Automated Titration of Vasopressor Infusion Using a Closed-loop Controller In Vivo Feasibility Study Using a Swine Model

被引:38
作者
Joosten, Alexandre [1 ,3 ]
Delaporte, Amelie [1 ]
Alexander, Brenton [4 ]
Su, Fuhong [2 ]
Creteur, Jacques [2 ]
Vincent, Jean-Louis [2 ]
Cannesson, Maxime [5 ]
Rinehart, Joseph [6 ]
机构
[1] Univ Libre Bruxelles, Erasme Univ Hosp, Dept Anesthesiol, Brussels, Belgium
[2] Univ Libre Bruxelles, Erasme Univ Hosp, Dept Intens Care, Brussels, Belgium
[3] Univ Paris Saclay, Univ Paris Sud, Dept Anesthesiol & Intens Care, Hop Bicetre,Hop Univ Paris Sud,AP HP, Le Kremlin Bicetre, France
[4] Univ Calif San Diego, Dept Anesthesiol, San Diego, CA 92103 USA
[5] David Geffen Sch Med, Dept Anesthesiol & Perioperat Med, Los Angeles, CA USA
[6] Univ Calif Irvine, Dept Anesthesiol & Perioperat Care, Orange, CA 92668 USA
关键词
DIRECTED FLUID THERAPY; BLOOD-PRESSURE MANAGEMENT; MEAN ARTERIAL-PRESSURE; INTRAOPERATIVE HYPOTENSION; NONCARDIAC SURGERY; MYOCARDIAL INJURY; CARDIAC-OUTPUT; SEVERE SEPSIS; ACUTE KIDNEY; ASSOCIATION;
D O I
10.1097/ALN.0000000000002581
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Background: Multiple studies have reported associations between intraoperative hypotension and adverse postoperative complications. One of the most common interventions in the management of hypotension is vasopressor administration. This approach requires careful and frequent vasopressor boluses and/or multiple adjustments of an infusion. The authors recently developed a closed-loop controller that titrates vasopressors to maintain mean arterial pressure (MAP) within set limits. Here, the authors assessed the feasibility and overall performance of this system in a swine model. The authors hypothesized that the closed-loop controller would be able to maintain MAP at a steady, predefined target level of 80 mmHg for greater than 85% of the time. Methods: The authors randomized 14 healthy anesthetized pigs either to a control group or a closed-loop group. Using infusions of sodium nitroprusside at doses between 65 and 130 mu g/min, we induced four normovolemic hypotensive challenges of 30 min each. In the control group, nothing was done to correct hypotension. In the closed-loop group, the system automatically titrated norepinephrine doses to achieve a predetermined MAP of 80 mmHg. The primary objective was study time spent within +/- 5 mmHg of the MAP target. Secondary objectives were performance error, median performance error, median absolute performance error, wobble, and divergence. Results: The controller maintained MAP within +/- 5 mmHg of the target for 98 +/- 1% (mean +/- SD) of the time. In the control group, the MAP was 80 +/- 5 mmHg for 14.0 +/- 2.8% of the time (P < 0.0001). The MAP in the closed-loop group was above the target range for 1.2 +/- 1.2% and below it for 0.5 +/- 0.9% of the time. Performance error, median performance error, median absolute performance error, wobble, and divergence were all optimal. Conclusions: In this experimental model of induced normovolemic hypotensive episodes in pigs, the automated controller titrated norepinephrine infusion to correct hypotension and keep MAP within +/- 5 mmHg of target for 98% of management time.
引用
收藏
页码:394 / 403
页数:10
相关论文
共 37 条
[1]   Norepinephrine plus dobutamine versus epinephrine alone for management of septic shock:: a randomised trial [J].
Annane, Djillali ;
Vignon, Philippe ;
Renault, Alain ;
Bollaert, Pierre-Edouard ;
Charpentier, Claire ;
Martin, Claude ;
Troche, Gilles ;
Ricard, Jean-Damien ;
Nitenberg, Gerard ;
Papazian, Laurent ;
Azoulay, Elie ;
Bellissant, Eric .
LANCET, 2007, 370 (9588) :676-684
[2]  
[Anonymous], 2017, R LANG ENV STAT COMP
[3]   High versus Low Blood-Pressure Target in Patients with Septic Shock [J].
