摘要
Diagnostic laboratories today often operate according to standard quality management procedures such as ISO/IEC 17025. This requires that only validated methods are used. Validation procedures help to document that a particular protocol used by the accredited laboratory has a guaranteed performance in that particular laboratory. Several study designs exist for validation procedures. Computer programmes are available to help with the statistical analysis of validation results. The agreement beyond chance of results obtained in the protocol that is to be validated can be compared to those achieved in an already established test (agreement). For a method that is used under routine conditions or for epidemiological studies, it is necessary to assess the diagnostic sensitivity and diagnostic specificity of the technique. These parameters can be estimated by comparing the method that needs to be validated with an existing reliable method ('gold standard'). This is done by testing a standard set of well-documented samples using both techniques in parallel. Approaches using Bayes' theorem are used to perform gold standard-free validations. Many PCR-based methods are characterised by an excellent analytical sensitivity and are thus good candidates for diagnostic tools of the required diagnostic sensitivity. However, the high level of analytical sensitivity can also make molecular techniques susceptible to cross-contamination and carry-over problems leading to false-positive results. Moreover, the presence of inhibitors can cause false-negative results. After an initial validation, test performance needs to be continuously monitored, e.g. by using combined Shewhart-CUSUM control routines, and test results compared to those obtained by other laboratories (proficiency testing). (c) 2005 Elsevier B.V. All rights reserved.
引用
收藏
页码:91 / 98
页数:8
相关论文
共 44 条
[1]  
[Anonymous], 17000 ISOIEC
[2]  
[Anonymous], VET EPIDEMIOLOGY
[3]  
[Anonymous], 17025 ISOIEC
[4]   The use of mimics as internal standards to avoid false negatives in diagnostic PCR [J].
BallagiPordany, A ;
Belak, S .
MOLECULAR AND CELLULAR PROBES, 1996, 10 (03) :159-164
[5]  
Bayes M., 1763, PHILOS T ROY SOC LON, V53, P370, DOI [10.1098/rstl.1763.0053, DOI 10.1098/RSTL.1763.0053]
[6]   The molecular diagnosis of porcine viral diseases:: A review [J].
Belák, S .
ACTA VETERINARIA HUNGARICA, 2005, 53 (01) :113-124
[7]  
Belak S, 2001, Expert Rev Mol Diagn, V1, P434
[8]   USE OF COMBINED SHEWHART-CUSUM CONTROL CHARTS IN INTERNAL QUALITY-CONTROL OF ENZYME-LINKED IMMUNOSORBENT ASSAYS FOR THE TYPING OF FOOT-AND-MOUTH-DISEASE VIRUS-ANTIGEN [J].
BLACKSELL, SD ;
GLEESON, LJ ;
LUNT, RA ;
CHAMNANPOOD, C .
REVUE SCIENTIFIQUE ET TECHNIQUE DE L OFFICE INTERNATIONAL DES EPIZOOTIES, 1994, 13 (03) :687-699
[9]   Implementation of internal laboratory quality control procedures for the monitoring of ELISA performance at a regional veterinary laboratory [J].
Blacksell, SD ;
Cameron, AR ;
Chamnanpood, C ;
Chamnanpood, P ;
Tatong, D ;
Monpolsiri, M ;
Westbury, HA .
VETERINARY MICROBIOLOGY, 1996, 51 (1-2) :1-9
[10]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310