Influence of cavity loss on an extrinsic Fabry-Perot cavity intensity-based pressure sensor

被引:1
作者
Lu, Tao [1 ,2 ]
机构
[1] China Univ Geosci, Sch Math & Phys, Wuhan 430074, Peoples R China
[2] China Univ Geosci, State Key Lab Geol Proc & Mineral Resource, Wuhan 430074, Peoples R China
基金
中国博士后科学基金; 中国国家自然科学基金;
关键词
OPTICAL-FIBER SENSORS; DIAPHRAGM; LASER;
D O I
10.1063/1.4929681
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
We present an extrinsic Fabry-Perot cavity intensity-based pressure sensor that mainly comprises a single-mode fiber end and an elastic monocrystalline silicon layer bonded to a silicon diaphragm. We investigated the influence of cavity loss on the performance indexes (PIS) of the intensity-based extrinsic Fabry-Perot cavity optical fiber pressure sensor. A buffer unit made of three incompressible oil cavities attenuated outside pressure and transformed pressure information into cavity length microchange information. Experimental results indicated that, under center quadrature-points within the linear regions of adjacent fringes, for an applied 40 kPa external pressure, cavity length was modulated by pressures of 69.9 kPa-109.9 kPa, 150.1 kPa-190 kPa, 220.1 kPa-259.9 kPa, and 279.9 kPa-319.9 kPa, output intensity ranges increased as 1 mu W, 1.02 mu W, 1.03 mu W, and 1.05 mu W, sensitivity increased as 0.01909 mu W/kPa, 0.01986 mu W/kPa, 0.02127 mu W/kPa, and 0.02387 mu W/kPa, but linearity degraded, as indicated by the standard deviation of linear fits of 0.02607, 0.02664, 0.02935, and 0.04879 due to cavity loss. Furthermore, the pressure ranges within the same quarter period decreased as 40 kPa, 37.45 kPa, 32.4 kPa, and 30.15 kPa. Consequently, the same lengths of linear regions within adjacent fringes of an approximately sinusoidal curve corresponded to different measurement ranges, linearities, and sensitivities. Initial cavity length must be chosen to optimize both signal strength and the PIS studied here in manufacturing this type sensor. (C) 2015 AIP Publishing LLC.
引用
收藏
页数:7
相关论文
共 25 条
  • [1] Analysis of the effect of imperfect fiber endfaces on the performance of extrinsic Fabry-Perot interferometric optical fiber sensors
    Arya, V
    deVries, MJ
    Athreya, M
    Wang, AB
    Claus, RO
    [J]. OPTICAL ENGINEERING, 1996, 35 (08) : 2262 - 2265
  • [2] Miniature Fabry-Perot pressure sensor created by using UV-molding process with an optical fiber based mold
    Bae, H.
    Yu, M.
    [J]. OPTICS EXPRESS, 2012, 20 (13): : 14573 - 14583
  • [3] Phase-shifting interferometry with equal phase steps by use of a frequency-tunable diode laser and a Fabry-Perot cavity
    Bitou, Y
    Inaba, H
    Hong, FL
    Takatsuji, T
    Onae, A
    [J]. APPLIED OPTICS, 2005, 44 (26) : 5403 - 5407
  • [4] Chitosan based fiber-optic Fabry-Perot humidity sensor
    Chen, L. H.
    Li, T.
    Chan, C. C.
    Menon, R.
    Balamurali, P.
    Shaillender, M.
    Neu, B.
    Ang, X. M.
    Zu, P.
    Wong, W. C.
    Leong, K. C.
    [J]. SENSORS AND ACTUATORS B-CHEMICAL, 2012, 169 : 167 - 172
  • [5] Miniature fiber-optic pressure sensor with a polymer diaphragm
    Cibula, E
    Donlagic, D
    [J]. APPLIED OPTICS, 2005, 44 (14) : 2736 - 2744
  • [6] In-line short cavity Fabry-Perot strain sensor for quasi distributed measurement utilizing standard OTDR
    Cibula, Edvard
    Donlagic, Denis
    [J]. OPTICS EXPRESS, 2007, 15 (14) : 8719 - 8730
  • [7] All-fiber high-sensitivity pressure sensor with SiO2 diaphragm
    Donlagic, D
    Cibula, E
    [J]. OPTICS LETTERS, 2005, 30 (16) : 2071 - 2073
  • [8] Filmed extrinsic Fabry-Perot interferometric sensors for the measurement of arbitrary refractive index of liquid
    Gao, Ran
    Jiang, Yi
    Ding, Wenhui
    Wang, Zhen
    Liu, Da
    [J]. SENSORS AND ACTUATORS B-CHEMICAL, 2013, 177 : 924 - 928
  • [9] A Coaxial Cable Fabry-Perot Interferometer for Sensing Applications
    Huang, Jie
    Wang, Tao
    Hua, Lei
    Fan, Jun
    Xiao, Hai
    Luo, Ming
    [J]. SENSORS, 2013, 13 (11): : 15252 - 15260
  • [10] Jiang Shao-Ji, 2005, Chinese Journal of Sensors and Actuators, V18, P165