Assessment of a technique for 2D-3D registration of cerebral intra-arterial angiography

被引:25
作者
Byrne, JV
Colominas, C
Hipwell, J
Cox, T
Noble, JA
Penney, GP
Hawkes, DJ
机构
[1] Univ Oxford, Radcliffe Infirm, Dept Neuroradiol, Nuffield Dept Surg, Oxford OX2 6HE, England
[2] Kings Coll London, Guys Hosp, Div Imaging Sci, London SE1 9RT, England
[3] Natl Hosp Neurol & Neurosurg, London WC1 3BG, England
[4] Univ Oxford, Dept Engn Sci, Oxford OX1 3PJ, England
关键词
D O I
10.1259/bjr/27339681
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
This study assesses the ability of a computer algorithm to perform automated 2D-3D registrations of digitally subtracted cerebral angiograms. The technique was tested on clinical studies of five patients with intracranial aneurysms. The automated procedure was compared against a gold standard manual registration, and achieved a mean registration accuracy of 1.3 mm (SD 0.6 mm). Two registration strategies were tested using coarse (128 x 128 pixel) or fine (256 x 256 pixel) images. The mean registration errors proved similar but registration of the lower resolution images was 3 times quicker (mean registration times 33 s, SD 13 s for low and 150 s SD 48 s for high resolution images). The automated techniques were considerably faster than manual registrations but achieved similar accuracy. The technique has several potential uses but is particularly applicable to endovascular treatment techniques.
引用
收藏
页码:123 / 128
页数:6
相关论文
共 13 条
  • [1] Anxionnat R, 1998, J NEURORADIOLOGY, V25, P251
  • [2] Intracranial aneurysms: Clinical value of 3D digital subtraction angiography in the therapeutic decision and endovascular treatment
    Anxionnat, R
    Bracard, S
    Ducrocq, X
    Trousset, Y
    Launay, L
    Kerrien, E
    Braun, M
    Vaillant, R
    Scomazzoni, F
    Lebedinsky, A
    Picard, L
    [J]. RADIOLOGY, 2001, 218 (03) : 799 - 808
  • [3] DECOMPOSITION OF TRANSFORMATION-MATRICES FOR ROBOT VISION
    GANAPATHY, S
    [J]. PATTERN RECOGNITION LETTERS, 1984, 2 (06) : 401 - 412
  • [4] Analysis of cerebrovascular diseases by a new 3-dimensional computerised X-ray angiography system
    Heautot, JF
    Chabert, E
    Gandon, Y
    Croci, S
    Romeas, R
    Campagnolo, R
    Chereul, B
    Scarabin, JM
    Carsin, M
    [J]. NEURORADIOLOGY, 1998, 40 (04) : 203 - 209
  • [5] HIPWELL JH, 2002, P MICCAI, V2489, P501
  • [6] HOFF DJ, 1994, AM J NEURORADIOL, V15, P1945
  • [7] Masutani Y, 1997, LECT NOTES COMPUT SC, V1205, P427, DOI 10.1007/BFb0029265
  • [8] MCLAUGHLIN RA, 2002, P MICCAI, V2489, P517
  • [9] International Subarachnoid Aneurysm Trial (ISAT) of neurosurgical clipping versus endovascular coiling in 2143 patients with ruptured intracranial aneurysms: a randomised trial
    Molyneux, A
    Kerr, R
    Stratton, I
    Sandercock, P
    Clarke, M
    Shrimpton, J
    Holman, R
    [J]. LANCET, 2002, 360 (9342) : 1267 - 1274
  • [10] A comparison of similarity measures for use in 2-D-3-D medical image registration
    Penney, GP
    Weese, J
    Little, JA
    Desmedt, P
    Hill, DLG
    Hawkes, DJ
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 1998, 17 (04) : 586 - 595