A phantom study to determine the optimum size of a single collimator for shortening the treatment time in CyberKnife stereotactic radiosurgery of spherical targets

被引:5
作者
Harikrishnaperumal, Sudahar [2 ]
Kurup, Gopalakrishna [2 ]
Venkatraman, Murali [2 ]
Jagadeesan, Velmurugan [1 ]
机构
[1] Anna Univ, Dept Med Phys, Madras 600025, Tamil Nadu, India
[2] Apollo Specialty Hosp, Dept Radiotherapy, Madras 600035, Tamil Nadu, India
关键词
CyberKnife; treatment time; optimal single collimator; nodes; monitor units; beams; GAMMA-KNIFE RADIOSURGERY; ROBOTIC RADIOSURGERY; TRIGEMINAL NEURALGIA; TREATMENT PLANS; CONFORMITY;
D O I
10.1120/jacmp.v13i5.3864
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Prolonged treatment execution time is a concern in CyberKnife robotic radiosurgery. Beam reduction and node reduction technique, and monitor unit optimization methods are adopted to reduce the treatment time. Usage of single collimator in the CyberKnife treatment plan can potentially reduce collimator exchange time. An optimal single collimator, which yields an acceptable dose distribution along with minimum number of nodes, beams, and monitor units, can be a versatile alternative for shortening treatment time. The aim of the present study is to find the optimal single collimator in CyberKnife treatment planning to shorten the treatment time with the acceptable dose distribution. A spherical planning target volume PTV1 was drawn in an anthropomorphic head and neck phantom. Plans with same treatment goals were generated for all the 12 collimators independently. D95% was selected as the prescribing isodose and the prescribed dose was 10 Gy. The plan of the optimal collimator size was evaluated for conformity, homogeneity, and dose spillage outside the target. The optimum collimator size and the target dimensions were correlated. The study was repeated with two other target volumes PTV2 and PTV3 for generalizing the results. Collimator sizes just above the diameter of the spherical PTVs were yielding least number of nodes and beams with acceptable dose distributions. The collimator size of 35 mm is optimum for the PTV1, whose diameter is 31.4 mm. Similarly, 50 mm collimator is optimum for PTV2 (diameter=45.2 mm) and 20 mm collimator is optimum for PTV3 (Diameter = 17.3 mm). The total number of monitor units is found to reduce with increasing collimator size. Optimal single collimator is found to be useful for shortening the treatment time in spherical targets. Studies on two clinical targets, (a brain metastasis and a liver metastasis cases) show comparable results with the phantom study.
引用
收藏
页码:33 / 45
页数:13
相关论文
共 12 条
[1]   A dosimetric comparison between Gamma Knife and CyberKnife treatment plans for trigeminal neuralgia [J].
Descovich, Martina ;
Sneed, Penny K. ;
Barbaro, Nicholas M. ;
McDermott, Michael W. ;
Chuang, Cynthia F. ;
Barani, Igor J. ;
Nakamura, Jean L. ;
Ma, Lijun .
JOURNAL OF NEUROSURGERY, 2010, 113 :199-206
[2]  
Dieterich S, 2011, FRONT RADIAT THER ON, V43, P181, DOI 10.1159/000322423
[3]   The design, physical properties and clinical utility of an iris collimator for robotic radiosurgery [J].
Echner, G. G. ;
Kilby, W. ;
Lee, M. ;
Earnst, E. ;
Sayeh, S. ;
Schlaefer, A. ;
Rhein, B. ;
Dooley, J. R. ;
Lang, C. ;
Blanck, O. ;
Lessard, E. ;
Maurer, C. R., Jr. ;
Schlegel, W. .
PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (18) :5359-5380
[4]  
Hamamoto Yasushi, 2004, Radiat Med, V22, P442
[5]   The CyberKnife® Robotic Radiosurgery System in 2010 [J].
Kilby, W. ;
Dooley, J. R. ;
Kuduvalli, G. ;
Sayeh, S. ;
Maurer, C. R., Jr. .
TECHNOLOGY IN CANCER RESEARCH & TREATMENT, 2010, 9 (05) :433-452
[6]  
Kondziolka D, 1997, NEUROSURG CLIN N AM, V8, P79
[7]   CyberKnife: A new paradigm in radiotherapy [J].
Kurup, Gopalakrishna .
JOURNAL OF MEDICAL PHYSICS, 2010, 35 (02) :63-64
[8]   Advantages of mini-multileaf in stereotactic radiotherapy [J].
Kurup, P. G. G. ;
Murali, V. ;
Sankar, A. .
JOURNAL OF MEDICAL PHYSICS, 2007, 32 (01) :12-17
[9]   Dose conformity of Gamma Knife radiosurgery and risk factors for complications [J].
Nakamura, JL ;
Verhey, LJ ;
Smith, V ;
Petti, PL ;
Lamborn, KR ;
Larson, DA ;
Wara, WM ;
McDermott, MW ;
Sneed, PK .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2001, 51 (05) :1313-1319
[10]   A simple scoring ratio to index the conformity of radiosurgical treatment plans - Technical note [J].
Paddick, I .
JOURNAL OF NEUROSURGERY, 2000, 93 :219-222