Laboratory evaluation of the Coasys® Plus C coagulation analyzer

被引:4
作者
de Bie, Prim [1 ]
Schornagel, Willem J. [1 ]
van den Dool, Erik-Jan [1 ]
Bakker, Bianca [1 ]
van Dam, Willem [1 ]
Heckman, Marion [1 ]
Agar, Cetin [1 ]
Sturk, Auguste [1 ]
Stroobants, An K. [1 ]
机构
[1] Univ Amsterdam, Acad Med Ctr, Lab Gen Clin Chem, NL-1105 AZ Amsterdam, Netherlands
关键词
Coasys (R) Plus C; Coagulation; ACL TOP ANALYZER; PERFORMANCE-CHARACTERISTICS; SYSMEX(R) CA-7000; D-DIMER;
D O I
10.1016/j.thromres.2013.02.004
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Introduction: The Coasys (R) Plus C (Behnk Elektronik, distributed by Roche Diagnostics) is a coagulation analyzer for small to midsize clinical chemistry laboratories. We performed a laboratory evaluation. Materials and Methods: After a familiarization period the dead volume, carry-over, capacity, within-assay reproducibility and imprecision were determined for the tests aPTT (STA APTT en STA Cephascreen), PT (STA Neoplastin Plus, STA Neoplastin R), INR (Hepato Quick), fibrinogen (STA Fibrinogen), antithrombin (Antithrombin III) and D-Dimer (Tina-quant D-Dimer Gen.2). A precision test and patient comparison with a STA-R Evolution (R) system were performed for aPTT (STA Cephascreen), PT (STA Neoplastin R), fibrinogen, antithrombin and D-Dimer. Results: No carry-over was detected. The analyzer performed on average 66 measurements per hour. Within-assay reproducibility (% with normal/abnormal/extreme result): STA APTT 2.4/1.7/2.7; aPTT Cephascreen 1.4/3.1/1.8; PT Neoplastin Plus 1.2/2.1/1.7; PT Neoplastin R 2.3/-/3.1; INR Hepato Quick 0.6/0.9/1.9; fibrinogen 4.8/3.0/4.5; antithrombin 1.3/4.1/3.8; D-Dimer 19.9/4.3/4.3. Total imprecision (% with control 1/control 2/human pooled plasma): STA APTT 3.1/1.4/2.1; APTT Cephascreen 2.3/2.6/2.0; PT Neoplastin Plus 3.3/1.3/4.1; PT Neoplastin R 3.7/4.0/3.5; INR Hepato Quick 1.5/4.3/1.4; fibrinogen 3.4/5.8/8.1; antithrombin 3.0/6.4/2.4; D-Dimer 4.6/2.2/30.3. The correlation between the Coasys (R) Plus C and the STA-R Evolution (R) was good for aPTT, PT, antithrombin and fibrinogen. Some differences were observed for extreme deviant results for fibrinogen. For the analysis of D-Dimer, no sufficient correlation was found between the two analyzers. Conclusions: The Coasys (R) Plus C analyzer is fast, easy to handle and safe for the personnel. Its analytical performance makes it suitable for use in a clinical chemistry laboratory. (C) 2013 Elsevier Ltd. All rights reserved.
引用
收藏
页码:357 / 362
页数:6
相关论文
共 8 条
[1]   Evaluation and performance characteristics of the automated coagulation analyzer ACL TOP [J].
Appert-Flory, Anny ;
Fischer, Florence ;
Jambou, Didier ;
Toulon, Pierre .
THROMBOSIS RESEARCH, 2007, 120 (05) :733-743
[2]   Evaluation of the automated coagulation analyzer Sysmex® CA-7000 [J].
Dorn-Beineke, A ;
Dempfle, CE ;
Bertsch, T ;
Wisser, H .
THROMBOSIS RESEARCH, 2005, 116 (02) :171-179
[3]  
Ermens A A, 2000, Clin Lab, V46, P463
[4]   Evaluation of the automated coagulation analyzer Sysmex® CA-7000 [J].
Fischer, F ;
Appert-Flory, A ;
Jambou, D ;
Toulon, P .
THROMBOSIS RESEARCH, 2006, 117 (06) :721-729
[5]   Evaluation and performance characteristics of the coagulation system: ACL TOP analyzer - HemosIL reagents [J].
Milos, M. ;
Herak, D. ;
Kuric, L. ;
Horvat, I. ;
Zadro, R. .
INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2009, 31 (01) :26-35
[6]  
Ricós C, 1999, SCAND J CLIN LAB INV, V59, P491
[7]   Performance of the automated and rapid HemosIL D-Dimer HS on the ACL TOP analyzer [J].
Salvagno, Gian L. ;
Lippi, Giuseppe ;
Montagnana, Martina ;
Poli, Giovanni ;
Giavarina, Davide ;
Manzato, Franco ;
Guidi, Gian C. .
BLOOD COAGULATION & FIBRINOLYSIS, 2008, 19 (08) :817-821
[8]   Performance of the automated and rapid STA Liatest D-dimer on the STA-R analyzer [J].
Waser, G ;
Kathriner, S ;
Wuillemin, WA .
THROMBOSIS RESEARCH, 2005, 116 (02) :165-170