 199
页数:6
相关论文
共 12 条
  • [1] Evaluation of the Quadrox-I neonatal oxygenator with an integrated arterial filter
    Salavitabar, Arash
    Qiu, Feng
    Kunselman, Allen
    Uendar, Akif
    PERFUSION-UK, 2010, 25 (06): : 409 - 415
  • [2] Evaluation of the Maquet Neonatal and Pediatric Quadrox I with an integrated arterial line filter during cardiopulmonary bypass
    Melchior, R. W.
    Schiavo, K.
    Frey, T.
    Rogers, D.
    Patel, J.
    Chelnik, K.
    Rosenthal, T.
    PERFUSION-UK, 2012, 27 (05): : 399 - 406
  • [3] Evaluation of Quadrox-i and Capiox FX neonatal oxygenators with integrated arterial filters in eliminating gaseous microemboli and retaining hemodynamic properties during simulated cardiopulmonary bypass
    Lin, J.
    Dogal, N. M.
    Mathis, R. K.
    Qiu, F.
    Kunselman, A.
    Uendar, A.
    PERFUSION-UK, 2012, 27 (03): : 235 - 243
  • [4] Clinical evaluation of the air removal characteristics of an oxygenator with integrated arterial filter in a minimized extracorporeal circuit
    Stehouwer, Marco C.
    Boers, Chris
    de Vroege, Roel
    Kelder, Johannes C.
    Yilmaz, Alaaddin
    Bruins, Peter
    INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2011, 34 (04) : 374 - 382
  • [5] Evaluation of air handling in a new generation neonatal oxygenator with integral arterial filter
    Gomez, D.
    Preston, T. J.
    Olshove, V. F.
    Phillips, A. B.
    Galantowicz, M. E.
    PERFUSION-UK, 2009, 24 (02): : 107 - 112
  • [6] Evaluation of Capiox RX25 and Quadrox-i Adult Hollow Fiber Membrane Oxygenators in a Simulated Cardiopulmonary Bypass Circuit
    Wang, Shigang
    Kunselman, Allen R.
    Undar, Akif
    ARTIFICIAL ORGANS, 2016, 40 (05) : E69 - E78
  • [7] Clinical evaluation of the air-handling properties of contemporary oxygenators with integrated arterial filter
    Stehouwer, Marco C.
    Legg, Kristina R.
    de Vroege, Roel
    Kelder, Johannes C.
    Hofman, Erik
    de Mol, Bastian A.
    Bruins, Peter
    PERFUSION-UK, 2017, 32 (02): : 118 - 125
  • [8] Evaluation of Capiox FX05 Oxygenator With an Integrated Arterial Filter on Trapping Gaseous Microemboli and Pressure Drop With Open and Closed Purge Line
    Qiu, Feng
    Peng, Sophia
    Kunselman, Allen
    Uendar, Akif
    ARTIFICIAL ORGANS, 2010, 34 (11) : 1053 - 1057
  • [9] Pre-clinical laboratory evaluation of the new 'AF' arterial line filter range
    Yarham, Gemma
    Mulholland, John
    PERFUSION-UK, 2010, 25 (04): : 267 - 276
  • [10] Clinical evaluation of emboli removal by integrated versus non-integrated arterial filters in new generation oxygenators
    Jabur, Ghazwan N. S.
    Sidhu, Karishma
    Willcox, Timothy W.
    Mitchell, Simon J.
    PERFUSION-UK, 2016, 31 (05): : 409 - 417