Condensation Particle Counter Proportionality Calibration from 1 Particle.cm-3 to 104 Particles.cm-3

被引:11
作者
Owen, Miles [1 ]
Mulholland, George [2 ]
Guthrie, Will [3 ]
机构
[1] USA, Primary Stand Lab, US Army TMDE Act, Redstone Arsenal, AL 35898 USA
[2] Univ Maryland, Dept Mech Engn, College Pk, MD 20742 USA
[3] Natl Inst Stand & Technol, Gaithersburg, MD 20899 USA
关键词
NIST;
D O I
10.1080/02786826.2011.637998
中图分类号
TQ [化学工业];
学科分类号
0817 ;
摘要
This article presents a process to calibrate a condensation particle counter (CPC) over the concentration range from 1 particle.cm(-3) to 10(4) particles.cm(-3) with traceability to SI units as realized by the National Institute of Standards and Technology (NIST). The process combines two independent steps: a proportionality assessment over the range of the CPC using a consistent diluter and an absolute calibration of the CPC with an aerosol electrometer (AE) at high concentrations. The proportionality assessment, which is the focus of the article, is analogous to the attenuation method for testing the proportionality of high-power laser detectors. This procedure tests the proportionality of one CPC by itself and does not require the use of a calibrated reference CPC. The primary calibration of the CPC with the AE at high concentrations allows for an absolute calibration with NIST traceability, resulting in a simple one-parameter correction to the measured CPC data. The proportionality test enables uncertainties to be assigned to the CPC over an extended range of concentration beyond the lower detection limit of the AE, thereby establishing measurement traceability for low concentrations. The relative expanded uncertainty of the CPC with a coverage factor of k = 2 is 2.8% over the range of about 1 particle.cm(-3) to 10(4) particles.cm(-3).
引用
收藏
页码:444 / 450
页数:7
相关论文
共 10 条
[1]   A SAMPLE EXTRACTION DILUTER FOR ULTRAFINE AEROSOL SAMPLING [J].
BROCKMANN, JE ;
LIU, BYH ;
MCMURRY, PH .
AEROSOL SCIENCE AND TECHNOLOGY, 1984, 3 (04) :441-451
[2]   Calibration of a Condensation Particle Counter Using a NIST Traceable Method [J].
Fletcher, R. A. ;
Mulholland, G. W. ;
Winchester, M. R. ;
King, R. L. ;
Klinedinst, D. B. .
AEROSOL SCIENCE AND TECHNOLOGY, 2009, 43 (05) :425-441
[3]   Calibration of Condensation Particle Counters for Legislated Vehicle Number Emission Measurements [J].
Giechaskiel, B. ;
Wang, X. ;
Horn, H. -G. ;
Spielvogel, J. ;
Gerhart, C. ;
Southgate, J. ;
Jing, L. ;
Kasper, M. ;
Drossinos, Y. ;
Krasenbrink, A. .
AEROSOL SCIENCE AND TECHNOLOGY, 2009, 43 (12) :1164-1173
[4]   An accurate, continuously adjustable dilution system (1:10 to 1:10(4)) for submicron aerosols [J].
Hueglin, C ;
Scherrer, L ;
Burtscher, H .
JOURNAL OF AEROSOL SCIENCE, 1997, 28 (06) :1049-1055
[5]   ELECTROSPRAY-CONDENSATION PARTICLE COUNTER - A MOLECULE-COUNTING LC-DETECTOR FOR MACROMOLECULES [J].
LEWIS, KC ;
DOHMEIER, DM ;
JORGENSON, JW ;
KAUFMAN, SL ;
ZARRIN, F ;
DORMAN, FD .
ANALYTICAL CHEMISTRY, 1994, 66 (14) :2285-2292
[6]   Nonlinearity measurements of high-power laser detectors at NIST [J].
Li, XY ;
Scott, T ;
Yang, S ;
Cromer, C ;
Dowell, M .
JOURNAL OF RESEARCH OF THE NATIONAL INSTITUTE OF STANDARDS AND TECHNOLOGY, 2004, 109 (04) :429-434
[7]  
NIST, 2009, NIST POL TRAC
[8]  
Taylor B., 1994, NIST TECHNICAL NOTE, V1297
[9]   Evaluation of a condensation particle counter for vehicle emission measurement: Experimental procedure and effects of calibration aerosol material [J].
Wang, Xiaoliang ;
Caldow, Robert ;
Sem, Gilmore J. ;
Hama, Naoya ;
Sakurai, Hiromu .
JOURNAL OF AEROSOL SCIENCE, 2010, 41 (03) :306-318
[10]   Towards traceable particle number concentration standard: Single charged aerosol reference (SCAR) [J].
Yli-Ojanpera, Jaakko ;
Makela, Jyrki M. ;
Marjamaki, Marko ;
Rostedt, Antti ;
Keskinen, Jorma .
JOURNAL OF AEROSOL SCIENCE, 2010, 41 (08) :719-728