Platinum subdural grid: MR imaging compatibility testing

被引:7
|
作者
Kanal, E
Meltzer, CC
Adelson, PD
Scheuer, MP
机构
[1] Univ Pittsburgh, Med Ctr, Dept Radiol, Pittsburgh, PA 15213 USA
[2] Univ Pittsburgh, Med Ctr, Dept Psychiat, Pittsburgh, PA 15213 USA
[3] Univ Pittsburgh, Med Ctr, Dept Neurol, Pittsburgh, PA 15213 USA
[4] Childrens Hosp Pittsburgh, Dept Neurosurg, Pittsburgh, PA 15213 USA
关键词
head; MR; magnetic resonance (MR); physics; safety; radiology and radiologists; iatrogenic injury;
D O I
10.1148/radiology.211.3.r99jn01886
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A platinum subdural grid electrode (used for seizure monitoring) was tested for potential heating during magnetic resonance (MR) imaging at 1.5 and 3.0 T. Fast spin-echo imaging with high radio-frequency transmitted power was performed while temperature measurements for the grid phantom were recorded by means of fiberoptic thermometry. Temperature variations remained within 1.0 degrees C. MR imaging examination in patients in whom such grid electrodes are implanted is expected to result in clinically insignificant temperature changes in the vicinity of the grid electrode or its leads.
引用
收藏
页码:886 / 888
页数:3
相关论文
共 50 条
  • [1] Testing MR safety and compatibility
    Schaefers, Gregor
    IEEE ENGINEERING IN MEDICINE AND BIOLOGY MAGAZINE, 2008, 27 (03): : 23 - 27
  • [2] SUBDURAL AND EPIDURAL EMPYEMAS - MR IMAGING
    WEINGARTEN, K
    ZIMMERMAN, RD
    BECKER, RD
    HEIER, LA
    HAIMES, AB
    DECK, MDF
    AMERICAN JOURNAL OF ROENTGENOLOGY, 1989, 152 (03) : 615 - 621
  • [3] SUBDURAL AND EPIDURAL EMPYEMAS - MR IMAGING
    WEINGARTEN, K
    ZIMMERMAN, RD
    BECKER, RD
    HEIER, LA
    HAIMES, AB
    DECK, MDF
    AMERICAN JOURNAL OF NEURORADIOLOGY, 1989, 10 (01) : 81 - 87
  • [4] THE EFFICACY OF MR IMAGING IN SUBDURAL EMPYEMA
    KOMORI, H
    TAKAGISHI, T
    OTAKI, E
    SASAKI, H
    MATSUISHI, T
    ABE, T
    KOJIMA, K
    MORITAKA, K
    BRAIN & DEVELOPMENT, 1992, 14 (02): : 123 - 125
  • [5] MR imaging of implanted depth and subdural electrodes: is it safe?
    Davis, LM
    Spencer, DD
    Spencer, SS
    Bronen, RA
    EPILEPSY RESEARCH, 1999, 35 (02) : 95 - 98
  • [6] Subdural grid implantation for intracranial EEG recording: CT and MR appearance
    Silberbusch, MA
    Rothman, MI
    Bergey, GK
    Zoarski, GH
    Zagardo, MT
    AMERICAN JOURNAL OF NEURORADIOLOGY, 1998, 19 (06) : 1089 - 1093
  • [7] Testing methods for MR safety and compatibility of medical devices
    Schaefers, G
    Melzer, A
    MINIMALLY INVASIVE THERAPY & ALLIED TECHNOLOGIES, 2006, 15 (02) : 71 - 75
  • [8] Evaluation of MR compatibility of a collimator material for simultaneous SPECT/MR imaging
    Chang, Yan
    Chen, Qiaoyan
    Rittenbach, Andrew
    Xu, Jingyan
    Sun, Mingshan
    Tsui, Benjamin
    Yang, Xiaodong
    JOURNAL OF NUCLEAR MEDICINE, 2016, 57
  • [9] Visualization of subdural strip and grid electrodes using curvilinear reformatting of 3D MR imaging data sets
    Schulze-Bonhage, AHJ
    Huppertz, HJ
    Comeau, RM
    Honegger, JB
    Spreer, JM
    Zentner, JK
    AMERICAN JOURNAL OF NEURORADIOLOGY, 2002, 23 (03) : 400 - 403
  • [10] Spetzler titanium aneurysm clips: Compatibility at MR imaging
    Shellock, FG
    Shellock, VJ
    RADIOLOGY, 1998, 206 (03) : 838 - 841