Predictors and Clinical Impact of Inappropriate Implantable Cardioverter-Defibrillator Shocks in Korean Patients

被引:14
作者
Yang, Jeong Hoon [1 ]
Byeon, Kyeongmin [1 ]
Yim, Hye Ran [1 ]
Park, Jung Wae [1 ]
Park, Seung-Jung [1 ]
Huh, June [2 ]
Kim, June Soo [1 ]
On, Young Keun [1 ]
机构
[1] Sungkyunkwan Univ, Sch Med, Samsung Med Ctr, Div Cardiol,Dept Med,Cardiac & Vasc Ctr, Seoul 135710, South Korea
[2] Sungkyunkwan Univ, Sch Med, Samsung Med Ctr, Dept Pediat,Cardiac & Vasc Ctr, Seoul 135710, South Korea
关键词
Implantable Cardioverter-Defibrillator; Inappropriate Shock; ATRIAL-FIBRILLATION; HEART-FAILURE; DISCRIMINATION ALGORITHMS; THERAPY; SURVIVAL; TRIAL;
D O I
10.3346/jkms.2012.27.6.619
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Limited data are available on inappropriate shocks in Korean patients implanted with an implantable cardioverter-defibrillator (ICD). We investigated the impact of inappropriate shocks on clinical outcomes. This retrospective, single-center study included 148 patients treated between October 1999 and June 2011. The primary outcome was a composite event of all-cause mortality or hospitalization for any cardiac reason. The median follow-up duration was 29 months (interquartile range: 8 to 53). One or more inappropriate shocks occurred in 34 (23.0%) patients. A history of atrial fibrillation was the only independent predictor of inappropriate shock (hazard ratio [HR]: 4.16, 95% confidence interval [CI]: 1.89-9.15, P < 0.001). Atrial fibrillation was the most common cause of inappropriate shock (67.7%), followed by supraventricular tachycardia (23.5%), and abnormal sensing (8.8%). A composite event of all-cause mortality or hospitalizations for any cardiac reason during follow-up was not significantly different between patients with or without inappropriate shock (inappropriate shock vs no inappropriate shock: 35.3% vs 35.4%, adjusted HR: 1.06, 95% Cl: 0.49-2.29, P = 0.877). Inappropriate shocks do not affect clinical outcomes in patients implanted with an ICD, although the incidence of inappropriate shocks is high.
引用
收藏
页码:619 / 624
页数:6
相关论文
共 24 条
[1]   Safety, efficacy, and performance of new discrimination algorithms to reduce inappropriate and unnecessary shocks: the PainFree SST clinical study design [J].
Auricchio, Angelo ;
Meijer, Albert ;
Kurita, Takashi ;
Schloss, Edward ;
Brinkman, Karin ;
Claessens-van Ooijen, Marieke ;
Sterns, Laurence .
EUROPACE, 2011, 13 (10) :1484-1493
[2]   Amiodarone or an implantable cardioverter-defibrillator for congestive heart failure [J].
Bardy, GH ;
Lee, KL ;
Mark, DB ;
Poole, JE ;
Packer, DL ;
Boineau, R ;
Domanski, M ;
Troutman, C ;
Anderson, J ;
Johnson, G ;
McNulty, SE ;
Clapp-Channing, N ;
Davidson-Ray, LD ;
Fraulo, ES ;
Fishbein, DP ;
Luceri, RM ;
Ip, JH .
NEW ENGLAND JOURNAL OF MEDICINE, 2005, 352 (03) :225-237
[3]   Association between statin therapy and reductions in atrial fibrillation or flutter and inappropriate shock therapy [J].
Bhavnani, Sanjeev P. ;
Coleman, Craig I. ;
White, Charles M. ;
Clyne, Christopher A. ;
Yarlagadda, Ravi ;
Guertin, Danette ;
Kluger, Jeffrey .
EUROPACE, 2008, 10 (07) :854-859
[4]   Implantable cardioverter defibrillators and their role in heart failure progression [J].
Cevik, Cihan ;
Perez-Verdia, Alejandro ;
Nugent, Kenneth .
EUROPACE, 2009, 11 (06) :710-715
[5]   Inappropriate implantable cardioverter-defibrillator shocks in MADIT II [J].
Daubert, James P. ;
Zareba, Wojciech ;
Cannom, David S. ;
McNitt, Scott ;
Rosero, Spencer Z. ;
Wang, Paul ;
Schuger, Claudio ;
Steinberg, Jonathan S. ;
Higgins, Steven L. ;
Wilber, David J. ;
Klein, Helmut ;
Andrews, Mark L. ;
Hall, W. Jackson ;
Moss, Arthur J. .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2008, 51 (14) :1357-1365
[6]   Appropriate Therapy But Not Inappropriate Shocks Predict Survival in Implantable Cardioverter Defibrillator Patients [J].
Dichtl, Wolfgang ;
Wolber, Thomas ;
Paoli, Ursula ;
Bruellmann, Simon ;
Stuehlinger, Markus ;
Berger, Thomas ;
Spuller, Karin ;
Strasak, Alexander ;
Pachinger, Otmar ;
Haegeli, Laurent M. ;
Duru, Firat ;
Hintringer, Florian .
CLINICAL CARDIOLOGY, 2011, 34 (07) :433-436
[7]   Frequency and causes of implantable cardioverter-defibrillator therapies: Is device therapy proarrhythmic? [J].
Germano, JJ ;
Reynolds, M ;
Essebag, V ;
Josephson, ME .
AMERICAN JOURNAL OF CARDIOLOGY, 2006, 97 (08) :1255-1261
[8]   Prospective comparison of discrimination algorithms to prevent inappropriate ICD therapy: Primary results of the Rhythm ID Going Head to Head Trial [J].
Gold, Michael R. ;
Ahmad, Saleem ;
Browne, Kevin ;
Berg, Kellie Chase ;
Thackeray, Lisa ;
Berger, Ronald D. .
HEART RHYTHM, 2012, 9 (03) :370-377
[9]   Stability: An ICD detection criterion for discriminating atrial fibrillation from ventricular tachycardia [J].
Higgins, SL ;
Lee, RS ;
Kramer, RL .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 1995, 6 (12) :1081-1088
[10]   Prophylactic defibrillator implantation in patients with nonischemic dilated cardiomyopathy [J].
Kadish, A ;
Dyer, A ;
Daubert, JP ;
Quigg, R ;
Estes, NAM ;
Anderson, KP ;
Calkins, H ;
Hoch, D ;
Goldberger, J ;
Shalaby, A ;
Sanders, WE ;
Schaechter, A ;
Levine, JH .