Robust motion correction for respiratory gated PET/CT using weighted averaging

被引:0
作者
Thielemans, Kris [1 ]
Gopalakrishnan, Girish [2 ]
Roy, Arunabha [2 ]
Srikrishnan, V [2 ]
Thiruvenkadam, Sheshadri [2 ]
Wollenweber, Scott D. [3 ]
Manjeshwar, Ravindra M. [4 ]
机构
[1] Hammersmith Hosp, Hammersmith Imanet Ltd, GE Healthcare, Du Cane Rd, London W12 0NN, England
[2] GE Global Res, Bangalore, Karnataka, India
[3] GE Healthcare, Waukesha, WI 53188 USA
[4] GE Global Res, Niskayuna, NY 12309 USA
来源
2011 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC) | 2011年
关键词
POSITRON-EMISSION-TOMOGRAPHY; ARTIFACTS;
D O I
暂无
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Movement degrades image quality in PET/CT. A common strategy is to gate the PET data, reconstruct the images, register each image to a reference gate, and average the registered images (Reconstruct, Registered, Average or RRA). Previously, we have shown that this technique can provide more quantitatively accurate PET images compared to the ungated case, while using all the data means that RRA images have better counting statistics than single gates. However, in some instances, e. g. large patient motion, low count statistics acquisitions and/or mismatch between PET and CT position, the registration might fail for a particular gate in some region of the image, resulting in artifacts in the final motion corrected image. We have developed techniques to automatically detect such failures for every image voxel based on an analysis of the motion vectors and the image values. We propose to use a weighted average, where the weights are gate and voxel dependent, and are determined based on estimated registration quality. We illustrate this idea using respiratory gated PET data with PET-PET registration using the level-sets non-rigid registration algorithm. We applied this technique to respiratory gated phantom and patient data from GE Discovery PET/CT scanners. The results demonstrate that the method reliably mitigates any image artifacts and is robust to large lesion displacements and low count density PET images.
引用
收藏
页码:2964 / 2967
页数:4
相关论文
共 12 条
  • [1] [Anonymous], 2005, The ITK Software Guide
  • [2] Detection and compensation of organ/lesion motion using 4D-PET/CT respiratory gated acquisition techniques
    Bettinardi, Valentino
    Picchio, Maria
    Di Muzio, Nadia
    Gianolli, Luigi
    Gilardi, Maria Carla
    Messa, Cristina
    [J]. RADIOTHERAPY AND ONCOLOGY, 2010, 96 (03) : 311 - 316
  • [3] Kinahan P, 2006, I S BIOMED IMAGING, P1104
  • [4] Klein G. J., 1996, 1996 IEEE Nuclear Science Symposium Conference Record (Cat. No.96CH35974), P1339, DOI 10.1109/NSSMIC.1996.591692
  • [6] Effect of respiratory gating on reducing lung motion artifacts in PET imaging of lung cancer
    Nehmeh, SA
    Erdi, YE
    Ling, CC
    Rosenzweig, KE
    Squire, OD
    Braban, LE
    Ford, E
    Sidhu, K
    Mageras, GS
    Larson, SM
    Humm, JL
    [J]. MEDICAL PHYSICS, 2002, 29 (03) : 366 - 371
  • [7] Respiratory motion in positron emission tomography/computed tomography: A review
    Nehmeh, Sadek A.
    Erdi, Yusuf E.
    [J]. SEMINARS IN NUCLEAR MEDICINE, 2008, 38 (03) : 167 - 176
  • [8] Motion correction of PET images using multiple acquisition frames
    Picard, Y
    Thompson, CJ
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 1997, 16 (02) : 137 - 144
  • [9] Reducing respiratory motion artifacts in positron emission tomography through retrospective stacking
    Thorndyke, Brian
    Schreibmann, Eduard
    Koong, Albert
    Xing, Lei
    [J]. MEDICAL PHYSICS, 2006, 33 (07) : 2632 - 2641
  • [10] Image registration via level-set motion: Applications to atlas-based segmentation
    Vemuri, BC
    Ye, J
    Chen, Y
    Leonard, CM
    [J]. MEDICAL IMAGE ANALYSIS, 2003, 7 (01) : 1 - 20