An investigation of image guidance dose for breast radiotherapy

被引:29
|
作者
Alvarado, Rosemerie [1 ]
Booth, Jeremy T. [1 ,2 ]
Bromley, Regina M. [1 ,2 ]
Gustafsson, Helen B. [1 ,2 ]
机构
[1] Univ Sydney, Sch Phys, Inst Med Phys, Sydney, NSW 2006, Australia
[2] Royal N Shore Hosp, Northern Sydney Canc Ctr, St Leonards, NSW 2065, Australia
来源
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS | 2013年 / 14卷 / 03期
关键词
breast cancer; cone-beam CT; absorbed dose; radiochromic film; MC simulation; BEAM COMPUTED-TOMOGRAPHY; SETUP; CT; DOSIMETRY; CANCER; FILM; REGISTRATION; IRRADIATION; PROTOCOL; RISK;
D O I
10.1120/jacmp.v14i3.4085
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Cone-beam computed tomography (CBCT) is used for external-beam radiation therapy setup and target localization. As with all medical applications of ionizing radiation, radiation exposure should be managed safely and optimized to achieve the necessary image quality using the lowest possible dose. The present study investigates doses from standard kilovoltage kV radiographic and CBCT imaging protocol, and proposes two novel reduced dose CBCT protocols for the setup of breast cancer patients undergoing external beam radiotherapy. The standard thorax kV and low-dose thorax CBCT protocols available on Varian's On-Board Imaging system was chosen as the reference technique for breast imaging. Two new CBCT protocols were created by modifying the low-dose thorax protocol, one with a reduced gantry rotation range ("Under breast" protocol) and the other with a reduced tube current-time product setting ("Low dose thorax 10ms" protocol). The absorbed doses to lungs, heart, breasts, and skin were measured using XRQA2 radiochromic film in an anthropomorphic female phantom. The absorbed doses to lungs, heart, and breasts were also calculated using the PCXMC Monte Carlo simulation software. The effective dose was calculated using the measured doses to the included organs and the ICRP 103 tissue weighting factors. The deviation between measured and simulated organ doses was between 3% and 24%. Reducing the protocol exposure time to half of its original value resulted in a reduction in the absorbed doses of the organs of 50%, while the reduced rotation range resulted in a dose reduction of at least 60%. Absorbed doses obtained from "Low dose thorax 10ms" protocol were higher than the doses from our departments orthogonal kV-kV imaging protocol. Doses acquired from "Under breast" protocol were comparable to the doses measured from the orthogonal kV-kV imaging protocol. The effective dose per fraction using the CBCT for standard low-dose thorax protocol was 5.00 +/- 0.30 mSv; for the "Low dose thorax 10ms" protocol it was 2.44 +/- 0.21 mSv; and for the "Under breast" protocol it was 1.23 +/- 0.25 mSv when the image isocenter was positioned at the phantom center and 1.17 +/- 0.30 mSv when the image isocenter was positioned in the middle of right breast. The effective dose per fraction using the orthogonal kV-kV protocol was 1.14 +/- 0.16 mSv. The reduction of the scan exposure time or beam rotation range of the CBCT imaging significantly reduced the dose to the organs investigated. The doses from the "Under breast" protocol and orthogonal kV-kV imaging protocol were comparable. Simulated organ doses correlated well with measured doses. Effective doses from imaging techniques should be considered with the increase use of kV imaging protocols in order to support the use of IGRT.
引用
收藏
页码:25 / 38
页数:14
相关论文
共 50 条
  • [1] The Effect of Image Guidance on Dose Distributions in Breast Boost Radiotherapy
    Donovan, E. M.
    Brooks, C.
    Mitchell, R. A.
    Mukesh, M.
    Coles, C. E.
    Evans, P. M.
    Harris, E. J.
    CLINICAL ONCOLOGY, 2014, 26 (11) : 671 - 676
  • [2] The Emerging Role of Image-Guidance for Breast Radiotherapy
    Al-Hallaq, H.
    Sonke, J.
    MEDICAL PHYSICS, 2012, 39 (06) : 3907 - 3907
  • [3] Management of Image-Guidance Dose in Radiotherapy Treatment Planning
    Ding, G.
    Deeley, M.
    Coffey, C.
    MEDICAL PHYSICS, 2008, 35 (06)
  • [4] Toward Molecular Image-guidance for Intraoperative Breast Radiotherapy
    Carpenter, C. M.
    Pratx, G.
    Sun, C.
    Xing, L.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2011, 81 (02): : S90 - S90
  • [5] A dose distribution overlay technique for image guidance during prostate radiotherapy
    Smyth, G.
    McCallum, H. M.
    Lambert, E. L.
    Lawrence, G. P.
    BRITISH JOURNAL OF RADIOLOGY, 2008, 81 (971): : 890 - 896
  • [6] Image guidance for FLASH radiotherapy
    El Naqa, Issam
    Pogue, Brian W.
    Zhang, Rongxiao
    Oraiqat, Ibrahim
    Parodi, Katia
    MEDICAL PHYSICS, 2022, 49 (06) : 4109 - 4122
  • [7] AN EXPERIENCE OF MALE BILATERAL BREAST TREATMENT FOR RADIOTHERAPY UNDER IMAGE GUIDANCE.
    Krishnapan, C.
    Velmurugan, T.
    Thirumalai Swamy, S.
    Devang, B.
    Jani, K.
    RADIOTHERAPY AND ONCOLOGY, 2010, 96 : S424 - S424
  • [8] The importance of image guidance for precision radiotherapy
    van Herk, M
    RADIOTHERAPY AND ONCOLOGY, 2004, 73 : S94 - S94
  • [9] NEW DEVELOPMENTS IN IMAGE GUIDANCE FOR RADIOTHERAPY
    Kron, Tomas
    CANCER FORUM, 2012, 36 (02) : 80 - 85
  • [10] Image guidance for precise conformal radiotherapy
    Mackie, TR
    Kapatoes, J
    Ruchala, K
    Lu, WG
    Wu, C
    Olivera, G
    Forrest, L
    Tome, W
    Welsh, J
    Jeraj, R
    Harari, P
    Reckwerdt, P
    Paliwal, B
    Ritter, M
    Keller, H
    Fowler, J
    Mehta, M
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2003, 56 (01): : 89 - 105