Accuracy of computer-aided implant placement

被引:267
作者
Van Assche, N. [1 ]
Vercruyssen, M. [1 ]
Coucke, W. [2 ]
Teughels, W. [1 ]
Jacobs, R. [3 ]
Quirynen, M. [1 ]
机构
[1] Katholieke Univ Leuven, Dept Periodontol, Fac Med, B-3000 Louvain, Belgium
[2] Sci Inst Publ Hlth, Dept Qual Med Labs, B-1050 Brussels, Belgium
[3] Catholic Univ Louvain, Oral Imaging Ctr, B-3000 Louvain, Belgium
关键词
accuracy implant placement; computer-assisted implant dentistry; dental implants; guided surgery; navigation; stereolithography; STEREOLITHOGRAPHIC SURGICAL GUIDES; TEMPLATE; SURGERY; TOMOGRAPHY; FLAPLESS; RELIABILITY; SYSTEMS; IMAGES;
D O I
10.1111/j.1600-0501.2012.02552.x
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Aim To assess the accuracy of static computer-guided implant placement. Material and methods Electronic and manual literature searches were conducted to collect information on the accuracy of static computer-guided implant placement and meta-regression analyses were performed to summarize and analyse the overall accuracy. The latter included a search for correlations between factors such as: support (teeth/mucosa/bone), number of templates, use of fixation pins, jaw, template production, guiding system, guided implant placement. Results Nineteen accuracy studies met the inclusion criteria. Meta analysis revealed a mean error of 0.99 mm (ranging from 0 to 6.5 mm) at the entry point and of 1.24 mm (ranging from 0 to 6.9 mm) at the apex. The mean angular deviation was 3.81 degrees (ranging from 0 to 24.9 degrees). Significant differences for all deviation parameters was found for implant-guided placement compared to placement without guidance. Number of templates used was significant, influencing the apical and angular deviation in favour for the single template. Study design and jaw location had no significant effect. Less deviation was found when more fixation pins were used (significant for entry). Conclusion Computer-guided implant placement can be accurate, but significant deviations have to be taken into account. Randomized studies are needed to analyse the impact of individual parameters in order to allow optimization of this technique. Moreover, a clear overview on indications and benefits would help the clinicians to find the right candidates.
引用
收藏
页码:112 / 123
页数:12
相关论文
共 43 条
[1]   Implant Placement Accuracy When Using Stereolithographic Template as a Surgical Guide: Preliminary Results [J].
Al-Harbi, Saad A. ;
Sun, Albert Y. T. .
IMPLANT DENTISTRY, 2009, 18 (01) :46-51
[2]   Accuracy of Two Stereolithographic Guide Systems for Computer-Aided Implant Placement: A Computed Tomography-Based Clinical Comparative Study [J].
Arisan, Volkan ;
Karabuda, Z. Cuneyt ;
Ozdemir, Tayfun .
JOURNAL OF PERIODONTOLOGY, 2010, 81 (01) :43-51
[3]   Factors influencing transfer accuracy of cone beam CT-derived template-based implant placement [J].
Behneke, Alexandra ;
Burwinkel, Matthias ;
Behneke, Nikolaus .
CLINICAL ORAL IMPLANTS RESEARCH, 2012, 23 (04) :416-423
[4]   Accuracy assessment of cone beam computed tomography-derived laboratory-based surgical templates on partially edentulous patients [J].
Behneke, Alexandra ;
Burwinkel, Matthias ;
Knierim, Kristian ;
Behneke, Nikolaus .
CLINICAL ORAL IMPLANTS RESEARCH, 2012, 23 (02) :137-143
[5]   Accuracy of image-guided implantology [J].
Brief, J ;
Edinger, D ;
Hassfeld, S ;
Eggers, G .
CLINICAL ORAL IMPLANTS RESEARCH, 2005, 16 (04) :495-501
[6]   Depth deviation and occurrence of early surgical complications or unexpected events using a single stereolithographic surgi-guide [J].
Cassetta, M. ;
Stefanelli, L. V. ;
Giansanti, M. ;
Di Mambro, A. ;
Calasso, S. .
INTERNATIONAL JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2011, 40 (12) :1377-1387
[7]  
Cassetta M., 2011, CLIN IMPLANT DENT R, DOI [10.1111/j1708-8208.2011.00369.x, DOI 10.1111/J1708-8208.2011.00369.X]
[8]   Modular Preoperative Planning Software for Computer-Aided Oral Implantology and the Application of a Novel Stereolithographic Template: A Pilot Study [J].
Chen, Xiaojun ;
Yuan, Jianbing ;
Wang, Chengtao ;
Huang, Yuanliang ;
Kang, Lu .
CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH, 2010, 12 (03) :181-193
[9]  
D'haese J., 2009, CLIN IMPLANT DENT R, V14, P321
[10]  
D'haese J., 2011, CLIN IMPLANT DENT R, DOI [10.1111/j1708-8208.2011.00353.x, DOI 10.1111/J1708-8208.2011.00353.X]