Precision of GE Lunar iDXA for the Measurement of Total and Regional Body Composition in Nonobese Adults

被引:91
作者
Rothney, Megan P. [1 ]
Martin, Francois-Pierre [2 ]
Xia, Yi [3 ]
Beaumont, Maurice [2 ]
Davis, Cynthia [1 ]
Ergun, David [3 ]
Fay, Laurent [2 ]
Ginty, Fiona [1 ]
Kochhar, Sunil [2 ]
Wacker, Wynn [3 ]
Rezzi, Serge [2 ]
机构
[1] GE Global Res Ctr, Computat Biol & Biostat Lab, Niskayuna, NY 12309 USA
[2] Nestec Ltd, Nestle Res Ctr, CH-1000 Lausanne, Switzerland
[3] GE Healthcare, Madison, WI USA
关键词
Body composition; DXA; precision; X-RAY ABSORPTIOMETRY; IMPAIRED GLUCOSE-TOLERANCE; LIFE-STYLE INTERVENTION; FAT DISTRIBUTION; WOMEN; PREVENTION; DISEASE; RISK; MEN; OBESITY;
D O I
10.1016/j.jocd.2012.02.009
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Dual-energy X-ray absorptiometry (DXA) is a well-accepted technique for measuring body composition. Knowledge of measurement precision is critical for monitoring of changes in bone mineral content (BMC), and fat and lean masses. The purpose of this study was to characterize in vivo precision of total body and regional body composition parameters using the GE Lunar iDXA (GE Healthcare Lunar, Madison, WI) system in a sample of nonobese subjects. We also evaluated the difference between expert and automatic region-of-interest (ROI) analysis on body composition precision. To this end, 2 total body scans were performed on each subject with repositioning between scans. Total body precision for BMC, fat and lean mass were 0.5%, 1.0%, and 0.5% coefficient of variation (CV), respectively. Regional body composition precision error was less than 2.5% CV for all regions except arms. Precision error was higher for the arms (CV: BMC 1.5%; fat mass 2.8%; lean mass 1.6%), likely owing to the placement of arms relative to torso leading to differences in ROI. There was a significant correlation between auto ROT and expert ROT (r > 0.99). Small, but statistically significant differences were found between auto and manual ROT. Differences were small in total body, leg, trunk, and android and gynoid regions (0.004-2.8%), but larger in arm region (3.0-6.3%). Total body and regional precision for iDXA are small and it is suggested that iDXA may be useful for monitoring changes in body composition during longitudinal trials.
引用
收藏
页码:399 / 404
页数:6
相关论文
共 51 条
[1]   Chronic diseases 1 - The burden and costs of chronic diseases in low-income and middle-income countries [J].
Abegunde, Dele O. ;
Mathers, Colin D. ;
Adam, Taghreed ;
Ortegon, Monica ;
Strong, Kathleen .
LANCET, 2007, 370 (9603) :1929-1938
[2]   Clinical applications of body composition measurements using DXA [J].
Albanese, CV ;
Diessel, E ;
Genant, HK .
JOURNAL OF CLINICAL DENSITOMETRY, 2003, 6 (02) :75-85
[4]  
Andreoli A, 2009, RADIOL MED, V114, P286, DOI 10.1007/s11547-009-0369-7
[5]  
[Anonymous], 2008, SOURC EFF ION RAD
[6]   Precision assessment and radiation safety for dual-energy X-ray absorptiometry [J].
Baim, S ;
Wilson, CR ;
Lewiecki, EM ;
Luckey, MM ;
Downs, RW ;
Lentle, BC .
JOURNAL OF CLINICAL DENSITOMETRY, 2005, 8 (04) :371-378
[7]   Body fat distribution and risk of coronary heart disease in men and women in the European prospective investigation into cancer and nutrition in Norfolk cohort - A population-based prospective study [J].
Canoy, Dexter ;
Boekholdt, S. Matthijs ;
Wareham, Nicholas ;
Luben, Robert ;
Welch, Ailsa ;
Bingham, Sheila ;
Buchan, Iain ;
Day, Nicholas ;
Khaw, Kay-Tee .
CIRCULATION, 2007, 116 (25) :2933-2943
[8]  
Chen R, 2005, J BONE MINER RES, V20, pS220
[9]   Assessment and prevalence of obesity - Application of new methods to a major problem [J].
Chumlea, WC ;
Guo, SS .
ENDOCRINE, 2000, 13 (02) :135-142
[10]   Anthropometric and body composition assessment in dialysis patients [J].
Chumlea, WC .
SEMINARS IN DIALYSIS, 2004, 17 (06) :466-470