Epidural multiorifice catheters function as single-orifice catheters:: An in vitro study

被引:23
作者
Fegley, Allison J. [2 ]
Lerman, Jerrold [1 ]
Wissler, Richard [2 ]
机构
[1] SUNY Buffalo, Womens & Childrens Hosp, Dept Anesthesia, Buffalo, NY 14222 USA
[2] Univ Rochester, Strong Mem Hosp, Rochester, NY 14642 USA
关键词
D O I
10.1213/ane.0b013e31817f1fc2
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
In an in vitro study, we determined the flow rates required to use selective orifices of multiorifice catheters. Saline was infused at rates between 1 and 360 mL/h through Portex and Braun 20-gauge multiorifice catheters using Baxter, Abbott Laboratories, and Alaris infusion pumps. The numbers of orifices used via infusion and manual injection, and the pressure within the catheter during Continuous In fusion, were recorded. Infusion rates < 80 mL/h used one orifice, between 100 and 280 mL/h used two orifices, and > 300 mL/h used three orifices. Catheter pressures with Braun catheters were 40%, greater than with Portex catheters. Manual injections by all 12 residents used all three orifices. Twenty-gauge multiorifice catheters function as single-orifice catheters at clinically relevant infusion rates, but function as multiorifice catheters during manual boluses.
引用
收藏
页码:1079 / 1081
页数:3
相关论文
共 8 条
[1]   Comparison of continuous and intermittent administration of extradural bupivacaine for analgesia after lower abdominal surgery [J].
Duncan, LA ;
Fried, MJ ;
Lee, A ;
Wildsmith, JAW .
BRITISH JOURNAL OF ANAESTHESIA, 1998, 80 (01) :7-10
[2]   Intermittent vs continuous administration of epidural ropivacaine with fentanyl for analgesia during labour [J].
Fettes, P. D. W. ;
Moore, C. S. ;
Whiteside, J. B. ;
Mcleod, G. A. ;
Wildsmith, J. A. W. .
BRITISH JOURNAL OF ANAESTHESIA, 2006, 97 (03) :359-364
[3]   Distribution of solution in the epidural space: Examination by cryomicrotome section [J].
Hogan, Q .
REGIONAL ANESTHESIA AND PAIN MEDICINE, 2002, 27 (02) :150-156
[4]   Multiport epidural catheters - Can orifice location be tested? [J].
Leighton, BL ;
Katsiris, SE ;
Halpern, SH ;
Wilson, DB ;
Kronberg, JE .
ANESTHESIOLOGY, 2000, 92 (06) :1840-1842
[5]   An in vitro evaluation of flow from multihole epidural catheters during continuous infusion with four different infusion pumps [J].
McAtamney, D ;
O'Hare, C ;
Fee, JPH .
ANAESTHESIA, 1999, 54 (07) :664-669
[6]  
POWER I, 1988, ANAESTHESIA, V43, P876
[7]   A comparative study of sequential epidural bolus technique and continuous epidural infusion [J].
Ueda, K ;
Ueda, W ;
Manabe, M .
ANESTHESIOLOGY, 2005, 103 (01) :126-129
[8]   A randomized comparison of programmed intermittent epidural bolus with continuous epidural infusion for labor analgesia [J].
Wong, CA ;
Ratliff, JT ;
Sullivan, JT ;
Scavone, BM ;
Toledo, P ;
McCarthv, RT .
ANESTHESIA AND ANALGESIA, 2006, 102 (03) :904-909