Volume measurement of intracranial aneurysms from 3D rotational angiography:: Improvement of accuracy by gradient edge detection

被引:0
作者
Bescós, JO
Slob, MJ
Slump, CH
Sluzewski, M
van Rooij, WJ
机构
[1] St Elizabeth Hosp, Dept Radiol, NL-5022 GC Tilburg, Netherlands
[2] Twente Univ Technol, Dept Elect Engn Math & Comp Sci, NL-7500 AE Enschede, Netherlands
关键词
D O I
暂无
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Manual volume measurement of intracranial aneurysms from 3D rotational angiography varies on different threshold settings and, therefore, is operator-dependent. We developed and validated a method based on automatic gradient edge detection that is independent on threshold settings and provides an accurate and reproducible volume measurement. This method was compared with manual volume calculation in 13 aneurysm phantoms, and the results were significantly more accurate.
引用
收藏
页码:2569 / 2572
页数:4
相关论文
共 8 条
[1]   COMPARING METHODS OF MEASUREMENT - WHY PLOTTING DIFFERENCE AGAINST STANDARD METHOD IS MISLEADING [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1995, 346 (8982) :1085-1087
[3]  
Cloft HJ, 2004, AM J NEURORADIOL, V25, P60
[4]   Endovascular occlusion of intracranial aneurysms with Guglielmi detachable coils: Correlation between coil packing density and coil compaction [J].
Kawanabe, Y ;
Sadato, A ;
Taki, W ;
Hashimoto, N .
ACTA NEUROCHIRURGICA, 2001, 143 (05) :451-455
[5]  
Piotin M, 2003, AM J NEURORADIOL, V24, P1446
[6]  
Slob MJ, 2005, AM J NEURORADIOL, V26, P901
[7]   Relation between aneurysm volume, packing, and compaction in 124 cerebral aneurysms treated with coils [J].
Sluzewski, M ;
van Rooij, WJ ;
Slob, MJ ;
Bescós, JO ;
Slump, CH ;
Wijnalda, D .
RADIOLOGY, 2004, 231 (03) :653-658
[8]  
Tamatani S, 2002, AM J NEURORADIOL, V23, P762