Standard dosimetry of 125I seeds against prostate cancer

被引:0
作者
Ito, Akira [1 ]
Takahashi, Yutaka [1 ]
Sumida, Iori [1 ]
Yunoki, Akira [2 ]
Hino, Yoshiho [2 ]
机构
[1] Japanese Fdn Canc Res, Inst Canc, Koto Ku, 3-10-6 Ariake, Tokyo 1358550, Japan
[2] Natl Inst Adv Ind Sci & Technol, Tsukuba, Ibaraki, Japan
来源
WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING 2006, VOL 14, PTS 1-6 | 2007年 / 14卷
关键词
I-125; seed; air-kerma; standard; calibration;
D O I
暂无
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
We have established the standard seed assay method using the well-type ionization chambers, after two years experiences of measurements. It is clear that the single seed measurement whose dose sensitivity is traceable to the standard is the basic method. Currently NIST traceable well-type ionization chamber is the only the system to be used for the seed assay. In order to establish Japanese national standard for the air kerma strength measurement and also to establish the traceability chain of calibration, we started to establish national standard ionization chamber at Japanese primary dosimetry laboratory (AIST).
引用
收藏
页码:2076 / +
页数:2
相关论文
共 2 条
  • [1] LARRY A, 2004, MED PHYS, V31, P675
  • [2] MARK J, 2004, MED PHYS, V31, P633