Asfar, Pierre ;
Meziani, Ferhat ;
Hamel, Jean-Francois ;
Grelon, Fabien ;
Megarbane, Bruno ;
Anguel, Nadia ;
Mira, Jean-Paul ;
Dequin, Pierre-Francois ;
Gergaud, Soizic ;
Weiss, Nicolas ;
Legay, Francois ;
Le Tulzo, Yves ;
Conrad, Marie ;
Robert, Rene ;
Gonzalez, Frederic ;
Guitton, Christophe ;
Tamion, Fabienne ;
Tonnelier, Jean-Marie ;
Guezennec, Pierre ;
Van der Linden, Thierry ;
Vieillard-Baron, Antoine ;
Mariotte, Eric ;
Pradel, Gael ;
Lesieur, Olivier ;
Ricard, Jean-Damien ;
Herve, Fabien ;
du Cheyron, Damien ;
Guerin, Claude ;
Mercat, Alain ;
Teboul, Jean-Louis ;
Radermacher, Peter .
NEW ENGLAND JOURNAL OF MEDICINE, 2014, 370 (17) :1583-1593
[4]   Administration of the nitric oxide synthase inhibitor NG-Methyl-L-arginine hydrochloride (546C88) by intravenous infusion for up to 72 hours can promote the resolution of shock in patients with severe sepsis:: Results of a randomized, double-blind, placebo-controlled multicenter study (study no. 144-002) [J].
Bakker, J ;
Grover, R ;
McLuckie, A ;
Holzapfel, L ;
Andersson, J ;
Lodato, R ;
Watson, D ;
Grossman, S ;
Donaldson, J ;
Takala, J .
CRITICAL CARE MEDICINE, 2004, 32 (01) :1-12
[5]   Intraoperative Hypotension and Perioperative Ischemic Stroke after General Surgery A Nested Case-control Study [J].
Bijker, Jilles B. ;
Persoon, Suzanne ;
Peelen, Linda M. ;
Moons, Karel G. M. ;
Kalkman, Cor J. ;
Kappelle, L. Jaap ;
van Klei, Wilton A. .
ANESTHESIOLOGY, 2012, 116 (03) :658-664
[6]   Effect of Individualized vs Standard Blood Pressure Management Strategies on Postoperative Organ Dysfunction Among High-Risk Patients Undergoing Major Surgery A Randomized Clinical Trial [J].
Futier, Emmanuel ;
Lefrant, Jean-Yves ;
Guinot, Pierre-Gregoire ;
Godet, Thomas ;
Lorne, Emmanuel ;
Cuvillon, Philippe ;
Bertran, Sebastien ;
Leone, Marc ;
Pastene, Bruno ;
Piriou, Vincent ;
Molliex, Serge ;
Albanese, Jacques ;
Julia, Jean-Michel ;
Tavernier, Benoit ;
Imhoff, Etienne ;
Bazin, Jean-Etienne ;
Constantin, Jean-Michel ;
Pereira, Bruno ;
Jaber, Samir .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 2017, 318 (14) :1346-1357
[7]   Impact of intraoperative hypotension and blood pressure fluctuations on early postoperative delirium after non-cardiac surgery [J].
Hirsch, J. ;
DePalma, G. ;
Tsai, T. T. ;
Sands, L. P. ;
Leung, J. M. .
BRITISH JOURNAL OF ANAESTHESIA, 2015, 115 (03) :418-426
[8]   The association between lactate, mean arterial pressure, central venous oxygen saturation and peripheral temperature and mortality in severe sepsis: a retrospective cohort analysis [J].
Houwink, Aletta P. I. ;
Rijkenberg, Saskia ;
Bosman, Rob J. ;
van der Voort, Peter H. J. .
CRITICAL CARE, 2016, 20
[9]   Goal-Directed fluid therapy with closed-loop assistance during moderate risk surgery using noninvasive cardiac output monitoring: A pilot study [J].
Joosten, A. ;
Huynh, T. ;
Suehiro, K. ;
Canales, C. ;
Cannesson, M. ;
Rinehart, J. .
BRITISH JOURNAL OF ANAESTHESIA, 2015, 114 (06) :886-892
[10]   Crystalloid versus Colloid for Intraoperative Goal-directed Fluid Therapy Using a Closed-loop System A Randomized, Double-blinded, Controlled Trial in Major Abdominal Surgery [J].
Joosten, Alexandre ;
Delaporte, Amelie ;
Ickx, Brigitte ;
Touihri, Karim ;
Stany, Ida ;
Barvais, Luc ;
Van Obbergh, Luc ;
Loi, Patricia ;
Rinehart, Joseph ;
Cannesson, Maxime ;
Van der Linden, Philippe .
ANESTHESIOLOGY, 2018, 128 (01) :55-66