X-ray beam monitor made by thin-film CVD single-crystal diamond

被引:9
|
作者
Marinelli, Marco [1 ]
Milani, E. [1 ]
Prestopino, G. [1 ]
Verona, C. [1 ]
Verona-Rinati, G. [1 ]
Angelone, M. [2 ]
Pillon, M. [2 ]
Kachkanov, V. [3 ]
Tartoni, N. [3 ]
Benetti, M. [4 ]
Cannata, D. [4 ]
Di Pietrantonio, F. [4 ]
机构
[1] Univ Roma Tor Vergata, Dip Ingn Ind, I-00133 Rome, Italy
[2] Assoc EURATOM ENEA Fus, I-00044 Frascati, Roma, Italy
[3] Diamond Light Source, Didcot OX11 0DE, Oxon, England
[4] Ist Acust OM Corbino, CNR, I-00133 Rome, Italy
来源
JOURNAL OF SYNCHROTRON RADIATION | 2012年 / 19卷
关键词
CVD diamond; beam position monitor; X-ray detector; Schottky diode; UV PHOTODETECTORS; CONFIGURATION; DETECTORS;
D O I
10.1107/S0909049512038186
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A novel beam position monitor, operated at zero bias voltage, based on high-quality chemical-vapor-deposition single-crystal Schottky diamond for use under intense synchrotron X-ray beams was fabricated and tested. The total thickness of the diamond thin-film beam monitor is about 60 mu m. The diamond beam monitor was inserted in the B16 beamline of the Diamond Light Source synchrotron in Harwell (UK). The device was characterized under monochromatic high-flux X-ray beams from 6 to 20 keV and a micro-focused 10 keV beam with a spot size of approximately 2 mu m x 3 mu m square. Time response, linearity and position sensitivity were investigated. Device response uniformity was measured by a raster scan of the diamond surface with the micro-focused beam. Transmissivity and spectral responsivity versus beam energy were also measured, showing excellent performance of the new thin-film single-crystal diamond beam monitor.
引用
收藏
页码:1015 / 1020
页数:6
相关论文
共 50 条
  • [21] X-ray photoelectron spectroscopy studies of CVD diamond films
    Fan, Y
    Fitzgerald, AG
    John, P
    Troupe, CE
    Wilson, JIB
    SURFACE AND INTERFACE ANALYSIS, 2002, 34 (01) : 703 - 707
  • [22] Modelling the response of polymeric thin-film devices subject to X-ray irradiation
    Alderson, A
    Vinton, SJ
    Wimbush, SC
    SYNTHETIC METALS, 1999, 102 (1-3) : 951 - 952
  • [23] Flexible X-Ray Detector Array Fabricated With Oxide Thin-Film Transistors
    Lujan, R. A.
    Street, R. A.
    IEEE ELECTRON DEVICE LETTERS, 2012, 33 (05) : 688 - 690
  • [24] Silicon X-Ray Detector With Integrated Thin-Film Transistor for Biomedical Applications
    Wang, Kai
    Karim, Karim S.
    IEEE ELECTRON DEVICE LETTERS, 2010, 31 (02) : 147 - 149
  • [25] Growth and characterisation of CVD diamond wires for x-ray detection
    Manfredotti, C
    Fizzotti, F
    LoGiudice, A
    Mucera, G
    Polesello, P
    Vittone, E
    Mankelevich, YA
    Suetin, NV
    DIAMOND AND RELATED MATERIALS, 1997, 6 (08) : 1051 - 1056
  • [26] Effects of microstructure of films on CVD diamond X-ray detectors
    Zhang, ML
    Xia, YB
    Wang, LJ
    Gu, BB
    SENSORS AND ACTUATORS A-PHYSICAL, 2005, 120 (02) : 349 - 353
  • [27] A single crystal chemical vapour deposition diamond soft X-ray spectrometer
    Bodie, C. S.
    Lioliou, G.
    Lefeuvre, G.
    Barnett, A. M.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2021, 989
  • [28] Very Fast and Primingless Single-Crystal-Diamond X-Ray Dosimeters
    Trucchi, D. M.
    Allegrini, P.
    Calvani, P.
    Galbiati, A.
    Oliver, K.
    Conte, G.
    IEEE ELECTRON DEVICE LETTERS, 2012, 33 (04) : 615 - 617
  • [29] Selection of CVD Diamond Crystals for X-ray Monochromator Applications Using X-ray Diffraction Imaging
    Stoupin, Stanislav
    Krawczyk, Thomas
    Liu, Zunping
    Franck, Carl
    CRYSTALS, 2019, 9 (08):
  • [30] X-ray detectors based on (100)-textured CVD diamond films
    Zhang, ML
    Gu, BB
    Wang, LJ
    Xia, YB
    PHYSICS LETTERS A, 2004, 332 (3-4) : 320 - 325