Optimization of the Reconstruction Interval in Neurovascular 4D-CTA Imaging A Technical Note

被引:8
作者
Hoogenboom, T. C. H. [1 ]
Van Beurden, R. M. J. [1 ]
Van Teylingen, B. [1 ]
Schenk, B. [2 ]
Willems, P. W. A. [2 ]
机构
[1] Inholland Univ Appl Sci, Div Med Technol, Haarlem, Netherlands
[2] Leiden Univ, Med Ctr, Dept Radiol, NL-2300 RC Leiden, Netherlands
关键词
multi detector row CT; 4D-CTA; cerebral; vascular;
D O I
10.1177/159101991201800402
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Time resolved whole brain CT angiography (4D-CTA) is a novel imaging technology providing information regarding blood flow. One of the factors that influence the diagnostic value of this examination is the temporal resolution, which is affected by the gantry rotation speed during acquisition and the reconstruction interval during post-processing. Post-processing determines the time spacing between two reconstructed volumes and, unlike rotation speed, does not affect radiation burden. The data sets of six patients who underwent a cranial 4D-CTA were used for this study. Raw data was acquired using a 320-slice scanner with a rotation speed of 2 Hz. The arterial to venous passage of an intravenous contrast bolus was captured during a 15 s continuous scan. The raw data was reconstructed using four different reconstruction-intervals: 0.2, 0.3, 0.5 and 1.0 s. The results were rated by two observers using a standardized score sheet. The appearance of each lesion was rated correctly in all readings. Scoring for quality of temporal resolution revealed a stepwise improvement from the 1.0 s interval to the 0.3 s interval, while no discernable improvement was noted between the 0.3 s and 0.2 s interval. An increase in temporal resolution may improve the diagnostic quality of cranial 4D-CTA. Using a rotation speed of 0.5 s, the optimal reconstruction interval appears to be 0.3 s, beyond which, changes can no longer be discerned.
引用
收藏
页码:377 / 379
页数:3
相关论文
共 8 条
[1]   A PROPOSED CLASSIFICATION FOR SPINAL AND CRANIAL DURAL ARTERIOVENOUS FISTULOUS MALFORMATIONS AND IMPLICATIONS FOR TREATMENT [J].
BORDEN, JA ;
WU, JK ;
SHUCART, WA .
JOURNAL OF NEUROSURGERY, 1995, 82 (02) :166-179
[2]   Dynamic 320-Section CT Angiography in Cranial Arteriovenous Shunting Lesions [J].
Brouwer, P. A. ;
Bosman, T. ;
van Walderveen, M. A. A. ;
Krings, T. ;
Leroux, A. A. ;
Willems, P. W. A. .
AMERICAN JOURNAL OF NEURORADIOLOGY, 2010, 31 (04) :767-770
[3]   An investigation of temporal resolution parameters in cine-mode four-dimensional computed tomography acquisition [J].
Mutaf, Yildirim D. ;
Brinkmann, Debra H. .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2008, 9 (04) :172-180
[4]  
Siebert E, 2009, BRIT J RADIOL, V82, P561, DOI 10.1259/bjr/27721218
[5]  
Siebert E, 2010, NEUROL SCI, V31, P1
[6]   A PROPOSED GRADING SYSTEM FOR ARTERIOVENOUS-MALFORMATIONS [J].
SPETZLER, RF ;
MARTIN, NA .
JOURNAL OF NEUROSURGERY, 1986, 65 (04) :476-483
[7]   Detection and Classification of Cranial Dural Arteriovenous Fistulas Using 4D-CT Angiography: Initial Experience [J].
Willems, P. W. A. ;
Brouwer, P. A. ;
Barfett, J. J. ;
terBrugge, K. G. ;
Krings, T. .
AMERICAN JOURNAL OF NEURORADIOLOGY, 2011, 32 (01) :49-53
[8]   The use of 4D-CTA in the diagnostic work-up of brain arteriovenous malformations [J].
Willems, Peter W. A. ;
Taeshineetanakul, Patamintita ;
Schenk, Barry ;
Brouwer, Patrick A. ;
Terbrugge, Karel G. ;
Krings, Timo .
NEURORADIOLOGY, 2012, 54 (02) :123-131